Reducing 4D CT artifacts using optimized sorting based on anatomic similarity.

Eric Johnston1, Maximilian Diehn, James D Murphy, Billy W Loo, Peter G Maxim.   

Abstract

PURPOSE: Four-dimensional (4D) computed tomography (CT) has been widely used as a tool to characterize respiratory motion in radiotherapy. The two most commonly used 4D CT algorithms sort images by the associated respiratory phase or displacement into a predefined number of bins, and are prone to image artifacts at transitions between bed positions. The purpose of this work is to demonstrate a method of reducing motion artifacts in 4D CT by incorporating anatomic similarity into phase or displacement based sorting protocols.
METHODS: Ten patient datasets were retrospectively sorted using both the displacement and phase based sorting algorithms. Conventional sorting methods allow selection of only the nearest-neighbor image in time or displacement within each bin. In our method, for each bed position either the displacement or the phase defines the center of a bin range about which several candidate images are selected. The two dimensional correlation coefficients between slices bordering the interface between adjacent couch positions are then calculated for all candidate pairings. Two slices have a high correlation if they are anatomically similar. Candidates from each bin are then selected to maximize the slice correlation over the entire data set using the Dijkstra's shortest path algorithm. To assess the reduction of artifacts, two thoracic radiation oncologists independently compared the resorted 4D datasets pairwise with conventionally sorted datasets, blinded to the sorting method, to choose which had the least motion artifacts. Agreement between reviewers was evaluated using the weighted kappa score.
RESULTS: Anatomically based image selection resulted in 4D CT datasets with significantly reduced motion artifacts with both displacement (P = 0.0063) and phase sorting (P = 0.00022). There was good agreement between the two reviewers, with complete agreement 34 times and complete disagreement 6 times.
CONCLUSIONS: Optimized sorting using anatomic similarity significantly reduces 4D CT motion artifacts compared to conventional phase or displacement based sorting. This improved sorting algorithm is a straightforward extension of the two most common 4D CT sorting algorithms.

Entities:  

Mesh:

Year:  2011        PMID: 21776777     DOI: 10.1118/1.3577601

Source DB:  PubMed          Journal:  Med Phys        ISSN: 0094-2405            Impact factor:   4.071


  11 in total

1.  Reduction of irregular breathing artifacts in respiration-correlated CT images using a respiratory motion model.

Authors:  Agung Hertanto; Qinghui Zhang; Yu-Chi Hu; Oleksandr Dzyubak; Andreas Rimner; Gig S Mageras
Journal:  Med Phys       Date:  2012-06       Impact factor: 4.071

2.  Reconstruction of four-dimensional computed tomography lung images by applying spatial and temporal anatomical constraints using a Bayesian model.

Authors:  Tiancheng He; Zhong Xue; Bin S Teh; Stephen T Wong
Journal:  J Med Imaging (Bellingham)       Date:  2015-05-13

3.  4D CT lung ventilation images are affected by the 4D CT sorting method.

Authors:  Tokihiro Yamamoto; Sven Kabus; Cristian Lorenz; Eric Johnston; Peter G Maxim; Maximilian Diehn; Neville Eclov; Cristian Barquero; Billy W Loo; Paul J Keall
Journal:  Med Phys       Date:  2013-10       Impact factor: 4.071

4.  Improving image-guided radiation therapy of lung cancer by reconstructing 4D-CT from a single free-breathing 3D-CT on the treatment day.

Authors:  Guorong Wu; Jun Lian; Dinggang Shen
Journal:  Med Phys       Date:  2012-12       Impact factor: 4.071

5.  Estimating the 4D respiratory lung motion by spatiotemporal registration and super-resolution image reconstruction.

Authors:  Guorong Wu; Qian Wang; Jun Lian; Dinggang Shen
Journal:  Med Phys       Date:  2013-03       Impact factor: 4.071

6.  Pulmonary ventilation imaging based on 4-dimensional computed tomography: comparison with pulmonary function tests and SPECT ventilation images.

Authors:  Tokihiro Yamamoto; Sven Kabus; Cristian Lorenz; Erik Mittra; Julian C Hong; Melody Chung; Neville Eclov; Jacqueline To; Maximilian Diehn; Billy W Loo; Paul J Keall
Journal:  Int J Radiat Oncol Biol Phys       Date:  2014-08-04       Impact factor: 7.038

7.  A Method for Assessing Ground-Truth Accuracy of the 5DCT Technique.

Authors:  Tai H Dou; David H Thomas; Dylan P O'Connell; James M Lamb; Percy Lee; Daniel A Low
Journal:  Int J Radiat Oncol Biol Phys       Date:  2015-08-07       Impact factor: 7.038

8.  Evaluation of 4D CT acquisition methods designed to reduce artifacts.

Authors:  Sarah J Castillo; Richard Castillo; Edward Castillo; Tinsu Pan; Geoffrey Ibbott; Peter Balter; Brian Hobbs; Thomas Guerrero
Journal:  J Appl Clin Med Phys       Date:  2015-03-08       Impact factor: 2.102

9.  An automated method for comparing motion artifacts in cine four-dimensional computed tomography images.

Authors:  Guoqiang Cui; Brian Jew; Julian C Hong; Eric W Johnston; Billy W Loo; Peter G Maxim
Journal:  J Appl Clin Med Phys       Date:  2012-11-08       Impact factor: 2.102

10.  Geodesic density regression for correcting 4DCT pulmonary respiratory motion artifacts.

Authors:  Wei Shao; Yue Pan; Oguz C Durumeric; Joseph M Reinhardt; John E Bayouth; Mirabela Rusu; Gary E Christensen
Journal:  Med Image Anal       Date:  2021-06-21       Impact factor: 13.828

View more

北京卡尤迪生物科技股份有限公司 © 2022-2023.