| Literature DB >> 10783276 |
G A Rinard1, R W Quine, G R Eaton.
Abstract
Our crossed-loop resonator design has been enhanced to increase the filling factor and has been extended from S-band to L-band. High isolation between the two modes results in shorter dead time in pulsed EPR experiments than would occur with a reflection resonator of the same Q. Copyright 2000 Academic Press.Mesh:
Year: 2000 PMID: 10783276 DOI: 10.1006/jmre.2000.2014