An image analysis method to quantify the lateral pivot shift test.

Yuichi Hoshino1, Paulo Araujo, James J Irrgang, Freddie H Fu, Volker Musahl.   

Abstract

PURPOSE: Although various kinematic measurements with advanced technology have been used for quantitative evaluation of the pivot shift test, there is no clinically available quantification method of the pivot shift test. The purpose was to describe a novel image analysis technique for quantitative assessment of the pivot shift test using universally available and affordable devices.
METHODS: Five ACL deficient knees were tested during examination under anesthesia. Three skin markers were attached to bony landmarks on the lateral side of the knee joint, (1) Gerdy's tubercle, (2) fibular head, and (3) lateral epicondyle. A standard digital video camera captured motion of the lateral aspect of the knee during the pivot shift test. The image was processed into a 2-dimensional (2-D) coordinate system with Image J software (National Institute of Health, USA) to trace the three landmarks. The anteroposterior (AP) position of the femur was calculated on consecutive still images extracted from the video recording. AP translation over time was reported.
RESULTS: The reduction phase of the pivot shift could be tracked consistently by a sudden anterior translation of the distal femur. The sudden anterior translation of the lateral epicondyle was on average 3.7 ± 2.1 mm and occurred within 0.2 ± 0.1 s from the start of this anterior translation till the end.
CONCLUSION: The sudden shift of the lateral compartment of the knee joint was successfully detected by this newly developed image analysis measurement method. This image analysis technique facilitates a simple and affordable method to evaluate the lateral pivot shift test. LEVEL OF EVIDENCE: Diagnostic studies, Level IV.

Entities:  

Mesh:

Year:  2011        PMID: 22203048      PMCID: PMC3309139          DOI: 10.1007/s00167-011-1845-x

Source DB:  PubMed          Journal:  Knee Surg Sports Traumatol Arthrosc        ISSN: 0942-2056            Impact factor:   4.342


Introduction

The lateral pivot shift test [5, 6] is widely used to evaluate rotational instability after ACL injury and treatment. The result of this test is subjectively determined by the examiner and therefore highly variable [13]. There is no quantitative evaluation system that can be used in clinical practice despite extensive research efforts to measure this instability. The primary reason for the difficulty to establish an evaluation system is the complexity of the pivot shift movement that is composed of a six degree-of-freedom tibial internal–external (i–e) rotation, varus–valgus (v–v) rotation, and anterior–posterior (a–p) translation [3]. Bedi et al. [1] used a navigation system to demonstrate a good correlation between a–p translation of the lateral compartment and the results of the clinical pivot shift test grade. The results of this study indicated that each clinical graded correlated to an increment of approximately 6 mm of lateral compartment translation [1]. This suggests that the lateral compartment translation could reflect the clinical pivot shift test grade. Most reported devices used in experimental trials of pivot shift measurement, such as navigation system [1, 2, 9, 16], electromagnetic sensor [4, 7], and accelerometer [10, 11], are not universally available and associated with considerable costs. Digital camera technology was developed over the past 20 years and today is available with high-resolution digital images (10.0 Mpixel) for <100 dollars. Digital video recordings can be processed with free-of-charge image analysis software. This report describes a newly developed image analysis for capturing the lateral pivot shift movement using simple and affordable devices.

Materials and methods

Capture of the video image during the pivot shift test

Ring-shaped white stickers 14 mm in diameter (white portable reinforcements, item #636156, Staples, Inc., Framingham, MA, USA) were attached as skin markers on the bony landmarks, (1) lateral epicondyle, (2) Gerdy’s tubercle, and (3) fibular head. The distance between the centers of the markers located over Gerdy’s tubercle and the fibular head was preliminarily measured by a ruler. A video recording of the lateral aspect of the knee during the manually performed pivot shift test is taken by an ordinary digital camera (Cyber-shot® W120 Digital, Sony, Tokyo, Japan). The pivot shift test is performed with the maneuver reported by Galway and McIntosh [5] (Fig. 1). The camera angle and distance to the knee is determined as to capture the three skin markers as clearly as possible in the center of the frame throughout the pivot shift test. The background behind the knee is covered by a monotone sheet to reduce noise during the processing stage.
Fig. 1

