| Literature DB >> 22116373 |
Daniela G Bittar1, Thaís Gimenez, Caroline C Morais, Monique S De Benedetto, Mariana M Braga, Fausto Medeiros Mendes.
Abstract
The aim of this study was to evaluate the influence of dental plaque and moisture on performance of a laser fluorescence device in detecting occlusal and proximal caries lesions in primary teeth. Fifty-five occlusal and 58 proximal sites on primary molars were evaluated using a DIAGNOdent pen (LFpen) device. For the drying time study, the evaluations were performed in: (I) moist teeth; (II) teeth dried for 3 s, or (III) dried for 15 s. For the plaque study, the evaluations were done in sites: (I) without plaque; (II) with plaque, and (III) after cleaning. Evaluation of the teeth sections in stereomicroscope was the reference standard method. LF pen values, sensitivity, specificity, and accuracy were compared. The values obtained after 15 s of drying were higher than those obtained with moist teeth or dried for 3 s at both occlusal and proximal surfaces. However, there was no change in the performance in detecting caries lesions. With regard to the presence of plaque, there was no significant change in the readings of the device, but specificity was decreased in occlusal surfaces with plaque. At proximal surfaces, however, no significant differences were observed. In conclusion, the moisture conditions do not influence significantly the performance of the LFpen, but the presence of plaque can affect its performance in detecting occlusal caries lesions in primary teeth.Entities:
Mesh:
Year: 2011 PMID: 22116373 DOI: 10.1007/s10103-011-1025-8
Source DB: PubMed Journal: Lasers Med Sci ISSN: 0268-8921 Impact factor: 3.161