Stent collapse caused by balloon occlusion and aspiration system in saphenous vein graft intervention.

Morihiko Takeda1, Nobuyuki Shiba1.   

Abstract

We present the case of an 82-year-old male patient with unstable angina pectoris referred for percutaneous coronary intervention in the saphenous vein graft (SVG) to left circumflex coronary artery. An everolimus-eluting cobalt chrome stent was successfully deployed, using an embolic protection device with balloon occlusion and aspiration system. During the procedure of aspiration thrombectomy, the deployed stent unexpectedly collapsed. We re-dilated the collapsed stent using the same distal protection and aspiration system. However, the stent collapsed again during the aspiration procedure. As we could successfully re-dilate the collapsed stent using the same balloon occlusion and aspiration system after pulling the guide catheter out from the ostium of SVG, we concluded that the collapse of the stent was caused by the excess negative pressure generated by the distal occlusion and aspiration procedure under the completely wedged guide catheter into the ostium of SVG. We believe this is the first report of stent collapse due to excess negative pressure during aspiration thrombectomy. <Learning objective: Since the radial strength of coronary stents is high enough, deployed stent collapse is a rare complication. This case report describes a novel mechanism of stent collapse, which was caused by the excess negative pressure generated by the distal occlusion and aspiration procedure in saphenous vein graft intervention. To avoid this type of complication, in using distal protection devices, the side-hole guide catheter should be used, or the guide catheter should be carefully manipulated not to be wedged.>.

Entities:  

Keywords:  Aspiration; Distal protection; Stent collapse; Wedge

Year:  2017        PMID: 30279813      PMCID: PMC6149625          DOI: 10.1016/j.jccase.2017.06.002

Source DB:  PubMed          Journal:  J Cardiol Cases        ISSN: 1878-5409


  10 in total

1.  Radial force of coronary stents: a comparative analysis.

Authors:  R Rieu; P Barragan; C Masson; J Fuseri; V Garitey; M Silvestri; P Roquebert; J Sainsous
Journal:  Catheter Cardiovasc Interv       Date:  1999-03       Impact factor: 2.692

2.  Longitudinal stent deformation: a retrospective analysis of frequency and mechanisms.

Authors:  Paul D Williams; Mamas A Mamas; Kenneth P Morgan; Magdi El-Omar; Bernard Clarke; Anthony Bainbridge; Farzin Fath-Ordoubadi; Douglas G Fraser
Journal:  EuroIntervention       Date:  2012-06-20       Impact factor: 6.534

3.  Diffuse saphenous vein graft spasm.

Authors:  Vaishali Ashar; Giampaolo Niccoli; Italo Porto
Journal:  J Invasive Cardiol       Date:  2004-11       Impact factor: 2.022

Review 4.  Coronary stents: a materials perspective.

Authors:  Gopinath Mani; Marc D Feldman; Devang Patel; C Mauli Agrawal
Journal:  Biomaterials       Date:  2006-12-22       Impact factor: 12.479

5.  Stent deformity caused by coronary artery spasm.

Authors:  Toshihiko Yoshida; Yoshio Kobayashi; Takashi Nakayama; Nakabumi Kuroda; Nobuyuki Komiyama; Issei Komuro
Journal:  Circ J       Date:  2006-06       Impact factor: 2.993

6.  Evaluation of poly(L-lactic acid) as a material for intravascular polymeric stents.

Authors:  C M Agrawal; K F Haas; D A Leopold; H G Clark
Journal:  Biomaterials       Date:  1992       Impact factor: 12.479

Review 7.  Aortocoronary saphenous vein graft disease: pathogenesis, predisposition, and prevention.

Authors:  J G Motwani; E J Topol
Journal:  Circulation       Date:  1998-03-10       Impact factor: 29.690

8.  Immediate stent recoil in an anastomotic vein graft lesion treated by cutting balloon angioplasty.

Authors:  Nuri Ilker Akkus; Jagan Budeepalli; Mehmet Cilingiroglu
Journal:  Rev Port Cardiol       Date:  2013-11-14       Impact factor: 1.374

9.  Longitudinal compression of the platinum-chromium everolimus-eluting stent during coronary implantation: predisposing mechanical properties, incidence, and predictors in a large patient cohort.

Authors:  Gregor Leibundgut; Michael Gick; Aurel Toma; Christian Valina; Nikolaus Löffelhardt; Heinz Joachim Büttner; Franz-Josef Neumann
Journal:  Catheter Cardiovasc Interv       Date:  2012-11-09       Impact factor: 2.692

10.  Comparison of thin-strut cobalt-chromium stents and stainless steel stents in a porcine model of neointimal hyperplasia.

Authors:  Krzysztof Milewski; Aleksander Zurakowski; Jacek Pajak; Ewa Pajak-Zielinska; Lukasz Liszka; Piotr P Buszman; Jaroslaw Bis; Marcin Debinski; Pawel E Buszman
Journal:  Med Sci Monit       Date:  2010-01
  10 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.