Evaluations of an adaptive planning technique incorporating dose feedback in image-guided radiotherapy of prostate cancer.

Han Liu1, Qiuwen Wu.   

Abstract

PURPOSE: Online image guidance (IG) has been used to effectively correct the setup error and inter-fraction rigid organ motion for prostate cancer. However, planning margins are still necessary to account for uncertainties such as deformation and intra-fraction motion. The purpose of this study is to investigate the effectiveness of an adaptive planning technique incorporating offline dose feedback to manage inter-fraction motion and residuals from online correction.
METHODS: Repeated helical CT scans from 28 patients were included in the study. The contours of prostate and organs-at-risk (OARs) were delineated on each CT, and online IG was simulated by matching center-of-mass of prostate between treatment CTs and planning CT. A seven beam intensity modulated radiation therapy (IMRT) plan was designed for each patient on planning CT for a total of 15 fractions. Dose distribution at each fraction was evaluated based on actual contours of the target and OARs from that fraction. Cumulative dose up to each fraction was calculated by tracking each voxel based on a deformable registration algorithm. The cumulative dose was compared with the dose from initial plan. If the deviation exceeded the pre-defined threshold, such as 2% of the D₉₉ to the prostate, an adaptive planning technique called dose compensation was invoked, in which the cumulative dose distribution was fed back to the treatment planning system and the dose deficit was made up through boost radiation in future treatment fractions. The dose compensation was achieved by IMRT inverse planning. Two weekly compensation delivery strategies were simulated: one intended to deliver the boost dose in all future fractions (schedule A) and the other in the following week only (schedule B). The D₉₉ to prostate and generalized equivalent uniform dose (gEUD) to rectal wall and bladder were computed and compared with those without the dose compensation.
RESULTS: If only 2% underdose is allowed at the end of the treatment course, then 11 patients fail. If the same criteria is assessed at the end of each week (every five fractions), then 14 patients fail, with three patients failing the 1st or 2nd week but passing at the end. The average dose deficit from these 14 patients was 4.4%. They improved to 2% after the weekly compensation. Out of these 14 patients who needed dose compensation, ten passed the dose criterion after weekly dose compensation, three patients failed marginally, and one patient still failed the criterion significantly (10% deficit), representing 3.6% of the patient population. A more aggressive compensation frequency (every three fractions) could successfully reduce the dose deficit to the acceptable level for this patient. The average number of required dose compensation re-planning per patient was 0.82 (0.79) per patient for schedule A (B) delivery strategy. The doses to OARs were not significantly different from the online IG only plans without dose compensation.
CONCLUSIONS: We have demonstrated the effectiveness of offline dose compensation technique in image-guided radiotherapy for prostate cancer. It can effectively account for residual uncertainties which cannot be corrected through online IG. Dose compensation allows further margin reduction and critical organs sparing.

Entities:  

Mesh:

Year:  2011        PMID: 22149819      PMCID: PMC3230638          DOI: 10.1118/1.3658567

Source DB:  PubMed          Journal:  Med Phys        ISSN: 0094-2405            Impact factor:   4.071


  28 in total

Review 1.  Emergent technologies for 3-dimensional image-guided radiation delivery.

Authors:  David A Jaffray
Journal:  Semin Radiat Oncol       Date:  2005-07       Impact factor: 5.934

2.  Accuracy of finite element model-based multi-organ deformable image registration.

Authors:  K K Brock; M B Sharpe; L A Dawson; S M Kim; D A Jaffray
Journal:  Med Phys       Date:  2005-06       Impact factor: 4.071

3.  Geometric and dosimetric evaluations of an online image-guidance strategy for 3D-CRT of prostate cancer.

Authors:  Qiuwen Wu; Giovanni Ivaldi; Jian Liang; David Lockman; Di Yan; Alvaro Martinez
Journal:  Int J Radiat Oncol Biol Phys       Date:  2006-04-01       Impact factor: 7.038

