| Literature DB >> 21961830 |
Alessandro Clivio1, Giorgia Nicolini, Eugenio Vanetti, Antonella Fogliata, Luca Cozzi.
Abstract
BACKGROUND: A new-generation low-energy linear accelerator (UNIQUE) was introduced in the clinical arena during 2009 by Varian Medical Systems. The world's first UNIQUE was installed at Oncology Institute of Southern Switzerland and put into clinical operation in June 2010. The aim of the present contribution was to report experience about its commissioning and first year results from clinical operation.Entities:
Mesh:
Year: 2011 PMID: 21961830 PMCID: PMC3197498 DOI: 10.1186/1748-717X-6-129
Source DB: PubMed Journal: Radiat Oncol ISSN: 1748-717X Impact factor: 3.481
Figure 1Standard star-shot test. for radiation isocenter determination with film: measured radius for gantry radiation isocenter sphere resulted <0.1 mm
Figure 2Dose accuracy for open fields as function of field size. (10 cm depth, SDD = 90 cm): percentage difference of calculated respect to measured dose, for 100 MU.
Figure 3Stability results over one year period for UNIQUE and Clinac iX. Error bars refer to one standard deviation. A) Output (weekly check): percentage dose deviation from reference. At week n.42, machine output was re-tuned for both machines according to institutional protocols. B) MU stability (monthly check): ion chamber reading ratio for a fixed number of MU (100) between delivery at a fixed dose rate (100, 200, 300, 400, 500, 600 MU/min) and reference reading at 300 MU/min. C) MU linearity (monthly check): ratio between ion chamber reading at fixed number of MU (5, 10, 50, 100, 200, 300 MU) with respect to the same for reference of 100 MU. Measurements are relative to a fixed dose rate of 300 MU/min.
Summary of the results of the periodic radiation beam quality assurance measurements.
| Unique | Clinac iX | |||
|---|---|---|---|---|
| -0.3 ± 0.2% [-0.8,+0.2] | -0.1 ± 0.4% [-0.9,+1.0] | |||
| %diff. ratio @5.6 cm/dmax | 0.0 ± 0.0 [-0.1,0.1] | 0.1 ± 0.1 [-0.2,0.6] | ||
| %diff. ratio @7.6 cm/dmax | -0.1 ± 0.1 [-0.3, 0.1] | -0.1 ± 0.1 [-0.3, 0.4] | ||
| %diff. ratio @11 cm/dmax | -0.1 ± 0.1 [-0.3, 0.1] | -0.0 ± 0.2 [-0.3, 0.6] | ||
| 0.0 ± 0.2 [-0.1,0.1] | -0.0 ± 0.3 [-0.1,0.5] | |||
| 10.03 ± 0.05(ref.10.02) [10.00, 10.14] | 10.06 ± 0.06(ref.10.07) [9.99, 10.13] | 10.10 ± 0.03(ref.10.02) [10.08, 10.11] | 9.99 ± 0.02(ref.10.04) [9.94, 10.05] | |
| 20.13 ± 0.01(ref.20.13) [20.10, 20.14] | 20.14 ± 0.02(ref.20.13) [20.11, 20.18] | 20.21 ± 0.01(ref.20.21) [20.18, 20.24] | 20.09 ± 0.02(ref.20.02) [20.02, 20.10] | |
| 0.8 ± 0.04 (ref.0.7) [0.7, 0.9] | 0.9 ± 0.04 (ref.0.9) [0.8, 0.9] | 1.2 ± 0.09 (ref.1.1) [1.0, 1.5] | 0.8 ± 0.05 (ref.0.9) [0.8,1.0] | |
| 1.5 ± 0.04 (ref.1.5) [1.4, 1.6] | 2.0 ± 0.06 (ref.1.8) [1.9, 2.1] | 1.1 ± 0.12 (ref.1.0) [0.9, 1.4] | 1.6 ± 0.10ref.1.7) [1.4,1.8] | |
| 100.6 ± 0.2 (ref.100.6) [100.4, 100.9] | 100.3 ± 0.2(ref.100.4) [100.1, 100.7] | 100.5 ± 0.2 (ref.100.3) [100.3, 101.1] | 100.5 ± 0.2 (ref.100.3) [100.2, 101.1] | |
| 101.1 ± 0.1(ref.101.3) [101.0, 101.2] | 100.4 ± 0.2 (ref.100.3) [100.2, 100.7] | 100.6 ± 0.2 (ref.100.4) [100.1, 101.1] | 101.4 ± 0.3 (ref.101.7) [100.5, 101.8] | |
Measurements were performed by means of the GLAaS method (except output, where a 0.6 cm3 ion chamber was used) on the UNIQUE linac and, as comparison, on the Clinac iX linac operational at authors institute. Results derived from weekly controls over a period of one year and included stability of: dose output, beam quality, EDW wedge factor, field size, beam homogeneity and symmetry. Measurement settings are reported in the first column. Mean results were shown together with their standard deviation and range (in square brackets) for UNIQUE and Clinac iX in the second and third columns. Reference baseline was reported within normal brackets
Figure 4Measured and calculated open fields. (10 cm depth, SDD = 90 cm); calculated data refer to AAA algorithm version 10.0.25. First row: examples of profiles and DD curves; second row: gamma analysis [1%,1 mm] on all data after beam processing phase.
