Literature DB >> 22052084

Longitudinal stent deformation: a retrospective analysis of frequency and mechanisms.

Paul D Williams1, Mamas A Mamas, Kenneth P Morgan, Magdi El-Omar, Bernard Clarke, Anthony Bainbridge, Farzin Fath-Ordoubadi, Douglas G Fraser.   

Abstract

AIMS: Modern drug-eluting stents are constructed with thin struts and are easy to deliver and highly conformable. However, although innovative designs have enabled maintenance of radial strength, longitudinal strength may be lower with these stents and there have been recent reports of longitudinal stent compression of ostially deployed stents. We report the experience in our centre on longitudinal stent deformation and explore mechanisms of this complication and its frequency with various drug-eluting stent platforms. METHODS AND
RESULTS: Nine cases of longitudinal stent deformation were identified over a four year period representing 0.2% of cases and affected 0.097% of stents deployed. There were several mechanisms for this complication including compression by post-dilatation balloons, guide catheter extensions and proximal embolic protection devices. The rate of stent deformation varied from 0% in several stent types to 0.86% in the case of the Promus Element stent. There was one case of late stent thrombosis attributable to longitudinal stent deformation.
CONCLUSIONS: Longitudinal stent deformation can occur secondary to a variety of mechanisms and identification is important as, left untreated, it may be associated with a risk of stent thrombosis. Although seen with several different stents, in our series it was more commonly observed with the Promus Element stent.

Entities:  

Mesh:

Year:  2012        PMID: 22052084     DOI: 10.4244/EIJV8I2A41

Source DB:  PubMed          Journal:  EuroIntervention        ISSN: 1774-024X            Impact factor:   6.534


  27 in total

1.  Stent fracture and longitudinal compression detected on coronary CT angiography in the first- and new-generation drug-eluting stents.

Authors:  Mi Sun Chung; Dong Hyun Yang; Young-Hak Kim; Jae-Hyung Roh; Jihyun Song; Joon-Won Kang; Jung-Min Ahn; Duk-Woo Park; Soo-Jin Kang; Seung-Whan Lee; Cheol Whan Lee; Seong-Wook Park; Seung-Jung Park; Tae-Hwan Lim
Journal:  Int J Cardiovasc Imaging       Date:  2015-10-26       Impact factor: 2.357

2.  Longitudinal stent compression of everolimus-eluting stent: A report of 2 cases.

Authors:  Rajesh Vijayvergiya; Alok Kumar; Smit Shrivastava; Naveen K Kamana
Journal:  World J Cardiol       Date:  2013-08-26

Review 3.  Drug-eluting stents: the past, present, and future.

Authors:  Gregory Katz; Bhisham Harchandani; Binita Shah
Journal:  Curr Atheroscler Rep       Date:  2015-03       Impact factor: 5.113

4.  Longitudinal Stent Deformation Caused by Retraction of the Looped Main Branch Guidewire.

Authors:  Hung-Hao Lee; Po-Chao Hsu; Wen-Hsien Lee; Chun-Yuan Chu; Ho-Ming Su; Tsung-Hsien Lin; Wen-Chol Voon; Wen-Ter Lai; Sheng-Hsiung Sheu; Cheng-An Chiu
Journal:  Acta Cardiol Sin       Date:  2016-09       Impact factor: 2.672

5.  The potential hazard of a non-slip element balloon causing distal longitudinal stent deformation: the first clinical experience and in vitro assessment.

Authors:  Hiroki Shibutani; Yuzo Akita; Yohei Oishi; Hiroyuki Sueyoshi; Yu Mukai; Kotaro Yutaka; Yumie Matsui; Masahiro Yoshinaga; Masahiro Karakawa
Journal:  Cardiol J       Date:  2018-06-20       Impact factor: 2.737

6.  Stent collapse caused by balloon occlusion and aspiration system in saphenous vein graft intervention.

Authors:  Morihiko Takeda; Nobuyuki Shiba
Journal:  J Cardiol Cases       Date:  2017-07-17

7.  Longitudinal deformation - price we pay for better deliverability of coronary stent platforms.

Authors:  Ezhilan Janakiraman; Vijayakumar Subban; Suma M Victor; Ajit S Mullasari
Journal:  Indian Heart J       Date:  2012-07-27

8.  Coronary artery perforation after crush technique for stent compression.

Authors:  A Avci; S Fidan; G Alıcı; G Acar; C Toprak; S Izci; A M Esen
Journal:  Herz       Date:  2014-04-17       Impact factor: 1.443

9.  Needles in Our Technology Haystacks: Defining Efficacy Is Easy, Characterizing Complications Is the Challenge.

Authors:  Elazer R Edelman; Pei-Jiang Wang
Journal:  Circ Cardiovasc Interv       Date:  2017-11       Impact factor: 6.546

10.  Clinical, angiographic and procedural characteristics of longitudinal stent deformation.

Authors:  A Guler; Y Guler; E Acar; S M Aung; S C Efe; A Kilicgedik; C Y Karabay; S Barutcu; M K Tigen; S Pala; A İzgi; A M Esen; C Kirma
Journal:  Int J Cardiovasc Imaging       Date:  2016-05-19       Impact factor: 2.357

View more

北京卡尤迪生物科技股份有限公司 © 2022-2023.