Literature DB >> 21751284

In situ active control of noise in a 4 T MRI scanner.

Mingfeng Li1, Brent Rudd, Teik C Lim, Jing-Huei Lee.   

Abstract

PURPOSE: To evaluate the effectiveness of the proposed active noise control (ANC) system for the reduction of the acoustic noise emission generated by a 4 T MRI scanner during operation and to assess the feasibility of developing an ANC device that can be deployed in situ.
MATERIALS AND METHODS: Three typical scanning sequences, EPI (echo planar imaging), GEMS (gradient echo multislice), and MDEFT (modified driven equilibrium Fourier transform), were used for evaluating the performance of the ANC system, which was composed of a magnetic compatible headset and a multiple reference feedforward filtered-x least mean square controller.
RESULTS: The greatest reduction, about 55 dB, was achieved at the harmonic at a frequency of 1.3 kHz in the GEMS case. Approximately 21 dB and 30 dBA overall reduction was achieved for GEMS noise across the entire audible frequency range. For the MDEFT sequence, the control system achieved 14 dB and 14 dBA overall reduction in the audible frequency range, while 13 dB and 14 dBA reduction was obtained for the EPI case.
CONCLUSION: The result is highly encouraging because it shows great potential for treating magnetic resonance imaging noise with an ANC application during real-time scanning.
Copyright © 2011 Wiley-Liss, Inc.

Entities:  

Mesh:

Year:  2011        PMID: 21751284      PMCID: PMC3193878          DOI: 10.1002/jmri.22694

Source DB:  PubMed          Journal:  J Magn Reson Imaging        ISSN: 1053-1807            Impact factor:   4.813


  22 in total

1.  Isolating the auditory system from acoustic noise during functional magnetic resonance imaging: examination of noise conduction through the ear canal, head, and body.

Authors:  M E Ravicz; J R Melcher
Journal:  J Acoust Soc Am       Date:  2001-01       Impact factor: 1.840

2.  Acoustic noise during functional magnetic resonance imaging.

Authors:  M E Ravicz; J R Melcher; N Y Kiang
Journal:  J Acoust Soc Am       Date:  2000-10       Impact factor: 1.840

3.  Acoustic absorption measurement of human hair and skin within the audible frequency range.

Authors:  B F Katz
Journal:  J Acoust Soc Am       Date:  2000-11       Impact factor: 1.840

Review 4.  Auditory noise associated with MR procedures: a review.

Authors:  M McJury; F G Shellock
Journal:  J Magn Reson Imaging       Date:  2000-07       Impact factor: 4.813

5.  Sound-level measurements and calculations of safe noise dosage during EPI at 3 T.

Authors:  J R Foster; D A Hall; A Q Summerfield; A R Palmer; R W Bowtell
Journal:  J Magn Reson Imaging       Date:  2000-07       Impact factor: 4.813

6.  Active control of the volume acquisition noise in functional magnetic resonance imaging: method and psychoacoustical evaluation.

Authors:  J Chambers; M A Akeroyd; A Q Summerfield; A R Palmer
Journal:  J Acoust Soc Am       Date:  2001-12       Impact factor: 1.840

7.  Simulation study on active noise control for a 4-T MRI scanner.

Authors:  Mingfeng Li; Teik C Lim; Jing-Huei Lee
Journal:  Magn Reson Imaging       Date:  2007-12-03       Impact factor: 2.546

8.  Acoustic noise characteristics of a 4 Telsa MRI scanner.

Authors:  Shashikant R More; Teik C Lim; Mingfeng Li; Christy K Holland; Suzanne E Boyce; Jing-Huei Lee
Journal:  J Magn Reson Imaging       Date:  2006-03       Impact factor: 4.813

9.  Acoustic noise reduction in a 4 T MRI scanner.

Authors:  Chris K Mechefske; Ryan Geris; Joseph S Gati; Brian K Rutt
Journal:  MAGMA       Date:  2002-01       Impact factor: 2.310

10.  Operational safety issues in MRI.

Authors:  R E Gangarosa; J E Minnis; J Nobbe; D Praschan; R W Genberg
Journal:  Magn Reson Imaging       Date:  1987       Impact factor: 2.546

View more
  3 in total

1.  Adaptive speech enhancement using directional microphone in a 4-T MRI scanner.

Authors:  Guohua Sun; Mingfeng Li; Brent W Rudd; Teik C Lim; Jeffrey Osterhage; Elizabeth M Fugate; Jing-Huei Lee
Journal:  MAGMA       Date:  2015-04-18       Impact factor: 2.310

Review 2.  Gradient and shim technologies for ultra high field MRI.

Authors:  Simone A Winkler; Franz Schmitt; Hermann Landes; Joshua de Bever; Trevor Wade; Andrew Alejski; Brian K Rutt
Journal:  Neuroimage       Date:  2016-11-30       Impact factor: 6.556

Review 3.  Methodological challenges and solutions in auditory functional magnetic resonance imaging.

Authors:  Jonathan E Peelle
Journal:  Front Neurosci       Date:  2014-08-21       Impact factor: 4.677

  3 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.