Literature DB >> 22037530

The effect of anterior-posterior shear on the wear of CHARITÉ total disc replacement.

Rachel Vicars1, Polina Prokopovich, Thomas D Brown, Joanne L Tipper, Eileen Ingham, John Fisher, Richard M Hall.   

Abstract

STUDY
DESIGN: An in vitro study of the wear rates of the CHARITÉ lumbar total disc replacement (TDR).
OBJECTIVE: To investigate the effect of anterior-posterior shear on the in vitro wear rates of the CHARITÉ lumbar TDR. SUMMARY OF BACKGROUND DATA: Current standards prescribe only 4-degrees of freedom (DOF) inputs for evaluating the in vitro wear of TDRs, despite the functional spinal unit incorporating 6 DOF. Anterior-posterior shear has been highlighted as a significant load, particularly in the lumbar spine. A previous study investigated the effect of an anterior-posterior shear on the ProDisc-L, finding that wear rates were not significantly different from 4-DOF wear tests.
METHODS: Six CHARITÉ lumbar discs were mounted in a 5 active DOF spine wear simulator and tested under 4-DOF (ISO18192) conditions. Six further CHARITÉ lumbar discs were tested under 5-DOF conditions, consisting of 4-DOF conditions plus an anterior-posterior shear displacement of +2/-1.5 mm. The displacement was decreased and then increased by a factor of 2 to investigate the effect of the magnitude of displacement. Micro-computed tomographic scans of the discs were taken before and after wear testing, and the height loss of the discs was calculated. These were compared with the same measurements taken from explanted CHARITÉ discs, micro-computed tomography scanned at another institution.
RESULTS: Wear rates for 4 DOF (12.2 ± 1.0 mg/MC) were not significantly different from 4-DOF tests on the ProDisc-L. Wear rates were significantly increased (P < 0.01) for "standard" 5-DOF conditions (22.3 ± 2.0 mg/MC), decreased 5 DOF (24.3 ± 4.9 mg/MC), and increased 5 DOF (29.1 ± 7.6 mg/MC). The height loss of the explants and in vitro tested discs were not significantly different (P > 0.05).
CONCLUSION: The addition of anterior-posterior shear to wear testing inputs of the CHARITÉ lumbar TDR increases the wear rate significantly, which is in direct contrast to the previous 5-DOF testing on the ProDisc. This study highlights the importance of clinically relevant testing regimens, and that test inputs may be different for dissimilar design philosophies.

Entities:  

Mesh:

Year:  2012        PMID: 22037530      PMCID: PMC3314720          DOI: 10.1097/BRS.0b013e31823cbd6e

Source DB:  PubMed          Journal:  Spine (Phila Pa 1976)        ISSN: 0362-2436            Impact factor:   3.468


  16 in total

1.  Preferential superior surface motion in wear simulations of the Charité total disc replacement.

Authors:  Curtis M Goreham-Voss; Rachel Vicars; Richard M Hall; Thomas D Brown
Journal:  Eur Spine J       Date:  2010-06-26       Impact factor: 3.134

Review 2.  Adjacent segment disease after lumbar or lumbosacral fusion: review of the literature.

Authors:  Paul Park; Hugh J Garton; Vishal C Gala; Julian T Hoff; John E McGillicuddy
Journal:  Spine (Phila Pa 1976)       Date:  2004-09-01       Impact factor: 3.468

3.  The accuracy and precision of a micro computer tomography volumetric measurement technique for the analysis of in-vitro tested total disc replacements.

Authors:  R Vicars; J Fisher; R M Hall
Journal:  Proc Inst Mech Eng H       Date:  2009-04       Impact factor: 1.617

4.  Stereo radiography of lumbar spine motion.

Authors:  M J Pearcy
Journal:  Acta Orthop Scand Suppl       Date:  1985

5.  Polyethylene wear and rim fracture in total disc arthroplasty.

Authors:  Steven M Kurtz; André van Ooij; Raymond Ross; Jan de Waal Malefijt; John Peloza; Lauren Ciccarelli; Marta L Villarraga
Journal:  Spine J       Date:  2006-12-06       Impact factor: 4.166

6.  Fatigue strength of polyethylene after sterilization by gamma irradiation or ethylene oxide.

Authors:  M D Ries; K Weaver; R M Rose; J Gunther; W Sauer; N Beals
Journal:  Clin Orthop Relat Res       Date:  1996-12       Impact factor: 4.176

Review 7.  Lumbar disc disorders and low-back pain: socioeconomic factors and consequences.

Authors:  Jeffrey N Katz
Journal:  J Bone Joint Surg Am       Date:  2006-04       Impact factor: 5.284

8.  The effect of anterior-posterior shear load on the wear of ProDisc-L TDR.

Authors:  R Vicars; P J Hyde; T D Brown; J L Tipper; E Ingham; J Fisher; R M Hall
Journal:  Eur Spine J       Date:  2010-04-18       Impact factor: 3.134

9.  What is the correlation of in vivo wear and damage patterns with in vitro TDR motion response?

Authors:  Steven M Kurtz; Avinash Patwardhan; Daniel MacDonald; Lauren Ciccarelli; André van Ooij; Mark Lorenz; Michael Zindrick; Patrick O'Leary; Jorge Isaza; Raymond Ross
Journal:  Spine (Phila Pa 1976)       Date:  2008-03-01       Impact factor: 3.468

Review 10.  Why a mechanical disc?

Authors:  Thomas J Errico
Journal:  Spine J       Date:  2004 Nov-Dec       Impact factor: 4.166

View more
  6 in total

1.  Biomechanical Comparison of Robotically Applied Pure Moment, Ideal Follower Load, and Novel Trunk Weight Loading Protocols on L4-L5 Cadaveric Segments during Flexion-Extension.

Authors:  Charles R Bennett; Denis J DiAngelo; Brian P Kelly
Journal:  Int J Spine Surg       Date:  2015-07-17

2.  We Need to Talk about Lumbar Total Disc Replacement.

Authors:  Stephen Beatty
Journal:  Int J Spine Surg       Date:  2018-08-03

3.  Wear of the Charité® lumbar intervertebral disc replacement investigated using an electro-mechanical spine simulator.

Authors:  Parshia Moghadas; Aziza Mahomed; Duncan E T Shepherd; David W L Hukins
Journal:  Proc Inst Mech Eng H       Date:  2015-03       Impact factor: 1.617

4.  Total disc arthroplasty for treating lumbar degenerative disc disease.

Authors:  Keyvan Mostofi
Journal:  Asian Spine J       Date:  2015-02-13

5.  ISASS Policy Statement - Lumbar Artificial Disc.

Authors:  Jack Zigler; Rolando Garcia
Journal:  Int J Spine Surg       Date:  2015-03-12

6.  Wear characteristics of an unconstrained lumbar total disc replacement under a range of in vitro test conditions.

Authors:  Philip J Hyde; John Fisher; Richard M Hall
Journal:  J Biomed Mater Res B Appl Biomater       Date:  2015-09-28       Impact factor: 3.368

  6 in total