Manual pivot shift test is captured with standard digital camera. Three points are tracked, (1) lateral epicondyle, (2) Gerdy tubercle, and (3) fibula head

Manual pivot shift test is captured with standard digital camera. Three points are tracked, (1) lateral epicondyle, (2) Gerdy tubercle, and (3) fibula head

Image analysis using NIH image J software (National Institute of Health, Bethesda, MD, USA)

The video recording containing the pivot shift test is reformatted as avi file, which is compatible with Image J. This movie file is trimmed to have one sequence of the pivot shift test, which normally takes about 1 s to perform. The avi file is opened as a stack of images by Image J (Fig. 2a). The images are cropped to capture only the three markers within the imaging frames throughout the test (Fig. 2b). The color information of the images is converted to 8-bit grayscale, and the brightness and contrast are adjusted to highlight the skin markers. The areas of the skin markers are selected by the area detection tool by the level of grayscale. The detecting threshold of the grayscale level is adjusted to maximize the areas of the markers and to minimize the unwanted area selection (Fig. 2c). The XY plots of the centroid of the detected areas and the outlines of the detected area are provided after conducting particle analysis. The XY plots of the centroids are used to calculate the a–p translation of the lateral compartment. The outline data facilitate identification of each detected skin marker area and the corresponding XY plot of its centroid location (Fig. 3).
Fig. 2

a A picture processed from a movie of the pivot shift test with image processing software (ImageJ). b A cropped picture with clearly captured skin markers. c The areas of the skin markers are detected and colored as red

Fig. 3

Outline drawings of the detected areas in consecutive frames during the test (in this case, n = 13 frames were processed during the pivot shift)

a A picture processed from a movie of the pivot shift test with image processing software (ImageJ). b A cropped picture with clearly captured skin markers. c The areas of the skin markers are detected and colored as red Outline drawings of the detected areas in consecutive frames during the test (in this case, n = 13 frames were processed during the pivot shift)

Calculation of lateral compartment translation

In each frame, the XY plot of the intersection point, termed as “pivot point (P)” between the tibial horizontal line, which is defined as a line connecting the centroids of Gerdy’s tubercle (G) and fibular head markers (F), and a perpendicular line from the centroid of the lateral epicondyle marker (L) to the tibial horizontal line is calculated (Fig. 4). The ratio of the perpendicular offset length of the lateral epicondyle point from the Gerdy’s tubercle point (distance “b” in Fig. 4) to the length of the tibial horizontal line (distance “a” in Fig. 4) is calculated from the XY plot data. The femoral AP position from the Gerdy’s point can be calculated by multiplying the ratio by the distance of the two tibial points (from preliminary measurement).
Fig. 4

Calculation of the position of the distal femur. Using ImageJ software, a line is drawn to connect Gerdy’s tubercle (G) with the fibula head (F; line a). A perpendicular line is drawn from the lateral epicondyle (L) to line a. The intersection between the two lines is termed “pivot point” (P). A third line is drawn from G to P (line b). The position of the distal femur is calculated as line [(a actual measurement) × (b/a from image analysis)]

Calculation of the position of the distal femur. Using ImageJ software, a line is drawn to connect Gerdy’s tubercle (G) with the fibula head (F; line a). A perpendicular line is drawn from the lateral epicondyle (L) to line a. The intersection between the two lines is termed “pivot point” (P). A third line is drawn from G to P (line b). The position of the distal femur is calculated as line [(a actual measurement) × (b/a from image analysis)]

Accuracy of the image analysis measurement

Preliminary validation of lateral compartment translation measured by our image analysis system was done using a cadaveric knee specimen in comparison with the actual bony movement that was measured by an electromagnetic tracking system [8, 14]. In the ACL dissected knee specimen, lateral compartment translation during the pivot shift test was 3.0 ± 0.8 mm for three repeated tests, whereas actual bony movement, i.e., posterior tibial translation, was 22.8 ± 0.4 mm. Values were smaller than the actual bony movement, but lateral compartment translation were consistently observed.

