Intravenous fluids cause systemic bias in a conductivity-based point-of-care hematocrit meter.

Patrick Wu1, Timothy E Morey, Neil S Harris, Nikolaus Gravenstein, Mark J Rice.   

Abstract

BACKGROUND: Point-of-care (POC) devices measuring hematocrit rely on determination of electrical conductivity of whole blood. We hypothesized that some frequently administered IV fluids independently alter blood conductivity and confound hematocrit determination.
METHODS: Whole human blood was diluted to predetermined hematocrit values with normal saline, lactated Ringer solution, hetastarch, or plasma. Electrical conductivity and hematocrit (i-STAT® and spun methods) were measured at each dilution. In separate experiments, the effects of propofol and heparin were noted on these variables.
RESULTS: Greater dilution significantly increased conductivity irrespective of diluent type. The magnitude of the conductivity slopes increased in order for plasma, hetastarch, lactated Ringer solution, and normal saline dilution. Moreover, each slope varied from every other slope (all P < 0.0001), and 94.2% of hematocrit values measured by i-STAT (n = 211 of 224) were less than those for the spun method. Dilution with plasma, normal saline, lactated Ringer solution, and hetastarch caused bias (Bland-Altman limits of agreement) of -2.7% (-6.9/1.4), -4.6% (-7.3/-2.0), -4.8% (-7.8/-1.7), and -2.0% (-5.6/1.9), respectively. The Cohen κ agreement values (5th-95th confidence interval) for a transfusion trigger of 30% were 0.90 (all values, 0.85-0.95), 0.25 (hematocrit <30%, 0.02-0.48), and 0.21 (hematocrit 18%-30%, 0.01-0.42). Clinically relevant concentrations of propofol and heparin had minimal effects on electrical conductivity or hematocrit determination.
CONCLUSIONS: Dilution of blood with frequently used IV solutions affects whole blood conductivity determinations and thereby decreases hematocrits measured by a POC device relying on this method as compared with spun hematocrit. Conductivity-based hematocrit POC devices should be cautiously interpreted when hemodilution is present.

Entities:  

Mesh:

Substances:

Year:  2011        PMID: 22156329     DOI: 10.1213/ANE.0b013e31823fecbd

Source DB:  PubMed          Journal:  Anesth Analg        ISSN: 0003-2999            Impact factor:   5.108


  7 in total

1.  Dynamic electrochemistry: a step in the right direction.

Authors:  Mark J Rice
Journal:  J Diabetes Sci Technol       Date:  2011-09-01

Review 2.  Accuracy of point-of-care glucose measurements.

Authors:  Annette Rebel; Mark A Rice; Brenda G Fahy
Journal:  J Diabetes Sci Technol       Date:  2012-03-01

3.  Evaluation of multiwave pulse total-hemoglobinometer during general anesthesia.

Authors:  Daisuke Toyoda; Rie Yasumura; Mitsue Fukuda; Ryoichi Ochiai; Yoshifumi Kotake
Journal:  J Anesth       Date:  2013-10-22       Impact factor: 2.078

4.  Noninvasively Measured Hemoglobin Concentration Reflects Arterial Hemoglobin Concentration Before but Not After Cardiopulmonary Bypass in Patients Undergoing Coronary Artery or Valve Surgery.

Authors:  Matthias L Riess; Paul S Pagel
Journal:  J Cardiothorac Vasc Anesth       Date:  2016-03-24       Impact factor: 2.628

5.  Performance of point-of-care diagnostics for glucose, lactate, and hemoglobin in the management of severe malaria in a resource-constrained hospital in Uganda.

Authors:  Michael Hawkes; Andrea L Conroy; Robert O Opoka; Sophie Namasopo; W Conrad Liles; Chandy C John; Kevin C Kain
Journal:  Am J Trop Med Hyg       Date:  2014-03-03       Impact factor: 2.345

Review 6.  Blood glucose measurement in the intensive care unit: what is the best method?

Authors:  Huong T Le; Neil S Harris; Abby J Estilong; Arvid Olson; Mark J Rice
Journal:  J Diabetes Sci Technol       Date:  2013-03-01

7.  Influence of measurement principle on total hemoglobin value.

Authors:  Keisuke Hayashi; Takashi Hitosugi; Yoshifumi Kawakubo; Norihisa Kitamoto; Takeshi Yokoyama
Journal:  BMC Anesthesiol       Date:  2020-04-07       Impact factor: 2.217

  7 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.