Analytic image concept combined to SENSE reconstruction.

Josiane Yankam Njiwa1, Christof Baltes, Markus Rudin.   

Abstract

OBJECT: Two approaches of reconstructing undersampled partial k-space data, acquired with multiple coils are compared: homodyne detection combined with SENSE (HM_SENSE) and analytic image reconstruction combined with SENSE (AI_SENSE). The latter overcomes limitations of HM_ SENSE by considering aliased images as analytic thus avoiding the need for phase correction required for HM_SENSE.
MATERIALS AND METHODS: In vivo imaging experiments were carried out in male Lewis rats using both gradient echo and spin echo sequences. Accelerated images obtained by using the various reconstruction algorithms were compared to fully sampled reference images both qualitatively and quantitatively.
RESULTS: For the various sampling patterns evaluated, both HM_SENSE and AI_SENSE were found to yield robust image reconstruction with small deviations from the reference image. Even for high acceleration factors AI_SENSE still provided useful results and was found superior compared to HM_SENSE.
CONCLUSION: Combination of partial k-space sampling and parallel image acquisition allows for further acceleration of data acquisition as compared to each method alone. Image reconstruction from undersampled data sets using the AI_SENSE algorithm was found to considerably reduce reconstruction errors and artifacts observed for HM_SENSE reconstruction caused by errors in phase estimation.

Entities:  

Mesh:

Year:  2011        PMID: 21833790     DOI: 10.1007/s10334-011-0274-7

Source DB:  PubMed          Journal:  MAGMA        ISSN: 0968-5243            Impact factor:   2.310


  17 in total

1.  SENSE: sensitivity encoding for fast MRI.

Authors:  K P Pruessmann; M Weiger; M B Scheidegger; P Boesiger
Journal:  Magn Reson Med       Date:  1999-11       Impact factor: 4.668

2.  VD-AUTO-SMASH imaging.

Authors:  R M Heidemann; M A Griswold; A Haase; P M Jakob
Journal:  Magn Reson Med       Date:  2001-06       Impact factor: 4.668

3.  Generalized autocalibrating partially parallel acquisitions (GRAPPA).

Authors:  Mark A Griswold; Peter M Jakob; Robin M Heidemann; Mathias Nittka; Vladimir Jellus; Jianmin Wang; Berthold Kiefer; Axel Haase
Journal:  Magn Reson Med       Date:  2002-06       Impact factor: 4.668

4.  Using UNFOLD to remove artifacts in parallel imaging and in partial-Fourier imaging.

Authors:  Bruno Madore
Journal:  Magn Reson Med       Date:  2002-09       Impact factor: 4.668

5.  UNFOLD-SENSE: a parallel MRI method with self-calibration and artifact suppression.

Authors:  Bruno Madore
Journal:  Magn Reson Med       Date:  2004-08       Impact factor: 4.668

6.  The NMR phased array.

Authors:  P B Roemer; W A Edelstein; C E Hayes; S P Souza; O M Mueller
Journal:  Magn Reson Med       Date:  1990-11       Impact factor: 4.668

7.  POCSENSE: POCS-based reconstruction for sensitivity encoded magnetic resonance imaging.

Authors:  Alexei A Samsonov; Eugene G Kholmovski; Dennis L Parker; Chris R Johnson
Journal:  Magn Reson Med       Date:  2004-12       Impact factor: 4.668

8.  Homodyne detection in magnetic resonance imaging.

Authors:  D C Noll; D G Nishimura; A Macovski
Journal:  IEEE Trans Med Imaging       Date:  1991       Impact factor: 10.048

9.  Cardiac cine MR reconstruction from partial k-space using the notion of analytic image.

Authors:  Josiane Yankam Njiwa; Bassem Hiba; Yuemin Zhu
Journal:  Annu Int Conf IEEE Eng Med Biol Soc       Date:  2007

10.  Simultaneous acquisition of spatial harmonics (SMASH): fast imaging with radiofrequency coil arrays.

Authors:  D K Sodickson; W J Manning
Journal:  Magn Reson Med       Date:  1997-10       Impact factor: 4.668

View more

北京卡尤迪生物科技股份有限公司 © 2022-2023.