Effects of the transcutaneous electrode temperature on the accuracy of transcutaneous carbon dioxide tension.

Line C Sørensen1, Lene Brage-Andersen, Gorm Greisen.   

Abstract

AIM: The harmful effect of hypocapnia on the neonatal brain emphasizes the importance of monitoring arterial carbon dioxide tension (PaCO2). Transcutaneous monitoring of carbon dioxide (tcPCO2) reduces the need for arterial blood sampling. Drawbacks are high electrode temperature causing risks of skin burning. The aim was to determine the accuracy and precision of tcPCO2 at reduced electrode temperature.
METHODS: Forty newborns (GA 24.9-41.7) were included. Two tc-monitors were applied (TCM4, Radiometer, Copenhagen). Arterial blood gas sampling and monitoring of tcPCO2-level at different electrode temperatures was done simultaneously (39°C, 40°C, 41°C, 42°C, 44°C). Difference of PaCO2 - tcPCO2 was expressed as a percentage of the mean.
RESULTS: Mean PaCO2 was 5.8kPa [3,2; 7.9]. Bias (PaCO2 - tcPCO2) increased from 5% at 44°C to 17% at 39°C, but did not differ significantly between 41°C and 40°C. The precision of the tcPCO2 at each temperature ranged from +7-10%. After correction for the temperature-dependent overreading, we found increasing PaCO2 - tcPCO2 difference with increasing PaCO2, approx. 2% pr. kPa increase of CO(2). Only mild transient erythema was observed.
CONCLUSION: A lower electrode temperature in tcPCO2-monitoring increases systematic overreading of the tc-electrode. However, in very preterm babies, monitoring at 40°C or 41°C is possible provided a bias correction of 12-15% is applied.

Entities:  

Mesh:

Substances:

Year:  2011        PMID: 21732731      PMCID: PMC3212912          DOI: 10.3109/00365513.2011.590601

Source DB:  PubMed          Journal:  Scand J Clin Lab Invest        ISSN: 0036-5513            Impact factor:   1.713


Introduction

The harmful effects of hyperventilation and hypocapnia on the neonatal brain emphasize the importance of a stable arterial carbon dioxide tension (PaCO2) especially in mechanically ventilated newborns [1]. Transcutaneous monitoring (TCM) allows continuous monitoring of the transcutaneous carbon dioxide (tcPCO2) – a surrogate measure of the PaCO2. Hence, the need of arterial blood sampling is reduced. The transcutaneous electrode conventionally is heated to 44°C to arterialize the capillary bed. The correlation between tcPCO2 and PaCO2 depends on the temperature in the capillary bed, the blood flow to the skin, and the local metabolic production of CO2. Drawbacks of the high electrode temperature are risk of skin burns [2,3]. The correlation between time and surface temperature in the causation of cutaneous burns has been known for a long time [4]. Early studies in preterm infants showed a direct correlation between the duration of electrode application and persistence of erythema; while the application time is shortened [2]. Besides disturbance of the neonate, relocation of the electrode gives unstable values during stabilization. In the early era of transcutaneous monitoring, several studies were performed, improving the monitors in use today [5,6]. This also includes studies with monitoring at different electrode temperatures [7-9]. Because of the fragile skin of preterm infants a lower electrode temperature would be appreciated. This could reduce the risk of skin burns and prolong the application time. However, a recent study in adults examined reduced electrode temperature using the TCM4, Radiometer Medical ApS, Denmark [10]. Due to bias, the authors concluded that the electrode should be heated to at least 43°C to measure reliable tcPCO2 [10]. The aim of this study was to determine the accuracy and precision of tcPCO2 in preterm infants using TCM4, Radiometer Medical ApS, Denmark, with reduced electrode temperatures. Our assumption was that electrode temperatures of 39-41°C would be less damaging to the skin, since these temperatures are reached during systemic fever.

