Literature DB >> 21804747

Study of Stent Deployment Mechanics Using a High-Resolution X-ray Imaging Detector.

Weiyuan Wang1, Ciprian N Ionita, Daniel R Bednarek, Stephen Rudin.   

Abstract

To treat or prevent some of the 795,000 annual strokes in the U.S., self-expanding endo-vascular stents deployed under fluoroscopic image guidance are often used. Neuro-interventionalists need to know the deployment behavior of each stent in order to place them in the correct position. Using the Micro-Angiographic Fluoroscope (MAF) which has about 3 times higher resolution than commercially available flat panel detectors (FPD) we studied the deployment mechanics of two of the most important commercially available nitinol stents: the Pipeline embolization device (EV3), and the Enterprise stent (Codman). The Pipeline stent's length extends to about 3 times that of its deployed length when it is contained inside a catheter. From the high-resolution images with the MAF we found that upon the sudden release of the distal end of the Pipeline from a helical wire cap, the stent expands radially but retracts to about 30% (larger than for patient deployments) of its length. When released from the catheter proximally, it retracts additionally about 50% contributing to large uncertainty in the final deployed location. In contrast, the MAF images clearly show that the Enterprise stent self expands with minimal length retraction during deployment from its catheter and can be retrieved and repositioned until the proximal markers are released from clasping structures on its guide-wire thus enabling more accurate placement at the center of an aneurysm or stenosis. The high-resolution imaging demonstrated in this study should help neurointerventionalists understand and control endovascular stent deployment mechanisms and hence perform more precise treatments.

Entities:  

Year:  2011        PMID: 21804747      PMCID: PMC3144509          DOI: 10.1117/12.877710

Source DB:  PubMed          Journal:  Proc SPIE Int Soc Opt Eng        ISSN: 0277-786X


  7 in total

1.  Infectious aneurysm of the cavernous carotid artery in a child treated with a new-generation of flow-diverting stent graft: case report.

Authors:  Geoffrey Appelboom; Khaled Kadri; Farouk Hassan; Xavier Leclerc
Journal:  Neurosurgery       Date:  2010-03       Impact factor: 4.654

2.  Implementation of a high-sensitivity Micro-Angiographic Fluoroscope (HS-MAF) for in-vivo endovascular image guided interventions (EIGI) and region-of-interest computed tomography (ROI-CT).

Authors:  C N Ionita; C Keleshis; V Patel; G Yadava; K R Hoffmann; D R Bednarek; A Jain; S Rudin
Journal:  Proc SPIE Int Soc Opt Eng       Date:  2008

3.  Heart disease and stroke statistics--2010 update: a report from the American Heart Association.

Authors:  Donald Lloyd-Jones; Robert J Adams; Todd M Brown; Mercedes Carnethon; Shifan Dai; Giovanni De Simone; T Bruce Ferguson; Earl Ford; Karen Furie; Cathleen Gillespie; Alan Go; Kurt Greenlund; Nancy Haase; Susan Hailpern; P Michael Ho; Virginia Howard; Brett Kissela; Steven Kittner; Daniel Lackland; Lynda Lisabeth; Ariane Marelli; Mary M McDermott; James Meigs; Dariush Mozaffarian; Michael Mussolino; Graham Nichol; Véronique L Roger; Wayne Rosamond; Ralph Sacco; Paul Sorlie; Véronique L Roger; Randall Stafford; Thomas Thom; Sylvia Wasserthiel-Smoller; Nathan D Wong; Judith Wylie-Rosett
Journal:  Circulation       Date:  2009-12-17       Impact factor: 29.690

4.  Navigability trumps all: stenting of acute middle cerebral artery occlusions with a new self-expandable stent.

Authors:  P T L Chiam; R M Samuelson; J Mocco; R A Hanel; A H Siddiqui; L N Hopkins; E I Levy
Journal:  AJNR Am J Neuroradiol       Date:  2008-09-03       Impact factor: 3.825

5.  Progress in the Development of a new Angiography Suite including the High Resolution Micro-Angiographic Fluoroscope (MAF), a Control, Acquisition, Processing, and Image Display System (CAPIDS), and a New Detector Changer Integrated into a Commercial C-Arm Angiography Unit to Enable Clinical Use.

Authors:  Weiyuan Wang; Ciprian N Ionita; Christos Keleshis; Andrew Kuhls-Gilcrist; Amit Jain; Daniel Bednarek; Stephen Rudin
Journal:  Proc SPIE Int Soc Opt Eng       Date:  2010-03-23

6.  Angiographic analysis of animal model aneurysms treated with novel polyurethane asymmetric vascular stent (P-AVS): feasibility study.

Authors:  Ciprian N Ionita; Andreea Dohatcu; Andrey Sinelnikov; Jason Sherman; Christos Keleshis; Ann M Paciorek; K R Hoffmann; D R Bednarek; S Rudin
Journal:  Proc SPIE Int Soc Opt Eng       Date:  2009-01-01

7.  Generalized Objective Performance Assessment of a New High-Sensitivity Microangiographic Fluoroscopic (HSMAF) Imaging System.

Authors:  Girijesh K Yadava; Stephen Rudin; Andrew T Kuhls-Gilcrist; Daniel R Bednarek
Journal:  Proc SPIE Int Soc Opt Eng       Date:  2008
  7 in total