Results

Five ACL deficient knees were tested. The pivot shift test was performed by a single surgeon and evaluated as grade 1 positive. Image analysis was conducted by an individual who was blinded to the clinical grade for the pivot shift test. The femoral AP position from the Gerdy’s point suddenly decreased by 3.7 ± 2.1 mm (max 6.4 mm/min 1.5 mm) in the ACL deficient knees during the reduction phase of the pivot shift test, while the contralateral intact knees did not demonstrate such anteroposterior shift (Fig. 5). The anterior translation of the distal femur relative to the tibia occurred within 0.2 ± 0.1 s.
Fig. 5

Graph showing the anterior-posterior (AP) translation of the distal femur (y axis) during the pivot shift test over time (x axis). Sudden anterior translation (5.7 mm) of the distal femur occurs within 0.17 s (arrow) during the reduction phase of the manual pivot shift test

Graph showing the anterior-posterior (AP) translation of the distal femur (y axis) during the pivot shift test over time (x axis). Sudden anterior translation (5.7 mm) of the distal femur occurs within 0.17 s (arrow) during the reduction phase of the manual pivot shift test

Discussion

The most important finding of the present study was that the lateral compartment translation during the pivot shift test in the ACL deficient knee was consistently observed and quantitatively measured by the newly developed image analysis technique. This study described the use of a standard digital camera system and free-of-charge image processing software to quantify the sudden anterior translation of the lateral compartment during the reduction phase of the lateral pivot shift test. As Galway et al. [5] described the lateral pivot shift phenomenon in their original report as “forward subluxation of the lateral tibial plateau on the femoral condyle and its spontaneous reduction” [5], the a–p translation in the lateral compartment is reflected in clinical grading of the pivot shift test [1]. Tracking of skin motion in this study was possible with separate markers on distal femur and proximal tibia. The different directional skin motion of bony landmarks on distal femur and proximal tibia could be detected by dynamic video motion image analysis. There is a lack of quantitative evaluation systems for the pivot shift test, which are available for use in daily clinical practice mainly because of cost, operating space, and complexity. Navigation systems are the most commonly used device for experimental trials of kinematics measurements during the pivot shift test [1, 2, 9, 16]. However, this system requires surgical intervention for fixing the tracking markers rigidly to the bones and is associated with high costs. Electromagnetic systems can be an economical alternative for accurate knee kinematics measurement during the pivot shift test [4, 7]. Noninvasive measurement of the pivot shift is possible using simple straps [7], but their clinical use is limited by skin motion artifacts and metal interference that is extremely common in an operating or examining room environment. Although these systems have a great advantage in their capability to comprehensively examine knee kinematics during pivot shift test, the mere detection of the pivot shift movement may not require such complex systems. Based on the report that the acceleration of the pivot shift movement is related to clinical grading [7], the accelerometer has been increasingly utilized for simplified evaluation of the pivot shift test [10, 11]. However, these systems are exclusively for experimental use, and, even if they become available in the future, it is still unknown whether they can be economical. In contrast, the image analysis technique introduced in this report demands only a standard digital camera, free-of-charge image analysis software provided by a public institution (National Institute of Health, USA), and a few stationery stickers, which can be easily obtained. One of the inevitable limitations of this system is the manual performance of the pivot shift test, similar to previous experimental studies for quantitative measurement of the pivot shift test [1, 2, 4, 7, 9–11, 15]. Although an examiner might involuntarily reduce the magnitude of the pivot shift in ACL deficient knee during the manual procedure, it is difficult to elicit the pivot shift in intact knees. Therefore, as long as the examiner makes an effort to maximize the pivot shift based on his/her professional skill, the resultant knee motion should be consistent and larger when compared to external loading, such as described with the mechanized pivot shift [12]. The quantitative measurement of the pivot shift test from a “best effort” clinical examination can be used for further analysis. Although preliminary validation study was conducted with a single specimen, the accuracy of the digital analysis measurement needs to be fully validated using larger sample size especially for the disparity between skin movement and actual translation of the bones comprising the lateral compartment, which is problematic in patients with a larger body habitus. The skin movement detected by this study is relatively small compared to the actual lateral compartment translation that is described as greater than 6 mm in the ACL deficient knees according to Bedi et al. [1]. This small magnitude of the skin movement might affect the sensitivity of this measurement system. Also, there was little skin movement in the short distance between Gerdy’s tubercle and the fibula head in this study. Further investigation is warranted to compare the skin surface shift detected by this image analysis to bony kinematics. Objective evaluation of the pivot shift test has been greatly needed for clinical and research needs to quantify rotational laxity in the ACL injured and reconstructed knees and could be achieved using this image analysis technique with a minimal effort and investment.