Patients and methods

The study was conducted in the NICU at the Copenhagen University Hospital, Rigshospitalet. The local ethics committee approved the study (J. nr. (KF) 01 283061). Parental consent was obtained in all cases. This paper consists of two studies. In the first study, pCO2 precision was examined at five different electrode temperatures (39, 40, 41, 42 and 44°C). Recruitment was from January to April 2006. Inclusion criteria were clinical indication of transcutaneous (TC) monitoring and presence of an umbilical or radial arterial catheter. A total of 20 newborns irrespective of gestational age (GA) were included (= GA mixed group). In the second study, based on results from study one, comparisons of pCO2 precision at selected electrode temperatures were done (40, 41 and 44°C). Recruitment was from November 2006 to June 2008. Inclusions criteria was GA below 34 weeks or birth weight below 1500 g, clinical indication of TC monitoring and presence of an umbilical or radial arterial catheter. A total of 20 preterm infants were included (= Preterm group). Inclusion and examination was within the neonatal period (first 28 days of life). All newborns were monitored using two TC monitors simultaneously (TCM 4 with combined CO2/O2 electrode E5280, Radiometer Medical ApS, Denmark). Because of higher values of pCO2 when elevating skin temperature (anaerobic factor 4.5%/°C), the CO2 production of epidermal cells ('metabolic’ constant) adding to the pCO2 measured transcutaneously, the monitor has a correction algorithm integrated [11]. We used standard recommendations with the anaerobic factor (Severinghaus constant) turned on, and with a metabolic constant set at 0.5 kPa. Using two TCM4 simultaneously in the first study, allowed one electrode to remain unchanged at 44°C while the temperature of the other electrode was increased every half hour (39, 40, 41, 42 and 44°C) (Figure 1).
Figure 1

The sequence of the transcutaneous electrode temperature changes in the two studies.

The sequence of the transcutaneous electrode temperature changes in the two studies. Using two TCM4 simultaneously in the second study allowed a cross over design to investigate the impact of preheating of the skin on pCO2 precision (Figure 1). The TC electrodes were fixed to the skin of the trunk using a sticky fixation ring (E5260/E5280: Fixation ring 904 – 891; 30 mm; Radiometer Medical ApS, Denmark). The localization was not changed during the study. The electrode housing was 15 mm in diameter. After a few droplets of an electrolyte solution enhancing contact between electrode and skin, the electrode was fixed in the fixation ring (Electrolyte solution; Radiometer Medical ApS, Denmark). The ‘gold standard’ to which the TC values were compared was the arterial pCO2 (PaCO2). All arterial blood samples were taken from the indwelling catheter. Blood was sampled in heparinized capillary tubes (100 microliter). Immediately after sampling the tube was turned slowly 20 times. Within 5 min the blood gases were analysed using Blood Gas Analyzer ABL 735 (Radiometer Medical ApS, Denmark). In the second study, the first blood sample was analysed in duplicate to estimate test-retest variation of arterial pCO2. Arterial blood gas sampling and readings of the tcPCO2-level was done simultaneously. After changing the electrode temperature, a minimum of 30 min. was allowed for temperature equilibration and for stabilizing tcPCO2 before blood sampling. Before measuring at a new electrode temperature, the sensor was calibrated. The monitor automatically performs this process using a built in calibration module and gas cylinder. Calibration of the sensor is recommended as a routine procedure in TC monitoring using TCM4. The sensor was re-membraned when needed (E5260/E5280: Membrane 904 – 892; Radiometer Medical ApS, Denmark). Statistical analysis was done using SPSS version 12.0 (Chicago). To estimate the repeatability of the blood gas analysis we calculated the standard deviation of the differences between the pairs of blood gas measurements. Univariate ANOVA was used to test the influence of electrode temperature on accuracy (ΔtcPCO2PaCO2). 95% limits of agreement were calculated as described by Bland and Altman [12]. As the difference between transcutaneous and arterial pCO2 increased with increasing pCO2, the difference between transcutaneous and arterial values was expressed as a percentage of the mean (100 × (diff. tcPCO2PaCO2/mean tcPCO2PaCO2)). The Kruskal-Wallis test was used for comparisons.

Results

The median gestational age in the GA mixed group was 31.1 weeks [24.9; 41.7] and median birth weight 1822 g [735; 3800 g] (Table I).
Table I

Group characteristics.

