Technical aspects of daily online positioning of the prostate for three-dimensional conformal radiotherapy using an electronic portal imaging device.

Michael G Herman1, Thomas M Pisansky, Jon J Kruse, Joann I Prisciandaro, Brian J Davis, Bernard F King.   

Abstract

PURPOSE: To develop a real-time electronic portal imaging device (EPID) procedure to identify intraprostatic gold markers and correct daily variations in target position during external beam radiotherapy for prostate cancer. METHODS AND MATERIALS: Pretherapy electronic portal images (EPIs) were acquired with a small portion of the therapeutic 18-MV dose from an orthogonal pair of treatment fields. The position of the intraprostatic gold markers on the EPIs was aligned with that on the simulation digitally reconstructed radiographs. If the initial three-dimensional target displacement (3DI) exceeded 5 mm or rotations exceeded 3 degrees, the beam was realigned before the remainder of the dose was delivered. Field-only EPIs were then acquired for all fields and offline analysis was performed to determine the final 3D target placement (3DF).
RESULTS: Twenty patients completed protocol-specified treatment, and all markers were identified on 99.6% of the pretherapy EPIs. Overall, 53% of treatment fractions were realigned. The mean 3DI was 5.6 mm in all patients (range 3.7-9.3), and the mean 3DF was 2.8 mm (range 1.6-4.0), which was statistically significant (p < 0.001). Rotational corrections were made on 15% of treatments. Mean treatment duration was 1.4 min greater for protocol patients than for similar patients in whom localization was not performed.
CONCLUSIONS: Frequent field misalignment occurs when external fiducial marks are used for patient alignment. Misalignments can be readily and rapidly identified and corrected with an EPID-based online correction procedure that integrates commercially available equipment and software.

Entities:  

Mesh:

Substances:

Year:  2003        PMID: 14575846     DOI: 10.1016/s0360-3016(03)00766-1

Source DB:  PubMed          Journal:  Int J Radiat Oncol Biol Phys        ISSN: 0360-3016            Impact factor:   7.038


  20 in total

1.  Evaluations of an adaptive planning technique incorporating dose feedback in image-guided radiotherapy of prostate cancer.

Authors:  Han Liu; Qiuwen Wu
Journal:  Med Phys       Date:  2011-12       Impact factor: 4.071

2.  Consideration of the likely benefit from implementation of prostate image-guided radiotherapy using current margin sizes: a radiobiological analysis.

Authors:  G S J Tudor; Y L Rimmer; T B Nguyen; M A Cowen; S J Thomas
Journal:  Br J Radiol       Date:  2012-02-14       Impact factor: 3.039

Review 3.  Technological advances in radiotherapy for esophageal cancer.

Authors:  Milan Vosmik; Jiri Petera; Igor Sirak; Miroslav Hodek; Petr Paluska; Jiri Dolezal; Marcela Kopacova
Journal:  World J Gastroenterol       Date:  2010-11-28       Impact factor: 5.742

4.  Effects of point configuration on the accuracy in 3D reconstruction from biplane images.

Authors:  Jacek Dmochowski; Kenneth R Hoffmann; Vikas Singh; Jinhui Xu; Daryl P Nazareth
Journal:  Med Phys       Date:  2005-09       Impact factor: 4.071

5.  Quality assurance for kilo- and megavoltage in-room imaging and localization for off- and online setup error correction.

Authors:  James M Balter; Larry E Antonuk
Journal:  Int J Radiat Oncol Biol Phys       Date:  2008       Impact factor: 7.038

6.  Prostate intrafraction motion evaluation using kV fluoroscopy during treatment delivery: a feasibility and accuracy study.

Authors:  Justus Adamson; Qiuwen Wu
Journal:  Med Phys       Date:  2008-05       Impact factor: 4.071

7.  Fast internal marker tracking algorithm for onboard MV and kV imaging systems.

Authors:  W Mao; R D Wiersma; L Xing
Journal:  Med Phys       Date:  2008-05       Impact factor: 4.071

Review 8.  Fiducial marker guided prostate radiotherapy: a review.

Authors:  Angela G M O'Neill; Suneil Jain; Alan R Hounsell; Joe M O'Sullivan
Journal:  Br J Radiol       Date:  2016-09-23       Impact factor: 3.039

9.  Dosimetric impact of intrafractional patient motion in pediatric brain tumor patients.

Authors:  Chris Beltran; John Trussell; Thomas E Merchant
Journal:  Med Dosim       Date:  2009-02-07       Impact factor: 1.482

10.  A fiducial detection algorithm for real-time image guided IMRT based on simultaneous MV and kV imaging.

Authors:  Weihua Mao; Nadeem Riaz; Louis Lee; Rodney Wiersma; Lei Xing
Journal:  Med Phys       Date:  2008-08       Impact factor: 4.071

View more

北京卡尤迪生物科技股份有限公司 © 2022-2023.