Non-invasive haemoglobin estimation in patients with thalassaemia major.

Murtadha K Al Khabori1, Arwa Z Al-Riyami1, Khalil Al-Farsi1, Mohammed Al-Huneini1, Abdulhakeem Al-Hashim2, Nasser Al-Kemyani3, Issa Al-Qarshoubi2, Hammad Khan4, Khalfan Al-Amrani5, Shahina Daar4.   

Abstract

OBJECTIVES: This study aimed to validate pulse CO-oximetry-based haemoglobin (Hb) estimation in children and adults with thalassaemia major (TM) and to determine the impact of different baseline variables on the accuracy of the estimation.
METHODS: This observational study was conducted over a five-week period from March to April 2012. A total of 108 patients with TM attending the daycare thalassaemia centre of a tertiary care hospital in Muscat, Oman, were enrolled. Spot (Sp) Hb measurements were estimated using a Pronto-7(®) pulse CO-oximetry device (Masimo Corp., Irvine, California, USA). These were compared to venous samples of Hb using the CELL-DYN Sapphire Hematology Analyzer (Abbott Diagnostics, Abbott Park, Illinois, USA) to determine the reference (Ref) Hb levels. A multivariable linear regression model was used to assess the impact of baseline variables such as age, gender, weight, height, Ref Hb and blood pressure on the Hb estimations.
RESULTS: Of the 108 enrolled patients, there were 54 males and 54 females with a mean age of 21.6 years (standard deviation [SD] = 7.3 years; range: 2.5-38 years). The mean Ref Hb and Sp Hb were 9.4 g/dL (SD = 0.9 g/dL; range: 7.5-12.3 g/dL) and 11.1 g/dL (SD = 1.2 g/dL; range: 7.5-14.7 g/dL), respectively. The coefficient of determination (R(2)) was 21% with a mean difference of 1.7 g/dL (SD = 1.1 g/dL; range: -0.9-4.3 g/dL). In the multivariable model, the Ref Hb level (P = 0.001) was the only statistically significant predictor.
CONCLUSION: The Pronto-7(®) pulse CO-oximetry device was found to overestimate Hb levels in patients with TM and therefore cannot be recommended. Further larger studies are needed to confirm these results.

Entities:  

Keywords:  Hemoglobin; Pulse Oximetry; Validation Studies; beta Thalassemia

Year:  2014        PMID: 25364548      PMCID: PMC4205057     

Source DB:  PubMed          Journal:  Sultan Qaboos Univ Med J        ISSN: 2075-051X


  16 in total

1.  Application of the Bland-Altman plot for interpretation of method-comparison studies: a critical investigation of its practice.

Authors:  Katy Dewitte; Colette Fierens; Dietmar Stöckl; Linda M Thienpont
Journal:  Clin Chem       Date:  2002-05       Impact factor: 8.327

2.  Validation of a non-invasive pulse CO-oximetry based hemoglobin estimation in normal blood donors.

Authors:  Murtadha Al-Khabori; Arwa Z Al-Riyami; Khalil Al-Farsi; Mohamed Al-Huneini; Abdulhakeem Al-Hashim; Nasser Al-Kemyani; Shahina Daar
Journal:  Transfus Apher Sci       Date:  2013-11-04       Impact factor: 1.764

3.  Forecasting hemoglobinopathy burden through neonatal screening in Omani neonates.

Authors:  Salam Alkindi; Shoaib Al Zadjali; Ali Al Madhani; Shahina Daar; Hamood Al Haddabi; Qamariya Al Abri; David Gravell; Tsouria Berbar; Sahaya Pravin; Anil Pathare; Rajagopal Krishnamoorthy
Journal:  Hemoglobin       Date:  2010-01       Impact factor: 0.849

4.  The accuracy of noninvasive hemoglobin measurement by multiwavelength pulse oximetry after cardiac surgery.

Authors:  Ba-Vinh Nguyen; Jean-Louis Vincent; Emmanuel Nowak; Michelle Coat; Nicolas Paleiron; Pierre Gouny; Mehdi Ould-Ahmed; Maité Guillouet; Charles Christian Arvieux; Gildas Gueret
Journal:  Anesth Analg       Date:  2011-09-14       Impact factor: 5.108

Review 5.  Pulse oximetry in pediatric practice.

Authors:  Sotirios Fouzas; Kostas N Priftis; Michael B Anthracopoulos
Journal:  Pediatrics       Date:  2011-09-19       Impact factor: 7.124

6.  Validation of a noninvasive pulse CO-oximetry-based hemoglobin estimation in patients with sickle cell disease.

Authors:  M Al-Khabori; A Al-Hashim; Z Jabeen; K Al-Farsi; M Al-Huneini; A Al-Riyami; N Al-Kemyani; S Daar
Journal:  Int J Lab Hematol       Date:  2013-03-11       Impact factor: 2.877

7.  Continuous noninvasive hemoglobin monitoring during complex spine surgery.

Authors:  Lauren Berkow; Stephanie Rotolo; Erin Mirski
Journal:  Anesth Analg       Date:  2011-09-29       Impact factor: 5.108

Review 8.  Pulse oximetry: applications and limitations.

Authors:  S J Barker; K K Tremper
Journal:  Int Anesthesiol Clin       Date:  1987

9.  Side effects of Deferasirox Iron Chelation in Patients with Beta Thalassemia Major or Intermedia.

Authors:  Murtadha Al-Khabori; Sunil Bhandari; Mohammed Al-Huneini; Khalil Al-Farsi; Vinodh Panjwani; Shahina Daar
Journal:  Oman Med J       Date:  2013-03

10.  Non-invasive measurement of hemoglobin: assessment of two different point-of-care technologies.

Authors:  Etienne Gayat; Jérôme Aulagnier; Emmanuel Matthieu; Mireille Boisson; Marc Fischler
Journal:  PLoS One       Date:  2012-01-06       Impact factor: 3.240

View more

北京卡尤迪生物科技股份有限公司 © 2022-2023.