Conclusion

The sudden shift of the lateral compartment of the knee joint was successfully detected by this newly developed video-based image analysis measurement method. This image analysis technique facilitates a simple, reliable, and affordable measurement to evaluate the lateral pivot shift test.
  14 in total

1.  An analysis of the pivot shift phenomenon. The knee motions and subluxations induced by different examiners.

Authors:  F R Noyes; E S Grood; J F Cummings; R R Wroble
Journal:  Am J Sports Med       Date:  1991 Mar-Apr       Impact factor: 6.202

2.  The effect of graft tensioning in anatomic 2-bundle ACL reconstruction on knee joint kinematics.

Authors:  Yuichi Hoshino; Ryosuke Kuroda; Kouki Nagamune; Koji Nishimoto; Masayoshi Yagi; Kiyonori Mizuno; Shinichi Yoshiya; Masahiro Kurosaka
Journal:  Knee Surg Sports Traumatol Arthrosc       Date:  2006-12-23       Impact factor: 4.342

3.  In vivo measurement of the pivot-shift test in the anterior cruciate ligament-deficient knee using an electromagnetic device.

Authors:  Yuichi Hoshino; Ryosuke Kuroda; Kouki Nagamune; Masayoshi Yagi; Kiyonori Mizuno; Motoi Yamaguchi; Hirotsugu Muratsu; Shinichi Yoshiya; Masahiro Kurosaka
Journal:  Am J Sports Med       Date:  2007-03-09       Impact factor: 6.202

4.  In vivo analysis of the pivot shift phenomenon during computer navigated ACL reconstruction.

Authors:  Clayton G Lane; Russell F Warren; Fatima C Stanford; Daniel Kendoff; Andrew D Pearle
Journal:  Knee Surg Sports Traumatol Arthrosc       Date:  2008-05       Impact factor: 4.342

5.  Analysis of the graft bending angle at the femoral tunnel aperture in anatomic double bundle anterior cruciate ligament reconstruction: a comparison of the transtibial and the far anteromedial portal technique.

Authors:  Koji Nishimoto; Ryosuke Kuroda; Kiyonori Mizuno; Yuichi Hoshino; Kouki Nagamune; Seiji Kubo; Masayoshi Yagi; Motoi Yamaguchi; Shinichi Yoshiya; Masahiro Kurosaka
Journal:  Knee Surg Sports Traumatol Arthrosc       Date:  2008-12-02       Impact factor: 4.342

6.  Navigation evaluation of the pivot-shift phenomenon during double-bundle anterior cruciate ligament reconstruction: is the posterolateral bundle more important?

Authors:  Yasuyuki Ishibashi; Eiichi Tsuda; Yuji Yamamoto; Harehiko Tsukada; Satoshi Toh
Journal:  Arthroscopy       Date:  2008-12-18       Impact factor: 4.772

7.  Clinical relevance of static and dynamic tests after anatomical double-bundle ACL reconstruction.

Authors:  Simone Bignozzi; Stefano Zaffagnini; Nicola Lopomo; Freddie H Fu; James J Irrgang; Maurilio Marcacci
Journal:  Knee Surg Sports Traumatol Arthrosc       Date:  2009-06-30       Impact factor: 4.342

8.  Evaluation of rotational instability in the anterior cruciate ligament deficient knee using triaxial accelerometer: a biomechanical model in porcine knees.

