Electromagnetic interference from wireless video-capsule endoscopy on implantable cardioverter-defibrillators.

Sergio Dubner1, Yael Dubner, Horacio Rubio, Eran Goldin.   

Abstract

OBJECTIVES: The purpose of this study was to evaluate potential interference between the Pill-cam video-capsule and implanted cardio-defibrillators (ICD) in both in vitro and in vivo test environments.
METHODS: Phase I consisted of in vitro testing utilized 6 ICD models. Each was placed in a saline gel bath in conjunction with a tool designed to emit the same wave length as the Pill-Cam (the Test Cap). Tests were performed at both the manufacturer's nominal and most sensitive settings with the test probe placed at 1, 5, 10, and 15 cm from three different points. There were emissions of 10-, 30-, and 60- second duration at each location. Phase II was the in vivo study utilizing patients with implanted ICDs for standard clinical reasons who were undergoing evaluation with the Test Cap.
RESULTS: In Phase I, 864 tests were performed involving the 6 ICDs. There was totally normal behavior in 5 of the devices. The Biotronik Belos (Berlin, Germany) demonstrated oversensing and delivery of inappropriate therapy. In phase II, 6 patients underwent a total number of 288 tests (48 in each patient). There were 4 men and 2 women, mean age 61 (33-77) and none demonstrated any adverse interactions.
CONCLUSIONS: In patients with devices identical to those evaluated that showed no interference, the use of the Pill-Cam is safe. However, in patients with a Biotronik Belos ICD (or any other nontested ICD) the use of the Pill-Cam should be done only in-hospital, after suspended ICD therapy and in conjunction with close monitoring.

Entities:  

Mesh:

Year:  2007        PMID: 17437569     DOI: 10.1111/j.1540-8159.2007.00695.x

Source DB:  PubMed          Journal:  Pacing Clin Electrophysiol        ISSN: 0147-8389            Impact factor:   1.976


  11 in total

1.  Safety of capsule endoscopy using human body communication in patients with cardiac devices.

Authors:  Joo Won Chung; Hye Jin Hwang; Moon Jae Chung; Jeong Youp Park; Hui-Nam Pak; Si Young Song
Journal:  Dig Dis Sci       Date:  2012-02-07       Impact factor: 3.199

Review 2.  What defines quality in small bowel capsule endoscopy.

Authors:  Ioannis V Mitselos; Dimitrios K Christodoulou
Journal:  Ann Transl Med       Date:  2018-07

3.  Lack of interference between small bowel capsule endoscopy and implantable cardiac defibrillators: an 'in vivo' electrophysiological study.

Authors:  Dario Moneghini; Alessandro Lipari; Guido Missale; Luigi Minelli; Gianpaolo Cengia; Luca Bontempi; Antonio Curnis; Renzo Cestari
Journal:  United European Gastroenterol J       Date:  2015-09-24       Impact factor: 4.623

Review 4.  Small bowel capsule endoscopy: Where are we after almost 15 years of use?

Authors:  Cedric Van de Bruaene; Danny De Looze; Pieter Hindryckx
Journal:  World J Gastrointest Endosc       Date:  2015-01-16

5.  Effects of external electrical and magnetic fields on pacemakers and defibrillators: from engineering principles to clinical practice.

Authors:  Roy Beinart; Saman Nazarian
Journal:  Circulation       Date:  2013-12-24       Impact factor: 29.690

6.  Capsule endoscopy in patients with cardiac pacemakers, implantable cardioverter defibrillators, and left heart devices: a review of the current literature.

Authors:  Dirk Bandorski; Martin Keuchel; Martin Brück; Reinhard Hoeltgen; Marcus Wieczorek; Ralf Jakobs
Journal:  Diagn Ther Endosc       Date:  2011-04-27

7.  Capsule Endoscopy in Patients with Cardiac Pacemakers and Implantable Cardioverter Defibrillators: (Re)evaluation of the Current State in Germany, Austria, and Switzerland 2010.

Authors:  Dirk Bandorski; Ralf Jakobs; Martin Brück; Reinhard Hoeltgen; Marcus Wieczorek; Martin Keuchel
Journal:  Gastroenterol Res Pract       Date:  2011-12-29       Impact factor: 2.260

Review 8.  Capsule endoscopy in patients with cardiac pacemakers, implantable cardioverter defibrillators and left heart assist devices.

Authors:  Dirk Bandorski; Reinhard Höltgen; Dominik Stunder; Martin Keuchel
Journal:  Ann Gastroenterol       Date:  2014

Review 9.  Contraindications for video capsule endoscopy.

Authors:  Dirk Bandorski; Niehls Kurniawan; Peter Baltes; Reinhard Hoeltgen; Matthias Hecker; Dominik Stunder; Martin Keuchel
Journal:  World J Gastroenterol       Date:  2016-12-07       Impact factor: 5.742

10.  Capsule Endoscopy in a Patient with an Implanted CCM System and an Implantable Defibrillator.

Authors:  Florian Streitner; Nina Schoene; Martin Borggrefe; Jürgen Kuschyk
Journal:  Case Rep Cardiol       Date:  2011-07-27
View more

北京卡尤迪生物科技股份有限公司 © 2022-2023.