GA mixed group (N = 20)Preterm group (N = 20)
Gestational age (weeks)31.1 [24.9; 41.7]28.9 [25.6; 32.0]
BW (grams)1822 [735; 3800]1006 [590; 1900]
Nasal CPAP/Mechanical ventilation50%/35%55%/45%
No oxygen demand55%30%
Oxygen supplement (% O2)21% [21%; 55%]28% [21%; 70%]
Dopamine0%10%
Group characteristics. Diagnoses were manifold: Prematurity (n = 11), asphyxia (n = 1), meconium aspiration syndrome (n = 1), necrotizing enterocolitis (n = 1), congenital heart disease (n = 3), oesophageal atresia (n = 1), extracorporeal membrane oxygenation (n = 1) and no diagnosis specified (n = 2). In the preterm group, the median gestational age was 28.9 weeks [25.6; 32.0] and median birth weight 1006 g [590 g; 1900 g] (Table I). All infants (n = 20) were preterm, however other diagnoses included pneumothorax (n = 4), rhesus immunization (n = 1), persistent fetal circulation (n = 1) and lung bleeding (n = 1). No infants were excluded due to instability, since this was not a part of the protocol. However, if there were any events that potentially could influence results, it was remarked. In the GA mixed group, two infants had remarks: ‘restless’ (n = 1) and ‘apnoea, bradycardia and desaturation of prematurity’ (n = 1), while in the preterm group four infants had remarks: ‘restless’ (n = 2), ‘increasing oxygen demand’ (n = 2), ‘apnoea, bradycardia and desaturation of prematurity’ (n = 4). Arterial pCO2 (n = 220) was normally distributed (range 3.2-7.9; mean 5.8 kPa ± 1.0 kPa) and was not associated with electrode temperature (p = 0.9). The repeatability of arterial pCO2 measurements (n = 19) was 0.07 ± 0.15 kPa. By defining a maximum of 1 kPa bias between tcPCO2 and PaCO2, we examined how many of the paired samples met these qualifications. When using an electrode temperature of 39°C 45% had a bias below 1 kPa (9/20 measurements), at 40°C 73% (44/60 measurements), at 41°C 68% (27/40 measurements), at 42°C 90% (18/20 measurements) and at 44°C 94% had a bias below 1 kPa (75/80 measurements). The mean PaCO2 - tcPCO2 difference (bias) increased from 5% at 44°C to 17% at 39°C, but did not differ significantly between 41°C and 40°C (median 14.8% vs. 11.8%) (Figures 2 & 3). The precision of the tcPCO2 at each temperature ranged from 7-10%. After correction for the temperature-dependent overreading, we found increasing PaCO2 -tcPCO2 difference with increasing PaCO2, an approx. 2% pr. kPa increase of pCO2.
Figure 2

The difference between transcutaneous pCO2 and arterial pCO2 increases at lower electrode temperatures. The median difference is +12% at 40-41°C. Outliers also occur at high electrode temperature.

Figure 3

This figure shows the mean pCO2 (kPa) versus the temperature corrected bias of TcPCO2 (fraction, %). The horizontal lines show the mean bias of TcPCO2 at 40°C (12%) and 44°C (5%). Correction with 12-15% when using 40°C will approximate measurements to the arterial pCO2.

The difference between transcutaneous pCO2 and arterial pCO2 increases at lower electrode temperatures. The median difference is +12% at 40-41°C. Outliers also occur at high electrode temperature. This figure shows the mean pCO2 (kPa) versus the temperature corrected bias of TcPCO2 (fraction, %). The horizontal lines show the mean bias of TcPCO2 at 40°C (12%) and 44°C (5%). Correction with 12-15% when using 40°C will approximate measurements to the arterial pCO2. All outliers were above the 95% confidence interval. These outliers occurred primarily at low electrode temperature. Sequence of electrode temperatures, including preheating of the skin to 44°C, had no influence on precision. We did not find any significant electrode drift after each session by control measurement using the calibration gas. No skin lesions apart from mild transient erythema were observed.

