Comparison of anti-scatter grids for digital imaging with use of a direct-conversion flat-panel detector.

Masayoshi Mizuta1, Shigeru Sanada, Hiroyuki Akazawa, Toshifumi Kasai, Shuji Abe, Yasuhiro Ikeno, Shigeki Mitou.   

Abstract

Our purpose in this study was to establish a selection standard for anti-scatter grids for a direct conversion flat-panel detector (FPD) system. As indices for grid evaluation, we calculated the selectivity, Bucky factor, and the signal-to-noise ratio improvement factor (SIF) by measuring rates of scatter transmission, primary transmission, and total transmission (based on the digitally displayed measurement values of the FPD system), using 4 acrylic phantoms of different thicknesses. The results showed that the SIF was less than 1.0 when the phantom thickness was 5 cm. When the phantom thickness was 25 cm and the grid ratio was 16:1, the SIF was 1.505 and 1.518 (maximum value) at 90 and 120 kV, respectively. Compared with the grid ratio of 12:1, the SIF at the grid ratio 16:1 was improved by 6.1% at 90 kV, and by 7.0% at 120 kV. In a direct-conversion FPD system, the grid ratio of 16:1 is considered adequate for eliminating the scattered-radiation effect when much scattered radiation is present, such as with a thick imaged object or a high X-ray tube voltage.

Mesh:

Year:  2011        PMID: 21971637     DOI: 10.1007/s12194-011-0134-1

Source DB:  PubMed          Journal:  Radiol Phys Technol        ISSN: 1865-0333


  13 in total

1.  A study and optimization of lumbar spine X-ray imaging systems.

Authors:  G McVey; M Sandborg; D R Dance; G Alm Carlsson
Journal:  Br J Radiol       Date:  2003-03       Impact factor: 3.039

2.  Studies of performance of antiscatter grids in digital radiography: effect on signal-to-noise ratio.

Authors:  H P Chan; K L Lam; Y Z Wu
Journal:  Med Phys       Date:  1990 Jul-Aug       Impact factor: 4.071

3.  Experimental evaluation of fiber-interspaced antiscatter grids for large patient imaging with digital x-ray systems.

Authors:  Kenneth A Fetterly; Beth A Schueler
Journal:  Phys Med Biol       Date:  2007-07-30       Impact factor: 3.609

4.  Technical note: a comparison of antiscatter grids for digital radiography.

Authors:  L Court; T Yamazaki
Journal:  Br J Radiol       Date:  2004-11       Impact factor: 3.039

5.  Monte Carlo study of grid performance in diagnostic radiology: task dependent optimization for screen-film imaging.

Authors:  M Sandborg; D R Dance; G A Carlsson; J Persliden
Journal:  Br J Radiol       Date:  1994-01       Impact factor: 3.039

6.  Monte Carlo study of grid performance in diagnostic radiology: factors which affect the selection of tube potential and grid ratio.

Authors:  M Sandborg; D R Dance; G A Carlsson; J Persliden
Journal:  Br J Radiol       Date:  1993-12       Impact factor: 3.039

7.  Selection of anti-scatter grids for different imaging tasks: the advantage of low atomic number cover and interspace materials.

Authors:  M Sandborg; D R Dance; G A Carlsson; J Persliden
Journal:  Br J Radiol       Date:  1993-12       Impact factor: 3.039

8.  Image quality of a digital chest radiography system based on a selenium detector.

Authors:  U Neitzel; I Maack; S Günther-Kohfahl
Journal:  Med Phys       Date:  1994-04       Impact factor: 4.071

9.  The validity of Monte Carlo simulation in studies of scattered radiation in diagnostic radiology.

Authors:  H P Chan; K Doi
Journal:  Phys Med Biol       Date:  1983-02       Impact factor: 3.609

10.  Performance of antiscatter grids in diagnostic radiology: experimental measurements and Monte Carlo simulation studies.

Authors:  H P Chan; Y Higashida; K Doi
Journal:  Med Phys       Date:  1985 Jul-Aug       Impact factor: 4.071

View more
  1 in total

1.  Investigation of grid performance using simple image quality tests.

Authors:  Dogan Bor; Ozlem Birgul; Umran Onal; Turan Olgar
Journal:  J Med Phys       Date:  2016 Jan-Mar
  1 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.