Comparison of intraocular pressure measured by Pascal dynamic contour tonometry and Goldmann applanation tonometry.

J Y F Ku1, H V Danesh-Meyer, J P Craig, G D Gamble, C N J McGhee.   

Abstract

AIMS: To compare the intraocular pressure (IOP) measurements obtained using the Pascal dynamic contour tonometer (PDCT) with the standard Goldmann applanation tonometer (GAT) and to correlate these with central corneal thickness (CCT) in patients with normal corneas.
METHODS: A prospective, masked, comparative case series of 116 eyes from patients attending a glaucoma clinic. IOP was measured with PDCT by one examiner and with GAT by a masked, independent examiner. A mean of six CCT readings was used for analysis.
RESULTS: IOP measured by the two instruments correlated significantly (r=0.77; P<0.0001). IOP measured by GAT correlated strongly with CCT (r=0.37, P=0.0001) whereas the relationship between IOP measured by PDCT and CCT approached significance (r=0.17, P=0.073). The differences between GAT and PDCT measured IOP also correlated strongly with CCT (r=0.37, P<0.0001). The 95% limits of agreement between GAT and PDCT were +/-4.2 mmHg. Dividing the eyes into three groups on the basis of CCT, demonstrated those in the thickest tertile showed a poorer agreement between instruments and the GAT measured significantly higher IOP in this group (P=0.003) while the PDCT showed no significant differences with different CCTs (P=0.37).
CONCLUSION: Demonstration of the relative independence of PDCT IOP measurements from CCT supports a potential clinical role for this instrument, particularly for subjects with CCT outside the normal range.

Entities:  

Mesh:

Year:  2006        PMID: 15803173     DOI: 10.1038/sj.eye.6701849

Source DB:  PubMed          Journal:  Eye (Lond)        ISSN: 0950-222X            Impact factor:   3.775


  18 in total

1.  The influence of soft contact lenses on the intraocular pressure measurement.

Authors:  P G Firat; C Cankaya; S Doganay; M Cavdar; S Duman; E Ozsoy; B Koc
Journal:  Eye (Lond)       Date:  2011-11-11       Impact factor: 3.775

2.  Comparison between Pascal dynamic contour tonometer and Goldmann applanation tonometer after different types of refractive surgery.

Authors:  Antonios P Aristeidou; Georgios Labiris; Andreas Katsanos; Michalis Fanariotis; Nikitas C Foudoulakis; Vassilios P Kozobolis
Journal:  Graefes Arch Clin Exp Ophthalmol       Date:  2010-06-25       Impact factor: 3.117

3.  [Evaluation of correction formulas for tonometry : The Goldmann applanation tonometry in approximation to dynamic contour tonometry].

Authors:  J Wachtl; M Töteberg-Harms; S Frimmel; C Kniestedt
Journal:  Ophthalmologe       Date:  2017-08       Impact factor: 1.059

4.  Correlation Between Dynamic Contour Tonometry, Uncorrected and Corrected Goldmann Applanation Tonometry, and Stage of Glaucoma.

Authors:  Josephine Wachtl; Marc Töteberg-Harms; Sonja Frimmel; Malgorzata Roos; Christoph Kniestedt
Journal:  JAMA Ophthalmol       Date:  2017-06-01       Impact factor: 7.389

5.  Goldmann applanation tonometry and dynamic contour tonometry in eyes with elevated intraocular pressure (IOP): comparison in the same eyes after subsequent medical normalization of IOP.

Authors:  Chungkwon Yoo; Young Sub Eom; Yong Yeon Kim
Journal:  Graefes Arch Clin Exp Ophthalmol       Date:  2010-07-22       Impact factor: 3.117

6.  Comparison of dynamic contour tonometry and Goldmann applanation tonometry in relation to central corneal thickness in primary congenital glaucoma.

Authors:  Marcelo Lopes Silva Jordão; Vital Paulino Costa; Maria de Lourdes Veronese Rodrigues; Jayter Silva Paula
Journal:  Graefes Arch Clin Exp Ophthalmol       Date:  2012-05-04       Impact factor: 3.117

7.  Physiological diurnal variability and characteristics of the ocular pulse amplitude (OPA) with the dynamic contour tonometer (DCT-Pascal).

Authors:  Sayeh Pourjavan; Pierre-Yves Boëlle; Michéle Detry-Morel; Patrick De Potter
Journal:  Int Ophthalmol       Date:  2007-10-23       Impact factor: 2.031

8.  Comparison of Goldmann applanation and dynamic contour tonometry in a population of Mexican open-angle glaucoma patients.

Authors:  J Jimenez-Roman; F Gil-Carrasco; A Martinez; O Albis-Donado; J D De la Torre-Tovar
Journal:  Int Ophthalmol       Date:  2012-11-16       Impact factor: 2.031

9.  Correlation of intraocular pressure measured with goldmann and dynamic contour tonometry in normal and glaucomatous eyes.

Authors:  Tony Realini; Robert N Weinreb; Gerald Hobbs
Journal:  J Glaucoma       Date:  2009-02       Impact factor: 2.503

10.  Evaluation of lamina cribrosa thickness and depth in ocular hypertension.

Authors:  Jong Chul Han; Da-Ye Choi; Young Kyo Kwun; Wool Suh; Changwon Kee
Journal:  Jpn J Ophthalmol       Date:  2015-09-11       Impact factor: 2.447