Spectral CT with monochromatic imaging and metal artifacts reduction software for artifacts reduction of ¹²⁵I radioactive seeds in liver brachytherapy.

Qiuxia Yang1, Sheng Peng1, Jing Wu1, Xiaohua Ban1, Mingyan He1, Chuanmiao Xie1, Rong Zhang2.   

Abstract

PURPOSE: To investigate the optimal monochromatic energy for artifacts reduction from (125)I seeds as well as image improvement in the vicinity of seeds on monochromatic images with and without metal artifacts reduction software (MARS) and to compare this with traditional 120-kVp images, so as to evaluate the application value of gemstone spectral imaging for reducing artifacts from (125)I seeds in liver brachytherapy.
MATERIALS AND METHODS: A total of 45 tumors from 25 patients treated with (125)I seed brachytherapy in the liver were enrolled in this study. Multiphasic spectral computed tomography (CT) scanning was performed for each patient. After a delay time of 15 s of portal vein phase, a traditional 120-kVp scan was performed, focusing on several planes of (125)I seeds only. The artifact index (AI) in the vicinity of seeds and the standard deviation (SD) of the CT density of region of interest in the outside liver parenchyma were calculated. Artifact appearance was evaluated and classified on reconstructed monochromatic S and 120-kVp images. Image quality in the vicinity of seeds of three data sets were evaluated using a 1-5 scale scoring method. The Friedman rank-sum test was used to estimate the scoring results of image quality.
RESULTS: The greatest noise in monochromatic images was found at 40 keV (SD = 27.38, AI = 206.40). The optimal monochromatic energy was found at 75 keV, which provided almost the least image noise (SD = 10.01) and good performance in artifact reduction (AI = 102.73). Image noise and AI reduction at 75 keV was decreased by 63.44 and 50.23%, compared with at 40 keV. Near-field thick artifacts were obvious in all 45 lesions, in 120-kVp images, and 75-keV images, but basically reduced in 75 keV MARS images and artifacts completely invisible in 7 lesions. The number of diagnosable images (score ≥3) was significantly more in the 75-keV MARS group (28/45), and the 75-keV group (22/45) than in the 120-kVp group (11/45) (p < 0.0167 for both). Compared with 120-kVp images alone, 75-keV images plus 75-keV MARS images can increase tumor visibility around seeds and increase the proportion of diagnostic images to 84.4% (38/45).
CONCLUSION: Spectral CT producing 75-keV MARS images could substantially reduce near-field thick artifacts caused by (125)I seeds and improve image quality, even to a state of being completely free from artifacts. Spectral CT imaging (with and without MARS) can provide more accurate CT images for estimating efficacy after (125)I seed brachytherapy in the liver.

Entities:  

Keywords:  125I seed brachytherapy; Artifacts reduction; Metal artifacts reduction software; Monochromatic imaging; Spectral computed tomography

Mesh:

Substances:

Year:  2015        PMID: 26456321     DOI: 10.1007/s11604-015-0482-6

Source DB:  PubMed          Journal:  Jpn J Radiol        ISSN: 1867-1071            Impact factor:   2.374


  27 in total

1.  CT metal artifact reduction method correcting for beam hardening and missing projections.

Authors:  Joost M Verburg; Joao Seco
Journal:  Phys Med Biol       Date:  2012-04-18       Impact factor: 3.609

2.  Objective characterization of GE discovery CT750 HD scanner: gemstone spectral imaging mode.

Authors:  Da Zhang; Xinhua Li; Bob Liu
Journal:  Med Phys       Date:  2011-03       Impact factor: 4.071

3.  CT image artifacts from brachytherapy seed implants: a postprocessing 3D adaptive median filter.

Authors:  Parminder S Basran; Andrew Robertson; Derek Wells
Journal:  Med Phys       Date:  2011-02       Impact factor: 4.071

4.  Dual-energy (spectral) CT: applications in abdominal imaging.

Authors:  Alvin C Silva; Brian G Morse; Amy K Hara; Robert G Paden; Norio Hongo; William Pavlicek
Journal:  Radiographics       Date:  2011 Jul-Aug       Impact factor: 5.333

5.  Evaluation of image quality for different kV cone-beam CT acquisition and reconstruction methods in the head and neck region.

