A Wire Crossed-Loop-Resonator for Rapid Scan EPR.

George A Rinard1, Richard W Quine, Joshua R Biller, Gareth R Eaton.   

Abstract

A crossed-loop (orthogonal mode) resonator (CLR) was constructed of fine wire to achieve design goals for rapid scan in vivo EPR imaging at VHF frequencies (in practice, near 250 MHz). This application requires the resonator to have a very open design to facilitate access to the animal for physiological support during the image acquisition. The rapid scan experiment uses large amplitude magnetic field scans, and sufficiently large resonator and detection bandwidths to record the rapidly-changing signal response. Rapid-scan EPR is sensitive to RF/microwave source noise and to baseline changes that are coherent with the field scan. The sensitivity to source noise is a primary incentive for using a CLR to isolate the detected signal from the RF source noise. Isolation from source noise of 44 and 47 dB was achieved in two resonator designs. Prior results showed that eddy currents contribute to background problems in rapid scan EPR, so the CLR design had to minimize conducting metal components. Using fine (AWG 38) wire for the resonators decreased eddy currents and lowered the resonator Q, thus providing larger resonator bandwidth. Mechanical resonances at specific scan frequencies are a major contributor to rapid scan backgrounds.

Entities:  

Year:  2010        PMID: 21603086      PMCID: PMC3098462          DOI: 10.1002/cmr.b.20161

Source DB:  PubMed          Journal:  Concepts Magn Reson Part B Magn Reson Eng        ISSN: 1552-5031            Impact factor:   1.176


  3 in total

1.  An L-band crossed-loop (Bimodal) EPR resonator.

Authors:  G A Rinard; R W Quine; G R Eaton
Journal:  J Magn Reson       Date:  2000-05       Impact factor: 2.229

2.  Comparison of maximum entropy and filtered back-projection methods to reconstruct rapid-scan EPR images.

Authors:  Mark Tseitlin; Amarjot Dhami; Sandra S Eaton; Gareth R Eaton
Journal:  J Magn Reson       Date:  2006-10-27       Impact factor: 2.229

3.  Background removal procedure for rapid scan EPR.

Authors:  Mark Tseitlin; Tomasz Czechowski; Richard W Quine; Sandra S Eaton; Gareth R Eaton
Journal:  J Magn Reson       Date:  2008-10-15       Impact factor: 2.229

  3 in total
  17 in total

1.  Rapid-Scan EPR of Nitroxide Spin Labels and Semiquinones.

Authors:  Sandra S Eaton; Gareth R Eaton
Journal:  Methods Enzymol       Date:  2015-08-01       Impact factor: 1.600

2.  Modular imaging system: Rapid scan EPR at 800 MHz.

Authors:  Oxana Tseytlin; Priyaankadevi Guggilapu; Andrey A Bobko; Hussien AlAhmad; Xuan Xu; Boris Epel; Ryan O'Connell; Emily H Hoblitzell; Timothy D Eubank; Valery V Khramtsov; Benoit Driesschaert; Eiad Kazkaz; Mark Tseytlin
Journal:  J Magn Reson       Date:  2019-06-08       Impact factor: 2.229

3.  Use of the Frank sequence in pulsed EPR.

Authors:  Mark Tseitlin; Richard W Quine; Sandra S Eaton; Gareth R Eaton; Howard J Halpern; J-H Ardenkjaer-Larsen
Journal:  J Magn Reson       Date:  2011-02-03       Impact factor: 2.229

4.  An X-Band Crossed-Loop EPR Resonator.

Authors:  George A Rinard; Richard W Quine; Joseph McPeak; Laura Buchanan; Sandra S Eaton; Gareth R Eaton
Journal:  Appl Magn Reson       Date:  2017-09-14       Impact factor: 0.831

5.  Imaging thiol redox status in murine tumors in vivo with rapid-scan electron paramagnetic resonance.

Authors:  Boris Epel; Subramanian V Sundramoorthy; Martyna Krzykawska-Serda; Matthew C Maggio; Mark Tseytlin; Gareth R Eaton; Sandra S Eaton; Gerald M Rosen; Joseph P Y Kao; Howard J Halpern
Journal:  J Magn Reson       Date:  2016-12-31       Impact factor: 2.229

Review 6.  Exchange Phenomena in the Electron Paramagnetic Resonance Spectra of the Nitroxyl and Trityl Radicals: Multifunctional Spectroscopy and Imaging of Local Chemical Microenvironment.

Authors:  Valery V Khramtsov; Andrey A Bobko; Mark Tseytlin; Benoit Driesschaert
Journal:  Anal Chem       Date:  2017-04-10       Impact factor: 6.986

7.  Resonators for In Vivo Imaging: Practical Experience.

Authors:  George A Rinard; Richard W Quine; Laura A Buchanan; Sandra S Eaton; Gareth R Eaton; Boris Epel; Subramanian V Sundramoorthy; Howard J Halpern
Journal:  Appl Magn Reson       Date:  2017-09-22       Impact factor: 0.831

8.  Imaging of nitroxides at 250MHz using rapid-scan electron paramagnetic resonance.

Authors:  Joshua R Biller; Mark Tseitlin; Richard W Quine; George A Rinard; Hilary A Weismiller; Hanan Elajaili; Gerald M Rosen; Joseph P Y Kao; Sandra S Eaton; Gareth R Eaton
Journal:  J Magn Reson       Date:  2014-03-01       Impact factor: 2.229

9.  Corrections for sinusoidal background and non-orthogonality of signal channels in sinusoidal rapid magnetic field scans.

Authors:  Mark Tseitlin; Deborah G Mitchell; Sandra S Eaton; Gareth R Eaton
Journal:  J Magn Reson       Date:  2012-08-08       Impact factor: 2.229

10.  Decoupling of excitation and receive coils in pulsed magnetic resonance using sinusoidal magnetic field modulation.

Authors:  Mark Tseytlin; Boris Epel; Subramanian Sundramoorthy; Dmitriy Tipikin; Howard J Halpern
Journal:  J Magn Reson       Date:  2016-09-07       Impact factor: 2.229

View more

北京卡尤迪生物科技股份有限公司 © 2022-2023.