Figure 5Wedges results : profiles and dose accuracy as function of field size, respectively A) Hard Wedges and B) EDW. Dose accuracy is defined as: percentage difference of calculated dose respect measured dose, with fixed MU, at 10 cm depth, SDD = 90 cm.
Figure 6Image guidance on UNIQUE. Verification of 2D-2D image matching with weekly quality assurance procedure: average results (and standard deviation) over one year period. For comparison results for the same procedure on Clinac iX are presented although these latter refer to images acquired with the MV and kV systems, for steps 1 and 3 of the protocol.
Figure 7Comprehensive test of the stability of the imaging center. (including PV mechanical stability and gantry sag motion) from monthly quality assurance procedures. Test was performed acquiring portal images in continuous (cine) mode with a full gantry rotation. The graph reported an example of the actual distance of the center C of the small radiation field from the physical image center in the X and Y directions as a function of the gantry angle; the histogram showed the average deviations observed over a test period of one year as well as the maximum distance, the maximum deviation in X and in Y directions and also the maximum excursion of the deviations; error bars are expressed as 1 standard deviation.
Figure 8Summary of the RapidArc commissioning tests. (according to Ling et al [13]) performed as monthly checks over a period of one year. The plot showed the dose output difference between readings in each uniform band from the average values for: test 0.1 (a dynamic IMRT field with a 0.5 cm slit at 4 different gantry positions); test 2 (seven different combinations of dose rate and gantry speed during a RapidArc delivery) and test 3 (four different combinations of MLC speed and dose rate during a RapidArc delivery). Recommended tolerance was ±2% for all tests. Images for tests 2 and 3 are corrected for the beam profile, rationing the band and the open field acquisitions.
Figure 9Summary of the Dynalog Files analysis. for 12 test cases from real clinical patients delivered on both UNIQUE and Clinac iX linacs. Plot showed the average deviations from reference or expected values during arc delivery of: gantry angle, accumulated MU and RMS of MLC positions.
Summary of the stability control and of the pre-treatment patients quality assurance results for RapidArc and IMRT treatments.
| Unique | Clinac iX | ||
|---|---|---|---|
| GAI [%] constancy on a pre-treatment QA case (1 year data with a periodicity of 2 weeks) | RapidArc case | 98.5 ± 1.1 [96.7, 99.6] | 99.0 ± 0.3 [98.3, 99.4] |
| IMRT case | 99.4 ± 0.1 [99.2, 99.2] | 99.0 ± 0.4 [98.2, 99.3] | |
| Clinical pre-treatment RapidArc QA | GAI [%] | 97.3 ± 1.6 [92.4, 99.9] | 97.4 ± 1.8 [91.5, 99.9] |
| Number of arcs (plans) | 348 (192) [12 months] | 1186 (797) [31 months] |
Beside UNIQUE summary, a comparison with similar results from the Clinac iX operational at the institute is shown. Data are expressed as Gamma Agreement Index GAI with Distance to Agreement and Dose Difference thresholds set to 3 mm and 3%. Measurements and calculations were performed according to the GLAaS method by means of portal dosimetry.