Motion monitoring for cranial frameless stereotactic radiosurgery using video-based three-dimensional optical surface imaging.

Guang Li1, Ase Ballangrud, Li Cheng Kuo, Hyejoo Kang, Assen Kirov, Michael Lovelock, Yoshiya Yamada, James Mechalakos, Howard Amols.   

Abstract

PURPOSE: To establish a new clinical procedure in frameless stereotactic radiosurgery (SRS) for patient setup verification at treatment couch angles as well as for head-motion monitoring during treatment using video-based optical surface imaging (OSI).
METHODS: A video-based three-dimensional (3D) OSI system with three ceiling-mounted camera pods was employed to verify setup at treatment couch angles as well as to monitor head motion during treatment. A noninvasive head immobilization device was utilized, which includes an alpha head mold and a dental mouthpiece with vacuum suction; both were locked to the treatment couch. Cone beam computed tomography (CBCT) was used as the standard for image-guided setup. Orthogonal 2D-kV imaging was applied for setup verification before treatment, between couch rotations, and after treatment at zero couch angle. At various treatment couch angles, OSI setup verification was performed, relative to initial OSI setup verification at zero couch angle after CBCT setup through a coordinate transformation. For motion monitoring, the setup uncertainty was decoupled by taking an on-site surface image as new reference to detect motion-induced misalignment in near real-time (1-2 frames per second). Initial thermal instability baseline of the real-time monitoring was corrected. An anthropomorphous head phantom and a 1D positioning platform were used to assess the OSI accuracy in motion detection in longitudinal and lateral directions. Two hypofractionated (9 Gy x 3 and 6 Gy x 5) frameless stereotactic radiotherapy (SRT) patients as well as two single-fraction (21 and 18 Gy) frameless SRS patients were treated using this frameless procedure. For comparison, 11 conventional frame-based SRS patients were monitored using the OSI to serve as clinical standards. Multiple noncoplanar conformal beams were used for planning both frameless and frame-based SRS with a micromultileaf collimator.
RESULTS: The accuracy of the OSI in 1D motion detection was found to be 0.1 mm with uncertainty of +/- 0.1 mm using the head phantom. The OSI registration against simulation computed tomography (CT) external contour was found to be dependent on the CT skin definition with -0.4 mm variation. For frame-based SRS patients, head-motion magnitude was detected to be <1.0 mm (0.3 +/- 0.2 mm) and <1.0 degree (0.2 degrees +/- 0.2 degrees) for 98% of treatment time, with exception of one patient with head rotation <1.5 degrees for 98% of the time. For frameless SRT/SRS patients, similar motion magnitudes were observed with an average of 0.3 +/- 0.2 mm and 0.2 degrees +/- 0.1 degree in ten treatments. For 98% of the time, the motion magnitude was <1.1 mm and 1.0 degree. Complex head-motion patterns within 1.0 mm were observed for frameless SRT/SRS patients. The OSI setup verification at treatment couch angles was found to be within 1.0 mm.
CONCLUSIONS: The OSI system is capable of detecting 0.1 +/- 0.1 mm 1D spatial displacement of a phantom in near real time and useful in head-motion monitoring. This new frameless SRS procedure using the mask-less head-fixation system provides immobilization similar to that of conventional frame-based SRS. Head-motion monitoring using near-real-time surface imaging provides adequate accuracy and is necessary for frameless SRS in case of unexpected head motion that exceeds a set tolerance.

Entities:  

Mesh:

Year:  2011        PMID: 21858995     DOI: 10.1118/1.3596526

Source DB:  PubMed          Journal:  Med Phys        ISSN: 0094-2405            Impact factor:   4.071


  38 in total

1.  Novel spirometry based on optical surface imaging.

Authors:  Guang Li; Hailiang Huang; Jie Wei; Diana G Li; Qing Chen; Carl P Gaebler; James Sullivan; Joan Zatcky; Andreas Rimner; James Mechalakos
Journal:  Med Phys       Date:  2015-04       Impact factor: 4.071

2.  An evaluation of the consistency of shifts reported by three different systems for non-coplanar treatments.

Authors:  Vikren Sarkar; Adam Paxton; Martin W Szegedi; Hui Zhao; Long Huang; Geoff Nelson; Yu-Huei Jessica Huang; Fanchi Su; Prema Rassiah-Szegedi; Bill J Salter
Journal:  J Radiosurg SBRT       Date:  2018

Review 3.  Image guidance in proton therapy for lung cancer.

Authors:  Miao Zhang; Wei Zou; Boon-Keng Kevin Teo
Journal:  Transl Lung Cancer Res       Date:  2018-04

4.  Towards frameless maskless SRS through real-time 6DoF robotic motion compensation.

Authors:  Andrew H Belcher; Xinmin Liu; Steven Chmura; Kamil Yenice; Rodney D Wiersma
Journal:  Phys Med Biol       Date:  2017-11-13       Impact factor: 3.609

5.  Accuracy of positional correction for the floor-mounted kV X-ray IGRT system in angled couch positions.

Authors:  Satoshi Koubuchi; Toru Takakura; Mitsuhiro Nakamura; Takashi Mizowaki; Manabu Nakata; Masahiro Hiraoka
Journal:  Radiol Phys Technol       Date:  2014-06-13

6.  Spatial and rotational quality assurance of 6DOF patient tracking systems.

Authors:  Andrew H Belcher; Xinmin Liu; Zachary Grelewicz; Rodney D Wiersma
Journal:  Med Phys       Date:  2016-06       Impact factor: 4.071

7.  Feasibility of differential geometry-based features in detection of anatomical feature points on patient surfaces in range image-guided radiation therapy.

Authors:  Mazen Soufi; Hidetaka Arimura; Katsumasa Nakamura; Fauzia P Lestari; Freddy Haryanto; Taka-Aki Hirose; Yoshiyuki Umedu; Yoshiyuki Shioyama; Fukai Toyofuku
Journal:  Int J Comput Assist Radiol Surg       Date:  2016-06-13       Impact factor: 2.924

8.  Fast intra-fractional image-guidance with 6D positioning correction reduces delivery uncertainty for stereotactic radiosurgery and radiotherapy.

Authors:  Nikolai Tarnavski; Svend Aage Engelholm; Per Munck Af Rosenschold
Journal:  J Radiosurg SBRT       Date:  2016

9.  Robust fluoroscopic tracking of fiducial markers: exploiting the spatial constraints.

Authors:  Rui Li; Gregory Sharp
Journal:  Phys Med Biol       Date:  2013-02-26       Impact factor: 3.609

10.  Characterization of optical-surface-imaging-based spirometry for respiratory surrogating in radiotherapy.

Authors:  Guang Li; Jie Wei; Hailiang Huang; Qing Chen; Carl P Gaebler; Tiffany Lin; Amy Yuan; Andreas Rimner; James Mechalakos
Journal:  Med Phys       Date:  2016-03       Impact factor: 4.071

View more

北京卡尤迪生物科技股份有限公司 © 2022-2023.