Authors:  Ulrik V Elstrøm; Ludvig P Muren; Jørgen B B Petersen; Cai Grau
Journal:  Acta Oncol       Date:  2011-08       Impact factor: 4.089

6.  Spectral CT with metal artifacts reduction software for improvement of tumor visibility in the vicinity of gold fiducial markers.

Authors:  Olga R Brook; Sofia Gourtsoyianni; Alexander Brook; Anand Mahadevan; Carol Wilcox; Vassilios Raptopoulos
Journal:  Radiology       Date:  2012-03-13       Impact factor: 11.105

7.  Virtual monochromatic spectral imaging with fast kilovoltage switching: improved image quality as compared with that obtained with conventional 120-kVp CT.

Authors:  Kazuhiro Matsumoto; Masahiro Jinzaki; Yutaka Tanami; Akihisa Ueno; Minoru Yamada; Sachio Kuribayashi
Journal:  Radiology       Date:  2011-02-17       Impact factor: 11.105

8.  Metal artifact reduction by dual energy computed tomography using monoenergetic extrapolation.

Authors:  Fabian Bamberg; Alexander Dierks; Konstantin Nikolaou; Maximilian F Reiser; Christoph R Becker; Thorsten R C Johnson
Journal:  Eur Radiol       Date:  2011-01-20       Impact factor: 5.315

9.  Monoenergetic imaging of dual-energy CT reduces artifacts from implanted metal orthopedic devices in patients with factures.

Authors:  Changsheng Zhou; Yan E Zhao; Song Luo; Hongyuan Shi; Lin Li; Ling Zheng; Long Jiang Zhang; Guangming Lu
Journal:  Acad Radiol       Date:  2011-10       Impact factor: 3.173

10.  Short- to mid-term evaluation of CT-guided 125I brachytherapy on intra-hepatic recurrent tumors and/or extra-hepatic metastases after liver transplantation for hepatocellular carcinoma.

Authors:  Fu-Jun Zhang; Chuan-Xing Li; Liang Zhang; Pei-Hong Wu; De-Chao Jiao; Guang-Feng Duan
Journal:  Cancer Biol Ther       Date:  2009-04-22       Impact factor: 4.742

View more
  4 in total

1.  Reduction of metal artifacts due to dental hardware in computed tomography angiography: assessment of the utility of model-based iterative reconstruction.

Authors:  Keita Kuya; Yuki Shinohara; Ayumi Kato; Makoto Sakamoto; Masamichi Kurosaki; Toshihide Ogawa
Journal:  Neuroradiology       Date:  2017-03-02       Impact factor: 2.804

2.  Combination of 125I brachytherapy and chemotherapy for unresectable recurrent breast cancer: A retrospective control study.

Authors:  Qixing Tan; Qinghong Qin; Weiping Yang; Bin Lian; Qinguo Mo; Changyuan Wei
Journal:  Medicine (Baltimore)       Date:  2016-11       Impact factor: 1.889

3.  Peritumoral abnormalities on dynamic-enhanced CT after brachytherapy for hepatic malignancies: local progression or benign changes?

Authors:  Guanyu Chen; Dechao Jiao; Sheng Peng; Xi Chen; Yanling Zhang; Letao Lin; Zhihui Zhong; Yong Li; Kaihao Xu; Fujun Zhang
Journal:  Eur Radiol       Date:  2022-08-18       Impact factor: 7.034

4.  Lobaplatin promotes 125I-induced apoptosis and inhibition of proliferation in hepatocellular carcinoma by upregulating PERK-eIF2α-ATF4-CHOP pathway.

Authors:  Dong Li; Wu-Jie Wang; Yong-Zheng Wang; Yi-Biao Wang; Yu-Liang Li
Journal:  Cell Death Dis       Date:  2019-10-03       Impact factor: 8.469

  4 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.