Discussion

The aim of this study was to investigate how reduction of the electrode temperature influences accuracy and precision in tcPCO2 measurements in newborn infants. Lower electrode temperature reduces the risk of skin burning and allows longer periods of continued monitoring [13]. The bias of tcPCO2 using the normal electrode temperature of 44°C was approximately 5%. We had set the metabolic constant at 0.53 kPa (4 mmHg), which appears to be too low for our population. Lower electrode temperatures were associated with higher bias, but at 41°C the bias was still typically less than 1 kPa. Precision was unchanged. We therefore conclude that it is reasonable to monitor tcPCO2 with a temperature of 41°C, particularly if the bias of 12-15% is accounted for. However, changing the electrode temperature requires an increased clinical awareness. With measurements close to the normal PaCO2 range an error of 1 kPa will have no great consequences, however readings in the subnormal range of PaCO2 can have a significant impact on cerebral perfusion. On our ward, we recommend an arterial blood gas if there is any doubt of the actual level of PaCO2 and in particular if the tcPCO2 is out of the normal range. At 41°C, the tcPO2 cannot be relied upon, because the capillary bed will not be sufficiently arterialized. The best procedure would be to deactivate the pO2 readout of the TCM4. In most situations, oxygen treatment of newborns can be sufficiently monitored by pulseoximetry. Our results agree reasonably with previous studies. A study in neonates found a good correlation between tcPCO2 and arterial CO2 at 42°C and even 37°C [7]. In adults, a good correlation was found between transcutaneous and capillary pCO2 at 37, 39, 41, 43 and 45°C, provided the skin was preheated to 45°C [14]. The correlation coefficient, however, ignores bias. Fanconi and colleagues reduced the electrode temperature to 42°C and had reliable results, but only after topical metabolic inhibition [13]. Furthermore, after 12 h continuous use some neonates had skin burns [13]. In adults, the TOSCA with an electrode temperature of 42°C had a bias of +0.02 kPa, only [15]. The TOSCA monitor initially preheats the skin to 45°C for rapid arterialization before measurements and measurements were performed at the earlobe, using an ear clips. We did not find any effect of using high temperature before low temperature, or vice versa. Surprisingly, in our study the bias was pCO2-dependent. The transcutaneous electrode uses an electrochemical principle: tcPCO2 is measured by determining the pH of an electrolyte layer separated from the skin by a highly permeable membrane. Diffusing CO2 changes the pH of the electrolyte solution and the pH changes are proportional to the logarithm of pCO2-changes. Since tissue heating increases tissue pCO2 by 4.5%/°C and some CO2 is produced locally, the tcPCO2 is higher than the arterial pCO2 and corrections for this are included in the monitor. For optimal use, the electrolyte and membrane must be changed every week, and the sensor must also be calibrated whenever a new patient is monitored. This was done meticulously for our two experimental monitors and furthermore control measurements using the calibration gas were done after measurements in all infants. If out of the acceptable range, recalibration and/or change of membrane was done. Therefore we cannot explain the pCO2 dependency of the bias. Combined with the upwards skew of the distribution of transcutaneous-arterial pCO2 difference, it meant that occasionally tcPCO2 overestimated significantly, and this occurred more often at high arterial tcPCO2. In practical terms this means that a very high tcPCO2 must be used with care. Safe practice would be taking a blood sample before significant clinical action, e.g. intubation for mechanical ventilation. In conclusion, a lower transcutaneous electrode temperature increases the systematic overreading of tcPCO2. Monitoring at 40 or 41°C, however, appears reasonable since the typical bias is less than 1 kPa. In some cases with high arterial pCO2 the overreading was clinically significant and care should be taken that this does not have undue clinical consequences.
  15 in total

Review 1.  Applying the right statistics: analyses of measurement studies.

Authors:  J M Bland; D G Altman
Journal:  Ultrasound Obstet Gynecol       Date:  2003-07       Impact factor: 7.299

2.  May hypocarbia cause ischaemic brain damage in the preterm infant?

Authors:  G Greisen; H Munck; H Lou
Journal:  Lancet       Date:  1986-08-23       Impact factor: 79.321

3.  Comparison of the transcutaneous oxygen and carbon dioxide tension in different electrode locations during general anaesthesia.

Authors:  T Nishiyama; S Nakamura; K Yamashita
Journal:  Eur J Anaesthesiol       Date:  2006-07-07       Impact factor: 4.330

4.  Optimal temperature for the measurement of transcutaneous carbon dioxide tension in the neonate.

Authors:  N Herrell; R J Martin; M Pultusker; M Lough; A Fanaroff
Journal:  J Pediatr       Date:  1980-07       Impact factor: 4.406

5.  Long-term transcutaneous monitoring of oxygen tension and carbon dioxide at 42 degrees C in critically ill neonates: improved performance of the tcpo2 monitor with topical metabolic inhibition.

