Determination of a standard deviation that could minimize radiation exposure in an automatic exposure control for pulmonary thin-section computed tomography.

Atsushi Nambu1, Eiichi Sawada, Satoshi Kato, Tsutomu Araki, Yoshihito Aikawa, Makoto Yuge, Akitoshi Saito.   

Abstract

PURPOSE: The aim of this study was to determine a standard deviation (SD) that most reduces the radiation dose without sacrificing the diagnostic accuracy of thin-section computed tomography (CT) for clinical use.
MATERIALS AND METHODS: A total of 120 patients were examined by multidetector CT. They were assigned to one of four SD groups: 8, 9, 11, and 12. Each SD group consisted of 30 patients. The CT images of the same patients with SD10 that had formerly been examined were used for comparison. Two radiologists independently evaluated the degrees of image noise and diagnostic acceptability of the pulmonary diseases using a point score grading system. We compared the scores between each SD and the SD10 group.
RESULTS: Generally, image noise was significantly more prominent in the higher-SD groups. The mean score of diagnostic acceptability was significantly lower in the SD12 group (4.2 ± 1.6) than in the SD10 group (4.6 ± 1.1) group (P < 0.001), whereas no difference was present between the SD8 (4.9 ± 0.7), SD9 (4.8 ± 1.0), and SD11 (4.4 ± 1.5) groups and the SD10 group (4.7 ± 1.1, 4.6 ± 1.4, 4.6 ± 1.1, respectively).
CONCLUSION: Thin-section CT with SD12 is not acceptable. SD11 seems to be the setting with the lowest radiation dose while providing acceptable imaging quality for pulmonary thin-section CT.

Entities:  

Mesh:

Substances:

Year:  2011        PMID: 21786096     DOI: 10.1007/s11604-011-0571-0

Source DB:  PubMed          Journal:  Jpn J Radiol        ISSN: 1867-1071            Impact factor:   2.374


  10 in total

1.  Multislice helical CT of focal and diffuse lung disease: comprehensive diagnosis with reconstruction of contiguous and high-resolution CT sections from a single thin-collimation scan.

Authors:  U J Schoepf; R D Bruening; C Hong; R Eibel; S Aydemir; A Crispin; C Becker; M F Reiser
Journal:  AJR Am J Roentgenol       Date:  2001-07       Impact factor: 3.959

2.  Automatic exposure control in computed tomography--an evaluation of systems from different manufacturers.

Authors:  Marcus Söderberg; Mikael Gunnarsson
Journal:  Acta Radiol       Date:  2010-07       Impact factor: 1.990

3.  Low-dose high-resolution CT of lung parenchyma.

Authors:  C V Zwirewich; J R Mayo; N L Müller
Journal:  Radiology       Date:  1991-08       Impact factor: 11.105

Review 4.  Reducing radiation dose in emergency computed tomography with automatic exposure control techniques.

Authors:  Mannudeep K Kalra; Stefania M R Rizzo; Robert A Novelline
Journal:  Emerg Radiol       Date:  2005-05-12

5.  Identification and characterization of focal ground-glass opacity in the lungs by high-resolution CT using thin-section multidetector helical CT: experimental study using a chest CT phantom.

Authors:  Duo Liu; Kazuo Awai; Yoshinori Funama; Seitaro Oda; Takeshi Nakaura; Yumi Yanaga; Masahiro Hatemura; Koichi Kawanaka; Yasuyuki Yamashita
Journal:  Radiat Med       Date:  2008-01-31

6.  Use of 3D adaptive raw-data filter in CT of the lung: effect on radiation dose reduction.

Authors:  Takeshi Kubo; Yoshiharu Ohno; Shiva Gautam; Pei-Jan P Lin; Hans-Ulrich Kauczor; Hiroto Hatabu
Journal:  AJR Am J Roentgenol       Date:  2008-10       Impact factor: 3.959

7.  Chest CT performed with z-axis modulation: scanning protocol and radiation dose.

Authors:  Mannudeep K Kalra; Stefania Rizzo; Michael M Maher; Elkan F Halpern; Thomas L Toth; Jo-Anne O Shepard; Suzanne L Aquino
Journal:  Radiology       Date:  2005-10       Impact factor: 11.105

8.  Further reduction of radiation dose in helical CT for lung cancer screening using small tube current and a newly designed filter.

Authors:  S Itoh; S Koyama; M Ikeda; M Ozaki; A Sawaki; S Iwano; T Ishigaki
Journal:  J Thorac Imaging       Date:  2001-04       Impact factor: 3.000

9.  Detection of nodules showing ground-glass opacity in the lungs at low-dose multidetector computed tomography: phantom and clinical study.

Authors:  Yoshinori Funama; Kazuo Awai; Duo Liu; Seitaro Oda; Yumi Yanaga; Takeshi Nakaura; Koichi Kawanaka; Masamichi Shimamura; Yasuyuki Yamashita
Journal:  J Comput Assist Tomogr       Date:  2009 Jan-Feb       Impact factor: 1.826

10.  CT of the chest: minimal tube current required for good image quality with the least radiation dose.

Authors:  J R Mayo; T E Hartman; K S Lee; S L Primack; S Vedal; N L Müller
Journal:  AJR Am J Roentgenol       Date:  1995-03       Impact factor: 3.959

  10 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.