Julia Stanley1, Karen Breitman, Peter Dunscombe, David P Spencer, Harold Lau.
Abstract
A database of clinically approved stereotactic radiosurgery treatment plans was created. One hundred and seventy targets in the database were then retrospectively evaluated using conformity indices suggested by RTOG, SALT-Lomax and Paddick. Relationships between the three alternative conformity indices were determined. The Paddick index combines the information provided by the RTOG and SALT-Lomax indices into a single index. The variation in the geometric overlap ratio, which is related to the SALT-Lomax index, was found to be not clinically relevant for our cohort of patients, and thus the Paddick and RTOG indices can be directly related. It was found that access to a dose volume histogram or dose distribution for a treatment plan renders the RTOG conformity index sufficient for plan quality evaluation.Entities:
Mesh:
Year: 2011 PMID: 21587193 PMCID: PMC5718686 DOI: 10.1120/jacmp.v12i2.3449
Source DB: PubMed Journal: J Appl Clin Med Phys ISSN: 1526-9914 Impact factor: 2.102
Distribution of target volumes.
|
|
|
|---|---|
| 0 to | 79 |
| 1 to | 27 |
| 2 to | 21 |
| 3 to | 14 |
| 4 to | 5 |
| 5 to | 14 |
|
| 10 |
Distribution of characteristics of the 105 patients in the database.
|
|
| |
|---|---|---|
| Age Median (Range) |
| |
| Sex Male:Female | 39:66 | |
| Primary Tumor Site | Breast | 33 |
| Lung | 46 | |
| Renal | 5 | |
| Melanoma | 9 | |
| Other | 12 | |
| Dose Prescription to 80% isodose for 170 targets |
| 22 |
|
| 123 | |
|
| 25 | |
| Number of targets per treatment | One target | 72 |
| (8 patients treated twice) | Two targets | 27 |
| Three targets | 12 | |
|
| 2 | |
Figure 1RTOG conformity index versus target volume. Points that fall within the horizontal lines at 1 and 2 are defined as being within protocol.
Figure 2RTOG quality of coverage versus target volume.
Figure 3RTOG homogeneity index versus target volume. Points that fall below the horizontal line are defined as being within protocol.
Figure 4Paddick conformity index versus target volume.
Figure 5Relationship between Paddick and RTOG conformity indices.
Geometric ratio overlap occurrences.
|
|
|
|---|---|
|
| 1 |
| 0.75 to 0.8 | 1 |
| 0.8 to 0.85 | 3 |
| 0.85 to 0.9 | 2 |
| 0.9 to 0.95 | 12 |
|
| 151 |
Figure 6An example where the geometric overlap ratio is less than one due to an adjacent critical structure. The target volume is outlined in orange and the occipital cortex is outlined in purple.
Figure 7SALT‐Lomax conformity index versus target volume.