Longitudinal deformation - price we pay for better deliverability of coronary stent platforms.

Ezhilan Janakiraman1, Vijayakumar Subban, Suma M Victor, Ajit S Mullasari.   

Abstract

Coronary stents, especially drug eluting stents (DES), have revolutionized the practice of interventional cardiology. Newer stents are manufactured by altering basic design characteristics to tackle complex coronary morphologies more effectively. Alteration in one particular attribute might affect other attributes adversely. Even though, reduction in the number and alteration of the orientation of the connectors improves the stent flexibility and deliverability, it adversely decreases the axial strength of the stent with resulting longitudinal stent deformation. A 67 year old female underwent percutaneous coronary intervention for a mid left anterior descending artery stenosis with a 2.75 × 16 mm Promus Element stent (Boston Scientific, Natick, Massachusetts). The stent got longitudinally distorted during post-stent balloon dilatation which was effectively managed with further dilatation with non-compliant balloon.
Copyright © 2012 Cardiological Society of India. Published by Elsevier B.V. All rights reserved.

Entities:  

Mesh:

Year:  2012        PMID: 23102394      PMCID: PMC3861144          DOI: 10.1016/j.ihj.2012.07.012

Source DB:  PubMed          Journal:  Indian Heart J        ISSN: 0019-4832


  7 in total

1.  Editorial: a new beginning and ending?

Authors:  Cindy L Grines
Journal:  J Interv Cardiol       Date:  2011-11-14       Impact factor: 2.279

2.  Stent longitudinal integrity bench insights into a clinical problem.

Authors:  John A Ormiston; Bruce Webber; Mark W I Webster
Journal:  JACC Cardiovasc Interv       Date:  2011-11-30       Impact factor: 11.195

3.  Longitudinal stent deformation: a retrospective analysis of frequency and mechanisms.

Authors:  Paul D Williams; Mamas A Mamas; Kenneth P Morgan; Magdi El-Omar; Bernard Clarke; Anthony Bainbridge; Farzin Fath-Ordoubadi; Douglas G Fraser
Journal:  EuroIntervention       Date:  2012-06-20       Impact factor: 6.534

4.  Engineering assessment of the longitudinal compression behaviour of contemporary coronary stents.

Authors:  Santosh Prabhu; Tanya Schikorr; Tamer Mahmoud; James Jacobs; Adriaan Potgieter; Charles Simonton
Journal:  EuroIntervention       Date:  2012-06-20       Impact factor: 6.534

5.  Longitudinal compression: a "new" complication with modern coronary stent platforms--time to think beyond deliverability?

Authors:  Colm G Hanratty; Simon J Walsh
Journal:  EuroIntervention       Date:  2011-11       Impact factor: 6.534

Review 6.  Coronary stents: current status.

Authors:  Scot Garg; Patrick W Serruys
Journal:  J Am Coll Cardiol       Date:  2010-08-31       Impact factor: 24.094

7.  Major stent deformation/pseudofracture of 7 Crown Endeavor/Micro Driver stent platform: incidence and causative factors.

Authors:  Mark Pitney; Kate Pitney; Nigel Jepson; Daniel Friedman; Tuan Nguyen-Dang; Joseph Matthews; Robert Giles; David Taylor
Journal:  EuroIntervention       Date:  2011-06       Impact factor: 6.534

  7 in total
  2 in total

1.  The potential hazard of a non-slip element balloon causing distal longitudinal stent deformation: the first clinical experience and in vitro assessment.

Authors:  Hiroki Shibutani; Yuzo Akita; Yohei Oishi; Hiroyuki Sueyoshi; Yu Mukai; Kotaro Yutaka; Yumie Matsui; Masahiro Yoshinaga; Masahiro Karakawa
Journal:  Cardiol J       Date:  2018-06-20       Impact factor: 2.737

2.  Coronary stent concertina in proximal left anterior descending artery: An unusual case.

Authors:  Shivanand Patil; Natraj Setty; Rangaraj Ramalingam; Jayashree Kharge; Cholenahally Nanjappa Manjunath
Journal:  Interv Med Appl Sci       Date:  2017-06
  2 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.