Radiochromic film dosimetry: reflection vs transmission scanning.

J Kalef-Ezra1, K Karava.   

Abstract

The objective of this study was to compare the transmission and the reflection scanning reading modes of radiochromic MD-55 films of a document scanner. The use of the red channel of the red-green-blue images results in measurement accuracy that did not differ between the two reading modes. On the other hand, the scanner uniformity and temporal stability, the long-term stability of the film readings and the dynamic range is superior using the reflection mode. The dynamic range of dose measurements can be extended up to least 500 Gy using the alternative color channels by both reading techniques. Similar backscattering dose factors close to high Z inhomogeneities were found using both scanning modes. In conclusion, the use of reflection scanning mode of MD55 films was superior to the traditionally used transmission mode.

Mesh:

Year:  2008        PMID: 18649463     DOI: 10.1118/1.2919092

Source DB:  PubMed          Journal:  Med Phys        ISSN: 0094-2405            Impact factor:   4.071


  5 in total

1.  Calibration of GafChromic XR-RV3 radiochromic film for skin dose measurement using standardized x-ray spectra and a commercial flatbed scanner.

Authors:  Bradley P McCabe; Michael A Speidel; Tina L Pike; Michael S Van Lysel
Journal:  Med Phys       Date:  2011-04       Impact factor: 4.071

2.  A simple 96 well microfluidic chip combined with visual and densitometry detection for resource-poor point of care testing.

Authors:  Minghui Yang; Steven Sun; Yordan Kostov; Avraham Rasooly
Journal:  Sens Actuators B Chem       Date:  2011-03-31       Impact factor: 7.460

3.  Scanning methodology for contact lens-type ocular in vivo dosimeter (CLOD) dosimetry applying a silicone material.

Authors:  Jaeman Son; Jin Dong Cho; Seongmoon Jung; Chang Heon Choi; Jong Min Park; Jung-In Kim
Journal:  Radiat Oncol       Date:  2022-05-07       Impact factor: 4.309

4.  Precise film dosimetry for stereotactic radiosurgery and stereotactic body radiotherapy quality assurance using Gafchromic™ EBT3 films.

Authors:  Ning Wen; Siming Lu; Jinkoo Kim; Yujiao Qin; Yimei Huang; Bo Zhao; Chang Liu; Indrin J Chetty
Journal:  Radiat Oncol       Date:  2016-10-04       Impact factor: 3.481

5.  Comparison of Epson scanner quality for radiochromic film evaluation.

Authors:  Hani Alnawaf; Peter K N Yu; Martin Butson
Journal:  J Appl Clin Med Phys       Date:  2012-09-06       Impact factor: 2.102

  5 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.