Evaluation of image quality for different kV cone-beam CT acquisition and reconstruction methods in the head and neck region.

Ulrik V Elstrøm1, Ludvig P Muren, Jørgen B B Petersen, Cai Grau.   

Abstract

PURPOSE: To evaluate the image quality obtained in a standard QA phantom with both clinical and non-clinical cone-beam computed tomography (CBCT) acquisition modes for the head and neck (HN) region as a step towards CBCT-based treatment planning. The impact of deteriorated Hounsfield unit (HU) accuracy was investigated by comparing results from clinical CBCT image reconstructions to those obtained from a pre-clinical scatter correction algorithm.
METHODS: Five different CBCT acquisition modes on a clinical system for kV CBCT-guided radiotherapy were investigated. Image reconstruction was performed in both standard clinical software and with an experimental reconstruction algorithm with improved beam hardening and scatter correction. Using the Catphan 504 phantom, quantitative measures of HU uniformity, HU verification and linearity, contrast-to-noise ratio (CNR), and spatial resolution using modulation transfer function (MTF) estimation were assessed. To benchmark the CBCT image properties, comparison to standard HN protocols on conventional CT scanners was performed by similar measures.
RESULTS: The HU uniformity within a water-equivalent homogeneous region was considerably improved using experimental vs. standard reconstruction, by factors of two for partial scans and four for full scans. Similarly, the amount of capping/cupping artifact was reduced by more than 1.5%. With mode and reconstruction specific HU calibration using seven inhomogeneity inserts comparable HU linearity was observed. CNR was on average 5% higher for experimental reconstruction (scaled with the square-root of dose between modes for both reconstruction methods).
CONCLUSIONS: Judged on parameters affecting the common diagnostic image properties, improved beam hardening and scatter correction diminishes the difference between CBCT and CT image quality considerably. In the pursuit of CBCT-based treatment adaptation, dedicated imaging protocols may be required.

Entities:  

Mesh:

Year:  2011        PMID: 21767191     DOI: 10.3109/0284186X.2011.590525

Source DB:  PubMed          Journal:  Acta Oncol        ISSN: 0284-186X            Impact factor:   4.089


  9 in total

1.  Optimization of an on-board imaging system for extremely rapid radiation therapy.

Authors:  Erica M Cherry Kemmerling; Meng Wu; He Yang; Peter G Maxim; Billy W Loo; Rebecca Fahrig
Journal:  Med Phys       Date:  2015-11       Impact factor: 4.071

2.  Regional 3D superimposition to assess temporomandibular joint condylar morphology.

Authors:  J Schilling; L C R Gomes; E Benavides; T Nguyen; B Paniagua; M Styner; V Boen; J R Gonçalves; L H S Cevidanes
Journal:  Dentomaxillofac Radiol       Date:  2013-10-29       Impact factor: 2.419

3.  Effect of dose reduction on image registration and image quality for cone-beam CT in radiotherapy.

Authors:  B Loutfi-Krauss; J Köhn; N Blümer; K Freundl; T Koch; E Kara; C Scherf; C Rödel; U Ramm; J Licher
Journal:  Strahlenther Onkol       Date:  2014-09-20       Impact factor: 3.621

4.  Characterization and comparison of imaging contrast enhancement with PEG-functionalized gold nanoparticles in kV cone beam computed tomography and computed tomography imaging.

Authors:  Tara Gray; Nema Bassiri; Shaquan David; Devanshi Yogeshkumar Patel; Mohammad Abdul-Moqueet; Neil Kirby; Kathryn M Mayer
Journal:  Biomed Phys Eng Express       Date:  2020-05-29

Review 5.  Cone Beam CT vs. Fan Beam CT: A Comparison of Image Quality and Dose Delivered Between Two Differing CT Imaging Modalities.

Authors:  Lawrence Lechuga; Georg A Weidlich
Journal:  Cureus       Date:  2016-09-12

6.  Spectral CT with monochromatic imaging and metal artifacts reduction software for artifacts reduction of ¹²⁵I radioactive seeds in liver brachytherapy.

Authors:  Qiuxia Yang; Sheng Peng; Jing Wu; Xiaohua Ban; Mingyan He; Chuanmiao Xie; Rong Zhang
Journal:  Jpn J Radiol       Date:  2015-10-11       Impact factor: 2.374

Review 7.  Applications of linac-mounted kilovoltage Cone-beam Computed Tomography in modern radiation therapy: A review.

Authors:  Kavitha Srinivasan; Mohammad Mohammadi; Justin Shepherd
Journal:  Pol J Radiol       Date:  2014-07-03

8.  Improving Image Quality of On-Board Cone-Beam CT in Radiation Therapy Using Image Information Provided by Planning Multi-Detector CT: A Phantom Study.

Authors:  Ching-Ching Yang; Fong-Lin Chen; Yeh-Chi Lo
Journal:  PLoS One       Date:  2016-06-09       Impact factor: 3.240

9.  On the improvement of CBCT image quality for soft tissue-based SRS localization.

Authors:  Weihua Mao; Stephen J Gardner; Karen Chin Snyder; Ning Winston Wen; Hualiang Zhong; Haisen Li; Paul Jackson; Mira Shah; Indrin J Chetty
Journal:  J Appl Clin Med Phys       Date:  2018-10-07       Impact factor: 2.102

  9 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.