Authors:  Akira Maeyama; Yuichi Hoshino; Anibal Debandi; Yuki Kato; Kazuhiko Saeki; Shigehiro Asai; Bunsei Goto; Patrick Smolinski; Freddie H Fu
Journal:  Knee Surg Sports Traumatol Arthrosc       Date:  2011-02-25       Impact factor: 4.342

9.  The lateral pivot shift: a symptom and sign of anterior cruciate ligament insufficiency.

Authors:  H R Galway; D L MacIntosh
Journal:  Clin Orthop Relat Res       Date:  1980 Mar-Apr       Impact factor: 4.176

10.  Intraoperative measurement of knee kinematics in reconstruction of the anterior cruciate ligament.

Authors:  A M J Bull; P H Earnshaw; A Smith; M V Katchburian; A N A Hassan; A A Amis
Journal:  J Bone Joint Surg Br       Date:  2002-09
View more
  25 in total

1.  Clinical grading of the pivot shift test correlates best with tibial acceleration.

Authors:  Mattias Ahldén; Paulo Araujo; Yuichi Hoshino; Kristian Samuelsson; Kellie K Middleton; Kouki Nagamune; Jón Karlsson; Volker Musahl
Journal:  Knee Surg Sports Traumatol Arthrosc       Date:  2012-01-04       Impact factor: 4.342

2.  Use of a gyroscope sensor to quantify tibial motions during a pivot shift test.

Authors:  Per Henrik Borgstrom; Keith L Markolf; Brock Foster; Frank A Petrigliano; David R McAllister
Journal:  Knee Surg Sports Traumatol Arthrosc       Date:  2013-07-25       Impact factor: 4.342

Review 3.  Posterior tibial translation resulting from the posterior drawer manoeuver in cadaveric knee specimens: a systematic review.

Authors:  Marcin Kowalczuk; Marie-Claude Leblanc; Benjamin B Rothrauff; Richard E Debski; Volker Musahl; Nicole Simunovic; Olufemi R Ayeni
Journal:  Knee Surg Sports Traumatol Arthrosc       Date:  2015-04-03       Impact factor: 4.342

4.  Methods to diagnose acute anterior cruciate ligament rupture: a meta-analysis of instrumented knee laxity tests.

Authors:  Carola F van Eck; Miette Loopik; Michel P van den Bekerom; Freddie H Fu; Gino M M J Kerkhoffs
Journal:  Knee Surg Sports Traumatol Arthrosc       Date:  2012-10-17       Impact factor: 4.342

5.  Quantitative evaluation of the pivot shift by image analysis using the iPad.

Authors:  Yuichi Hoshino; Paulo Araujo; Mattias Ahldén; Kristian Samuelsson; Bart Muller; Marcus Hofbauer; Megan R Wolf; James J Irrgang; Freddie H Fu; Volker Musahl
Journal:  Knee Surg Sports Traumatol Arthrosc       Date:  2013-01-23       Impact factor: 4.342

Review 6.  Quantifying the pivot shift test: a systematic review.

Authors:  Nicola Lopomo; Stefano Zaffagnini; Andrew A Amis
Journal:  Knee Surg Sports Traumatol Arthrosc       Date:  2013-03-02       Impact factor: 4.342

7.  Individualized ACL reconstruction.

Authors:  Paulo H Araujo; Mauricio Kfuri Junior; Bruno Ohashi; Yuichi Hoshino; Stephano Zaffagnini; Kristian Samuelsson; Jon Karlsson; Freddie Fu; Volker Musahl
Journal:  Knee Surg Sports Traumatol Arthrosc       Date:  2014-03-13       Impact factor: 4.342

8.  Quantitative pivot shift assessment using combined inertial and magnetic sensing.

Authors:  David R Labbé; Di Li; Guy Grimard; Jacques A de Guise; Nicola Hagemeister
Journal:  Knee Surg Sports Traumatol Arthrosc       Date:  2014-05-15       Impact factor: 4.342

Review 9.  Electromagnetic tracking of the pivot-shift.

Authors:  Ryosuke Kuroda; Yuichi Hoshino
Journal:  Curr Rev Musculoskelet Med       Date:  2016-06

Review 10.  Knee instability scores for ACL reconstruction.

Authors:  Ata A Rahnemai-Azar; Jan-Hendrik Naendrup; Ashish Soni; Adam Olsen; Jason Zlotnicki; Volker Musahl
Journal:  Curr Rev Musculoskelet Med       Date:  2016-06
View more

北京卡尤迪生物科技股份有限公司 © 2022-2023.