4.  An adaptive off-line procedure for radiotherapy of prostate cancer.

Authors:  Tonnis T Nuver; Mischa S Hoogeman; Peter Remeijer; Marcel van Herk; Joos V Lebesque
Journal:  Int J Radiat Oncol Biol Phys       Date:  2007-02-15       Impact factor: 7.038

5.  Application of dose compensation in image-guided radiotherapy of prostate cancer.

Authors:  Qiuwen Wu; Jian Liang; Di Yan
Journal:  Phys Med Biol       Date:  2006-02-21       Impact factor: 3.609

6.  A gradient inverse planning algorithm with dose-volume constraints.

Authors:  S V Spirou; C S Chui
Journal:  Med Phys       Date:  1998-03       Impact factor: 4.071

7.  Strategies to reduce the systematic error due to tumor and rectum motion in radiotherapy of prostate cancer.

Authors:  Mischa S Hoogeman; Marcel van Herk; Josien de Bois; Joos V Lebesque
Journal:  Radiother Oncol       Date:  2005-02       Impact factor: 6.280

8.  Clinical application of a repositioning scheme, using gold markers and electronic portal imaging.

Authors:  Frank Van den Heuvel; James Fugazzi; Ed Seppi; Jeffrey D Forman
Journal:  Radiother Oncol       Date:  2006-04-03       Impact factor: 6.280

9.  A material sensitivity study on the accuracy of deformable organ registration using linear biomechanical models.

Authors:  Y Chi; J Liang; D Yan
Journal:  Med Phys       Date:  2006-02       Impact factor: 4.071

10.  Technical aspects of daily online positioning of the prostate for three-dimensional conformal radiotherapy using an electronic portal imaging device.

Authors:  Michael G Herman; Thomas M Pisansky; Jon J Kruse; Joann I Prisciandaro; Brian J Davis; Bernard F King
Journal:  Int J Radiat Oncol Biol Phys       Date:  2003-11-15       Impact factor: 7.038

View more
  7 in total

1.  A new method to reconstruct intra-fractional prostate motion in volumetric modulated arc therapy.

Authors:  Y Chi; N H Rezaeian; C Shen; Y Zhou; W Lu; M Yang; R Hannan; X Jia
Journal:  Phys Med Biol       Date:  2017-07-07       Impact factor: 3.609

Review 2.  Target margins in radiotherapy of prostate cancer.

Authors:  Slav Yartsev; Glenn Bauman
Journal:  Br J Radiol       Date:  2016-07-20       Impact factor: 3.039

3.  Validation of a dose warping algorithm using clinically realistic scenarios.

Authors:  Y G Roussakis; H Dehghani; S Green; G J Webster
Journal:  Br J Radiol       Date:  2015-03-20       Impact factor: 3.039

4.  Dose tracking assessment for image-guided radiotherapy of the prostate bed and the impact on clinical workflow.

Authors:  Lucia Clara Orlandini; Marianna Coppola; Christian Fulcheri; Luna Cernusco; Pei Wang; Luca Cionini
Journal:  Radiat Oncol       Date:  2017-04-28       Impact factor: 3.481

5.  Application programming in C# environment with recorded user software interactions and its application in autopilot of VMAT/IMRT treatment planning.

Authors:  Henry Wang; Lei Xing
Journal:  J Appl Clin Med Phys       Date:  2016-11-08       Impact factor: 2.102

6.  Varian ethos online adaptive radiotherapy for prostate cancer: Early results of contouring accuracy, treatment plan quality, and treatment time.

Authors:  Mikel Byrne; Ben Archibald-Heeren; Yunfei Hu; Amy Teh; Rhea Beserminji; Emma Cai; Guilin Liu; Angela Yates; James Rijken; Nick Collett; Trent Aland
Journal:  J Appl Clin Med Phys       Date:  2021-11-29       Impact factor: 2.102

7.  A framework for deformable image registration validation in radiotherapy clinical applications.

Authors:  Raj Varadhan; Grigorios Karangelis; Karthik Krishnan; Susanta Hui
Journal:  J Appl Clin Med Phys       Date:  2013-01-02       Impact factor: 2.102

  7 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.