Literature DB >> 21243037

Progress in the Development of a new Angiography Suite including the High Resolution Micro-Angiographic Fluoroscope (MAF), a Control, Acquisition, Processing, and Image Display System (CAPIDS), and a New Detector Changer Integrated into a Commercial C-Arm Angiography Unit to Enable Clinical Use.

Weiyuan Wang1, Ciprian N Ionita, Christos Keleshis, Andrew Kuhls-Gilcrist, Amit Jain, Daniel Bednarek, Stephen Rudin.   

Abstract

Due to the high-resolution needs of angiographic and interventional vascular imaging, a Micro-Angiographic Fluoroscope (MAF) detector with a Control, Acquisition, Processing, and Image Display System (CAPIDS) was installed on a detector changer which was attached to the C-arm of a clinical angiographic unit. The MAF detector provides high-resolution, high-sensitivity, and real-time imaging capabilities and consists of a 300 μm-thick CsI phosphor, a dual stage micro-channel plate light image intensifier (LII) coupled to a fiber optic taper (FOT), and a scientific grade frame-transfer CCD camera, providing an image matrix of 1024×1024 35 μm square pixels with 12 bit depth. The Solid-State X-Ray Image Intensifier (SSXII) is an EMCCD (Electron Multiplying charge-coupled device) based detector which provides an image matrix of 1k×1k 32 μm square pixels with 12 bit depth. The changer allows the MAF or a SSXII region-of-interest (ROI) detector to be inserted in front of the standard flat-panel detector (FPD) when higher resolution is needed during angiographic or interventional vascular imaging procedures. The CAPIDS was developed and implemented using LabVIEW software and provides a user-friendly interface that enables control of several clinical radiographic imaging modes of the MAF or SSXII including: fluoroscopy, roadmapping, radiography, and digital-subtraction-angiography (DSA). The total system has been used for image guidance during endovascular image-guided interventions (EIGI) using prototype self-expanding asymmetric vascular stents (SAVS) in over 10 rabbit aneurysm creation and treatment experiments which have demonstrated the system's potential benefits for future clinical use.

Entities:  

Year:  2010        PMID: 21243037      PMCID: PMC3021378          DOI: 10.1117/12.844909

Source DB:  PubMed          Journal:  Proc SPIE Int Soc Opt Eng        ISSN: 0277-786X


  4 in total

1.  Implementation of a high-sensitivity Micro-Angiographic Fluoroscope (HS-MAF) for in-vivo endovascular image guided interventions (EIGI) and region-of-interest computed tomography (ROI-CT).

Authors:  C N Ionita; C Keleshis; V Patel; G Yadava; K R Hoffmann; D R Bednarek; A Jain; S Rudin
Journal:  Proc SPIE Int Soc Opt Eng       Date:  2008

2.  LabVIEW Graphical User Interface for a New High Sensitivity, High Resolution Micro-Angio-Fluoroscopic and ROI-CBCT System.

Authors:  C Keleshis; Cn Ionita; G Yadava; V Patel; Dr Bednarek; Kr Hoffmann; A Verevkin; S Rudin
Journal:  Proc SPIE Int Soc Opt Eng       Date:  2008

3.  The Solid-State X-Ray Image Intensifier (SSXII): An EMCCD-Based X-Ray Detector.

Authors:  Andrew Kuhls-Gilcrist; Girijesh Yadava; Vikas Patel; Amit Jain; Daniel R Bednarek; Stephen Rudin
Journal:  Proc SPIE Int Soc Opt Eng       Date:  2008

4.  Generalized Objective Performance Assessment of a New High-Sensitivity Microangiographic Fluoroscopic (HSMAF) Imaging System.

Authors:  Girijesh K Yadava; Stephen Rudin; Andrew T Kuhls-Gilcrist; Daniel R Bednarek
Journal:  Proc SPIE Int Soc Opt Eng       Date:  2008
  4 in total
  12 in total

1.  Graphics Processing Unit (GPU) implementation of image processing algorithms to improve system performance of the Control, Acquisition, Processing, and Image Display System (CAPIDS) of the Micro-Angiographic Fluoroscope (MAF).

Authors:  S N Swetadri Vasan; Ciprian N Ionita; A H Titus; A N Cartwright; D R Bednarek; S Rudin
Journal:  Proc SPIE Int Soc Opt Eng       Date:  2012-02-23

2.  High Resolution Emission and Transmission Imaging Using the Same Detector.

Authors:  Ashish S Panse; A Jain; W Wang; R Yao; D R Bednarek; S Rudin
Journal:  IEEE Nucl Sci Symp Conf Rec (1997)       Date:  2010-10-30

3.  A theoretical and experimental evaluation of the microangiographic fluoroscope: A high-resolution region-of-interest x-ray imager.

Authors:  Amit Jain; D R Bednarek; Ciprian Ionita; S Rudin
Journal:  Med Phys       Date:  2011-07       Impact factor: 4.071

4.  Endovascular coil embolization of a very small ruptured aneurysm using a novel microangiographic technique: technical note.

Authors:  Peter Kan; Parham Yashar; Ciprian N Ionita; Amit Jain; Stephen Rudin; Elad I Levy; Adnan H Siddiqui
Journal:  J Neurointerv Surg       Date:  2012-01-21       Impact factor: 5.836

5.  The Micro-Angiographic Fluoroscope (MAF) in High Definition (HD) Mode for Improved Contrast-to-Noise Ratio and Resolution in Fluoroscopy and Roadmapping.

Authors:  Ashish Panse; C N Ionita; W Wang; S K Natarajan; A Jain; D R Bednarek; S Rudin
Journal:  IEEE Nucl Sci Symp Conf Rec (1997)       Date:  2010-10-30

6.  Region-of-Interest Micro-Angiographic Fluoroscope Detector Used in Aneurysm and Artery Stenosis Diagnoses and Treatment.

Authors:  Weiyuan Wang; Ciprian Ionita; Ying Huang; Bin Qu; Ashish Panse; Amit Jain; Daniel R Bednarek; Stephen Rudin
Journal:  Proc SPIE Int Soc Opt Eng       Date:  2012-02-23

7.  Study of Stent Deployment Mechanics Using a High-Resolution X-ray Imaging Detector.

Authors:  Weiyuan Wang; Ciprian N Ionita; Daniel R Bednarek; Stephen Rudin
Journal:  Proc SPIE Int Soc Opt Eng       Date:  2011-03-01

8.  Use of the microangiographic fluoroscope for coiling of intracranial aneurysms.

Authors:  Mandy J Binning; David Orion; Parham Yashar; Sharon Webb; Ciprian N Ionita; Amit Jain; Stephen Rudin; L Nelson Hopkins; Adnan H Siddiqui; Elad I Levy
Journal:  Neurosurgery       Date:  2011-11       Impact factor: 4.654

9.  Evaluation of intracranial aneurysm coil embolization in phantoms and patients using a high-resolution Microangiographic Fluoroscope (MAF).

Authors:  Ciprian N Ionita; Amit Jain; Brendan Loughran; Swetadri Vasan S N; Daniel R Bednarek; Elad Levy; Adnan H Siddiqui; Kenneth V Snyder; L N Hopkins; Stephen Rudin
Journal:  Proc SPIE Int Soc Opt Eng       Date:  2012-03-02

10.  Workflow for the use of a high-resolution image detector in endovascular interventional procedures.

Authors:  R Rana; B Loughran; S N Swetadri Vasan; L Pope; C N Ionita; A Siddiqui; N Lin; D R Bednarek; S Rudin
Journal:  Proc SPIE Int Soc Opt Eng       Date:  2014-03-19
View more