Authors:  S Fanconi; A Tschupp; L Molinari
Journal:  Eur J Pediatr       Date:  1996-12       Impact factor: 3.183

6.  Skin craters--a complication of transcutaneous oxygen monitoring.

Authors:  S M Golden
Journal:  Pediatrics       Date:  1981-04       Impact factor: 7.124

7.  The measurement of TcPO2 and TcPCO2 in newborn infants at 44 degrees C, 42 degrees C and 37 degrees C after initial heating to 44 degrees C.

Authors:  B Friis-Hansen; P Voldsgaard; J Witt; K G Pedersen; P S Frederiksen
Journal:  Adv Exp Med Biol       Date:  1987       Impact factor: 2.622

8.  Transcutaneous analysis of arterial PCO2.

Authors:  T A Hazinski; J W Severinghaus
Journal:  Med Instrum       Date:  1982 May-Jun

Review 9.  The design, use, and results of transcutaneous carbon dioxide analysis: current and future directions.

Authors:  Patrick Eberhard
Journal:  Anesth Analg       Date:  2007-12       Impact factor: 5.108

10.  Transcutaneous carbon dioxide and oxygen tension measured at different temperatures in healthy adults.

Authors:  P D Wimberley; K Grønlund Pedersen; J Olsson; O Siggaard-Andersen
Journal:  Clin Chem       Date:  1985-10       Impact factor: 8.327

View more
  7 in total

Review 1.  Cardiorespiratory events in preterm infants: etiology and monitoring technologies.

Authors:  J M Di Fiore; C F Poets; E Gauda; R J Martin; P MacFarlane
Journal:  J Perinatol       Date:  2015-11-19       Impact factor: 2.521

2.  Agreement of Mixed Venous Carbon Dioxide Tension (PvCO2) and Transcutaneous Carbon Dioxide (PtCO2) Measurements in Ventilated Infants.

Authors:  Sinan Uslu; Ali Bulbul; Mesut Dursun; Umut Zubarioglu; Ebru Turkoglu; Omer Guran
Journal:  Iran J Pediatr       Date:  2015-01-19       Impact factor: 0.364

Review 3.  Transcutaneous carbon dioxide monitoring for the prevention of neonatal morbidity and mortality.

Authors:  Matteo Bruschettini; Olga Romantsik; Simona Zappettini; Luca Antonio Ramenghi; Maria Grazia Calevo
Journal:  Cochrane Database Syst Rev       Date:  2016-02-13

4.  The application of transcutaneous CO2 pressure monitoring in the anesthesia of obese patients undergoing laparoscopic bariatric surgery.

Authors:  Shijiang Liu; Jie Sun; Xing Chen; Yingying Yu; Xuan Liu; Cunming Liu
Journal:  PLoS One       Date:  2014-04-03       Impact factor: 3.240

5.  Accuracy of Transcutaneous Carbon Dioxide Measurement in Premature Infants.

Authors:  Marie Janaillac; Sonia Labarinas; Riccardo E Pfister; Oliver Karam
Journal:  Crit Care Res Pract       Date:  2016-06-08

6.  Feasibilty of Transcutaneous pCO2 Monitoring During Immediate Transition After Birth-A Prospective Observational Study.

Authors:  Ilia Bresesti; Marlies Bruckner; Christian Mattersberger; Nariae Baik-Schneditz; Bernhard Schwaberger; Lukas Mileder; Alexander Avian; Berndt Urlesberger; Gerhard Pichler
Journal:  Front Pediatr       Date:  2020-01-29       Impact factor: 3.418

Review 7.  Non-invasive carbon dioxide monitoring in neonates: methods, benefits, and pitfalls.

Authors:  Deepika Sankaran; Lida Zeinali; Sameeia Iqbal; Praveen Chandrasekharan; Satyan Lakshminrusimha
Journal:  J Perinatol       Date:  2021-06-19       Impact factor: 2.521

  7 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.