Abstract
A rapid and specific high-performance liquid chromatography method with UV detection (HPLC-UV) for the simultaneous determination of 12 beta-lactam antibiotics (amoxicillin, cefepime, cefotaxime, ceftazidime, ceftriaxone, cloxacillin, imipenem, meropenem, oxacillin, penicillin G, piperacillin, and ticarcillin) in small samples of human plasma is described. Extraction consisted of protein precipitation by acetonitrile. An Atlantis T3 analytical column with a linear gradient of acetonitrile and a pH 2 phosphoric acid solution was used for separation. Wavelength photodiode array detection was set either at 210 nm, 230 nm, or 298 nm according to the compound. This method is accurate and reproducible (coefficient of variation [CV] < 8%), allowing quantification of beta-lactam plasma levels from 5 to 250 μg/ml without interference with other common drugs. This technique is easy to use in routine therapeutic drug monitoring of beta-lactam antibiotics.Entities:
Mesh:
Substances:
Year: 2011 PMID: 21788467 PMCID: PMC3187013 DOI: 10.1128/AAC.00533-11