Monte Carlo study of grid performance in diagnostic radiology: factors which affect the selection of tube potential and grid ratio.

M Sandborg1, D R Dance, G A Carlsson, J Persliden.   

Abstract

A Monte Carlo computational model has been developed for the study of the performance of anti-scatter grids in diagnostic radiology. It is used here to estimate the scatter in the image plane from soft tissue phantoms (representing the patient) and to calculate image contrast and the mean absorbed dose in the phantom. Different scattering conditions, representative of various examinations, have been investigated: adult lumbar spine; small field radiography and fluoroscopy; adult chest and paediatric pelvis and chest. For each scattering condition, the combinations of tube potential and grid ratio have been found which, for a well designed grid, result in the lowest mean absorbed dose in the phantom for a fixed contrast level. In examinations which generate large amounts of scatter, the use of high grid ratios in combination with high tube potentials is favourable with regard to both mean absorbed dose in the phantom and tube charge. When less scatter is generated, either the grid ratio or the tube potential can be varied to achieve the desired contrast level. High grid ratios require shorter exposure times, but need careful alignment in the beam to prevent primary radiation cut-off. It is shown that the air gap technique can be used to reduce patient dose in examinations with small amounts of scatter, but in combinations with a lower tube potential than when a grid is used.

Entities:  

Mesh:

Year:  1993        PMID: 8293262     DOI: 10.1259/0007-1285-66-792-1164

Source DB:  PubMed          Journal:  Br J Radiol        ISSN: 0007-1285            Impact factor:   3.039


  4 in total

1.  Comparison of anti-scatter grids for digital imaging with use of a direct-conversion flat-panel detector.

Authors:  Masayoshi Mizuta; Shigeru Sanada; Hiroyuki Akazawa; Toshifumi Kasai; Shuji Abe; Yasuhiro Ikeno; Shigeki Mitou
Journal:  Radiol Phys Technol       Date:  2011-10-05

2.  On image quality metrics and the usefulness of grids in digital mammography.

Authors:  Han Chen; Mats Danielsson; Cheng Xu; Björn Cederström
Journal:  J Med Imaging (Bellingham)       Date:  2015-02-17

3.  Pediatric interventional radiography equipment: safety considerations.

Authors:  Keith J Strauss
Journal:  Pediatr Radiol       Date:  2006-09

4.  Optimisation in general radiography.

Authors:  Cj Martin
Journal:  Biomed Imaging Interv J       Date:  2007-04-01
  4 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.