Clinical accuracy of ExacTrac intracranial frameless stereotactic system.

T Ackerly1, C M Lancaster, M Geso, K J Roxby.   

Abstract

PURPOSE: In this paper, the authors assess the accuracy of the Brainlab ExacTrac system for frameless intracranial stereotactic treatments in clinical practice.
METHODS: They recorded couch angle and image fusion results (comprising lateral, longitudinal, and vertical shifts, and rotation corrections about these axes) for 109 stereotactic radiosurgery and 166 stereotactic radiotherapy patient treatments. Frameless stereotactic treatments involve iterative 6D image fusion corrections applied until the results conform to customizable pass criteria, theirs being 0.7 mm and 0.5° for each axis. The planning CT slice thickness was 1.25 mm. It has been reported in the literature that the CT slices' thickness impacts the accuracy of localization to bony anatomy. The principle of invariance with respect to patient orientation was used to determine spatial accuracy.
RESULTS: The data for radiosurgery comprised 927 image pairs, of which 532 passed (pass ratio of 57.4%). The data for radiotherapy comprised 15983 image pairs, of which 10 050 passed (pass ratio of 62.9%). For stereotactic radiotherapy, the combined uncertainty of ExacTrac calibration, image fusion, and intrafraction motion was (95% confidence interval) 0.290-0.302 and 0.306-0.319 mm in the longitudinal and lateral axes, respectively. The combined uncertainty of image fusion and intrafraction motion in the anterior-posterior coordinates was 0.174-0.182 mm. For stereotactic radiosurgery, the equivalent ranges are 0.323-0.393, 0.337-0.409, and 0.231-0.281 mm. The overall spatial accuracy was 1.24 mm for stereotactic radiotherapy (SRT) and 1.35 mm for stereotactic radiosurgery (SRS).
CONCLUSIONS: The ExacTrac intracranial frameless stereotactic system spatial accuracy is adequate for clinical practice, and with the same pass criteria, SRT is more accurate than SRS. They now use frameless stereotaxy exclusively at their center.

Entities:  

Mesh:

Year:  2011        PMID: 21978048     DOI: 10.1118/1.3611044

Source DB:  PubMed          Journal:  Med Phys        ISSN: 0094-2405            Impact factor:   4.071


  15 in total

1.  An evaluation of the consistency of shifts reported by three different systems for non-coplanar treatments.

Authors:  Vikren Sarkar; Adam Paxton; Martin W Szegedi; Hui Zhao; Long Huang; Geoff Nelson; Yu-Huei Jessica Huang; Fanchi Su; Prema Rassiah-Szegedi; Bill J Salter
Journal:  J Radiosurg SBRT       Date:  2018

2.  Accuracy of positional correction for the floor-mounted kV X-ray IGRT system in angled couch positions.

Authors:  Satoshi Koubuchi; Toru Takakura; Mitsuhiro Nakamura; Takashi Mizowaki; Manabu Nakata; Masahiro Hiraoka
Journal:  Radiol Phys Technol       Date:  2014-06-13

3.  Single fraction volumetric modulated arc radiosurgery of brain metastases.

Authors:  A Serna; P P Escolar; V Puchades; F Mata; D Ramos; M A Gómez; A Iglesias; J Salinas; M Alcaraz
Journal:  Clin Transl Oncol       Date:  2015-03-17       Impact factor: 3.405

4.  Clinical Assessment of 2D/3D Registration Accuracy in 4 Major Anatomic Sites Using On-Board 2D Kilovoltage Images for 6D Patient Setup.

Authors:  Guang Li; T Jonathan Yang; Hugo Furtado; Wolfgang Birkfellner; Åse Ballangrud; Simon N Powell; James Mechalakos
Journal:  Technol Cancer Res Treat       Date:  2014-09-15

5.  Real-time image-processing algorithm for markerless tumour tracking using X-ray fluoroscopic imaging.

Authors:  S Mori
Journal:  Br J Radiol       Date:  2014-03-25       Impact factor: 3.039

6.  Spatial variations of multiple off-axial targets for a single isocenter SRS treatment in Novalis Tx linac system.

Authors:  Sangroh Kim; Tzu-Chi Tseng; Andrew Morrow
Journal:  J Radiosurg SBRT       Date:  2015

7.  Targeting accuracy at couch kick for a frameless image guided radiosurgery system.

Authors:  Yimei Huang; Bo Zhao; Joshua Kim; Ning Wen; Indrin J Chetty; Salim Siddiqui
Journal:  J Radiosurg SBRT       Date:  2018

8.  Heterogeneity correction for intensity-modulated frameless SRS in pituitary and cavernous sinus tumors: a retrospective study.

Authors:  Lisa B E Shields; Cindy Bond; Aaron Odom; David A Sun; Aaron C Spalding
Journal:  Radiat Oncol       Date:  2015-09-17       Impact factor: 3.481

9.  Dosimetric effects of positioning shifts using 6D-frameless stereotactic Brainlab system in hypofractionated intracranial radiotherapy.

Authors:  Hosang Jin; Vance P Keeling; Imad Ali; Salahuddin Ahmad
Journal:  J Appl Clin Med Phys       Date:  2016-01-08       Impact factor: 2.102

10.  Clinical experience with two frameless stereotactic radiosurgery (fSRS) systems using optical surface imaging for motion monitoring.

Authors:  Guang Li; Ase Ballangrud; Maria Chan; Ruimei Ma; Kathryn Beal; Yoshiya Yamada; Timothy Chan; James Lee; Pretti Parhar; James Mechalakos; Margie Hunt
Journal:  J Appl Clin Med Phys       Date:  2015-07-08       Impact factor: 2.102

View more

北京卡尤迪生物科技股份有限公司 © 2022-2023.