Dirk Bandorski1, Martin Keuchel, Martin Brück, Reinhard Hoeltgen, Marcus Wieczorek, Ralf Jakobs.
Abstract
Background and Study Aims. Capsule endoscopy is an established tool for investigation of the small intestine. Because of limited clinical experience in patients with cardiac devices, the Food and Drug Administration and the manufacturer recommended not to use capsule endoscopy in these patients. The vast majority of investigations did not reveal any interference between capsule endoscopy and cardiac devices. Methods. Studies investigating interference between CE and cardiac devices were analysed. For the review we considered studies published in English or German and indexed in Medline, as well as highly relevant abstracts. Results. In vitro and in vivo studies mainly revealed no interference between capsule endoscopy and cardiac devices. Technical data of capsule endoscopy (Given Imaging) reveal that interference with cardiac pacemakers and implantable cardioverter defibrillator is impossible. Telemetry can interfere with CE video. Conclusion. The clinical use of capsule endoscopy (Given Imaging) is unproblematic in patients with cardiac pacemakers.Entities:
Year: 2011 PMID: 21603020 PMCID: PMC3095403 DOI: 10.1155/2011/376053
Source DB: PubMed Journal: Diagn Ther Endosc ISSN: 1026-714X
Studies investigating patients with cardiac pacemakers who underwent capsule endoscopy.
| Author | Year | Number of patients/cardiac pacemakers ( | Brand of cardiac pacemaker | Kind of study | Interference | Brand of CE |
|---|---|---|---|---|---|---|
| Dirks et al. | 2008 | 5 | No specification | In vivo | No | Given Imaging |
| Bandorski et al. | 2008 | 21 | Medtronic, Osypka, Siemens, Vitatron, Ela, Guidant, St. Jude Medical | In vitro | No | Given Imaging + Olympus |
| Bandorski et al. | 2006 | 1 | Biotronik | In vitro | No | Given Imaging |
| Payeras et al. | 2005 | 20 | No specification | In vitro + In vivo | No | Given Imaging + Test Cap |
| Bandorski et al. | 2005 | 45 | No specification | In vivo | No | Given Imaging |
| Dubner et al. | 2005 | 100 | St. Jude Medical, Medtronic, Guidant, Biotronik, Sorin | In vivo | Yes ( | Given Imaging |
| Guyomar et al. | 2004 | 1 | ELA | In vivo | No | Given Imaging |
| Leighton et al. | 2004 | 5 | No specification | In vivo | No | Given Imaging |
Studies investigating patients with implantable cardioverter defibrillators who underwent capsule endoscopy.
| Author | Year | Number of patients/ICD ( | Brand of ICD | Kind of study | Interference | Brand of CE |
|---|---|---|---|---|---|---|
| Bandorski et al. | 2009 | 45 | Biotronik, Guidant, Medtronic, St. Jude Medical | In vivo | No | Given Imaging + Olympus |
| Dubner et al. | 2008 | 6 (In vitro) | Medtronic, St. Jude Medical, Guidant, Biotronik | In vitro + In vivo | Yes | Given Imaging |
| Pelargino et al. | 2005 | 1 | Medtronic | In vivo | No | Given Imaging |
| Bandorski et al. | 2005 | 8 | No specification | In vivo | No | Given Imaging |
| Leighton et al. | 2005 | 5 | Guidant, Medtronic, St. Jude Medical | In vivo | No | Given Imaging |