Literature DB >> 21498340

Clinical experience of photon counting breast tomosynthesis: comparison with traditional mammography.

Gunilla Svane1, Edward Azavedo, Karin Lindman, Mattias Urech, Jonas Nilsson, Niclas Weber, Lars Lindqvist, Christer Ullberg.   

Abstract

BACKGROUND: In two-dimensional mammography, a well-known problem is over- and underlying tissue which can either obstruct a lesion or create a false-positive result. Tomosynthesis, with an ability to layer the tissue in the image, has the potential to resolve these issues.
PURPOSE: To compare the diagnostic quality, sensitivity and specificity of a single tomosynthesis mammography image and a traditional two-view set of two-dimensional mammograms and to assess the comfort of the two techniques.
MATERIAL AND METHODS: One hundred and forty-four women, mainly chosen because of suspicious features on standard mammograms (76 malignant), had a single tomosynthesis image taken of one breast using a novel photon counting system. On average, the dose of the tomosynthesis images was 0.63 times that of the two-view images and the compression force during the procedure was halved. The resulting images were viewed by two radiologists and assessed both individually and comparing the two techniques.
RESULTS: In 56% of the cases the radiologists rated the diagnostic quality of the lesion details higher in the tomosynthesis images than in the conventional images (and in 91% equal or higher), which means there is a statistically significant preference for the tomosynthesis technique. This included the calcifications which were rated as having better quality in 41% of the cases. While sensitivity was slightly higher for traditional mammography the specificity was higher for tomosynthesis. However, neither of these two differences was large enough to be statistically significant.
CONCLUSION: The overall accuracy of the two techniques was virtually equal despite the radiologist's very limited experience with tomosynthesis images and vast experience with two-dimensional mammography. As the diagnostic quality of the lesion details in the tomosynthesis images was valued considerably higher this factor should improve with experience. The patients also favored the tomosynthesis examination, rating the comfort of the procedure as much higher than regular mammography which might affect screening attendance.

Entities:  

Mesh:

Year:  2011        PMID: 21498340     DOI: 10.1258/ar.2010.100262

Source DB:  PubMed          Journal:  Acta Radiol        ISSN: 0284-1851            Impact factor:   1.990


  17 in total

1.  Digital Breast Tomosynthesis: State of the Art.

Authors:  Srinivasan Vedantham; Andrew Karellas; Gopal R Vijayaraghavan; Daniel B Kopans
Journal:  Radiology       Date:  2015-12       Impact factor: 11.105

2.  Diagnostic accuracy of digital breast tomosynthesis versus digital mammography for benign and malignant lesions in breasts: a meta-analysis.

Authors:  Junqiang Lei; Pin Yang; Li Zhang; Yinzhong Wang; Kehu Yang
Journal:  Eur Radiol       Date:  2014-03       Impact factor: 5.315

3.  Digital breast tomosynthesis within a symptomatic "one-stop breast clinic" for characterization of subtle findings.

Authors:  G J Bansal; P Young
Journal:  Br J Radiol       Date:  2015-07-02       Impact factor: 3.039

Review 4.  A review of breast tomosynthesis. Part II. Image reconstruction, processing and analysis, and advanced applications.

Authors:  Ioannis Sechopoulos
Journal:  Med Phys       Date:  2013-01       Impact factor: 4.071

5.  Digital breast tomosynthesis versus mammography and breast ultrasound: a multireader performance study.

Authors:  Fabienne Thibault; Clarisse Dromain; Catherine Breucq; Corinne S Balleyguier; Caroline Malhaire; Luc Steyaert; Anne Tardivon; Enrica Baldan; Harir Drevon
Journal:  Eur Radiol       Date:  2013-05-15       Impact factor: 5.315

6.  Letter to the Editor re: Diagnostic accuracy of digital breast tomosynthesis versus digital mammography for benign and malignant lesions in breasts: a meta-analysis.

Authors:  Tony Martin Svahn
Journal:  Eur Radiol       Date:  2014-01-24       Impact factor: 5.315

7.  Reply to Letter to the Editor re: Diagnostic accuracy of digital breast tomosynthesis versus digital mammography for benign and malignant lesions in breasts: a meta-analysis.

Authors:  Junqiang Lei; Pin Yang; Li Zhang; Yinzhong Wang; Kehu Yang
Journal:  Eur Radiol       Date:  2014-01-18       Impact factor: 5.315

Review 8.  Main controversies in breast cancer.

Authors:  Stephane Zervoudis; George Iatrakis; Eirini Tomara; Anastasia Bothou; George Papadopoulos; George Tsakiris
Journal:  World J Clin Oncol       Date:  2014-08-10

Review 9.  Review of radiation dose estimates in digital breast tomosynthesis relative to those in two-view full-field digital mammography.

Authors:  T M Svahn; N Houssami; I Sechopoulos; S Mattsson
Journal:  Breast       Date:  2014-12-29       Impact factor: 4.380

10.  Comparison of radiation dose, workflow, patient comfort and financial break-even of standard digital radiography and a novel biplanar low-dose X-ray system for upright full-length lower limb and whole spine radiography.

Authors:  Tobias J Dietrich; Christian W A Pfirrmann; Alexander Schwab; Katja Pankalla; Florian M Buck
Journal:  Skeletal Radiol       Date:  2013-03-28       Impact factor: 2.199