Placement and removal torque values of orthodontic miniscrew implants.

Eduardo Yugo Suzuki1, Boonsiva Suzuki.   

Abstract

INTRODUCTION: The purpose of this study was to analyze the maximum insertion torque (MIT) and maximum removal torque (MRT) values of orthodontic miniscrews.
METHODS: Two hundred eighty titanium miniscrews were placed in several sites in the maxillae and mandibles of orthodontic patients to provide skeletal anchorage. Two types of miniscrews were used: predrilling, cylindrical miniscrews, and self-drilling, conical miniscrews. A force of 50 g was applied with nickel-titanium coil springs 2 weeks after placement. MIT and MRT values were assessed with a digital torque gauge. Torque values were subjected to the paired t test, Student t test, chi-square test, and 1-way analysis of variance (ANOVA) (P <0.05).
RESULTS: MIT values were significantly higher for the self-drilling miniscrews (mean, 14.5 Ncm; 95% CI, 13.6-15.3) than for the predrilling miniscrews (mean, 9.2 Ncm; 95% CI, 8.6-9.9) in all implant sites. For both predrilling and self-drilling miniscrews, the highest MIT values were observed at the midpalatal suture site followed by the dentoalveolar bones of the mandible and maxilla, respectively. In contrast, MRT values were significantly higher for the predrilling miniscrews (mean, 22.6 Ncm; 95% CI, 22.0-23.3) than for the self-drilling miniscrews (mean, 17.6 Ncm; 95% CI, 16.8-18.4). Accordingly, the mean torque ratio of predrilling miniscrews was twice that of the self-drilling miniscrews.
CONCLUSIONS: An inverse relationship between MIT and MRT values was observed. The results suggest that relatively lower MIT values were more favorable to osseointegration than higher values.
Copyright © 2011 American Association of Orthodontists. Published by Mosby, Inc. All rights reserved.

Entities:  

Mesh:

Substances:

Year:  2011        PMID: 21536211     DOI: 10.1016/j.ajodo.2010.11.017

Source DB:  PubMed          Journal:  Am J Orthod Dentofacial Orthop        ISSN: 0889-5406            Impact factor:   2.650


  25 in total

1.  Surgical management of fractured orthodontic mini- implant- a case report.

Authors:  Manthan Desai; Anoop Jain; Nida Sumra
Journal:  J Clin Diagn Res       Date:  2015-01-01

2.  Accuracy of torque-limiting devices for mini-implant removal: an in vitro study.

Authors:  A Pauls; M Nienkemper; D Drescher
Journal:  J Orofac Orthop       Date:  2013-05-08       Impact factor: 1.938

3.  Experimental investigation of the fracture torque of orthodontic anchorage screws.

Authors:  Susanne Reimann; Mustafa Ayubi; Fraser McDonald; Christoph Bourauel
Journal:  J Orofac Orthop       Date:  2016-04-20       Impact factor: 1.938

4.  Patient and miniscrew implant factors influence the success of orthodontic miniscrew implants.

Authors:  Kelvin I Afrashtehfar
Journal:  Evid Based Dent       Date:  2016-12

5.  Insertion torque values and success rates for paramedian insertion of orthodontic mini-implants : A retrospective study.

Authors:  Bruno Di Leonardo; Björn Ludwig; Jörg Alexander Lisson; Luca Contardo; Rossano Mura; Jan Hourfar
Journal:  J Orofac Orthop       Date:  2018-02-20       Impact factor: 1.938

6.  Stability of smooth and rough mini-implants: clinical and biomechanical evaluation - an in vivostudy.

Authors:  Giselle Naback Lemes Vilani; Antônio Carlos de Oliveira Ruellas; Carlos Nelson Elias; Cláudia Trindade Mattos
Journal:  Dental Press J Orthod       Date:  2015-10

7.  Effects of predrilling depth and implant shape on the mechanical properties of orthodontic mini-implants during the insertion procedure.

Authors:  Keun-Chul Cho; Seung-Hak Baek
Journal:  Angle Orthod       Date:  2011-11-03       Impact factor: 2.079

8.  Angled-predrilling depth and mini-implant shape effects on the mechanical properties of self-drilling orthodontic mini-implants during the angled insertion procedure.

Authors:  Yoon-Young Heo; Keun-Chul Cho; Seung-Hak Baek
Journal:  Angle Orthod       Date:  2012-03-03       Impact factor: 2.079

9.  Effect of vertical placement angle on the insertion torque of mini-implants in human alveolar bone.

Authors:  Rafael Ribeiro Maya; Célia Regina Maio Pinzan-Vercelino; Julio de Araujo Gurgel
Journal:  Dental Press J Orthod       Date:  2016 Sep-Oct

10.  Effect of surface treatment on the mechanical stability of orthodontic miniscrews.

Authors:  Yousef Al-Thomali; Sakeenabi Basha; Roshan Noor Mohamed
Journal:  Angle Orthod       Date:  2022-01-01       Impact factor: 2.079

View more

北京卡尤迪生物科技股份有限公司 © 2022-2023.