The Shack-Hartmann wavefront sensor (SHWS) spots upon which ocular aberration measurements depend have poor quality in mice due to light reflected from multiple retinal layers. We have designed and implemented a SHWS that can favor light from a specific retinal layer and measured monochromatic aberrations in 20 eyes from 10 anesthetized C57BL/6J mice. Using this instrument, we show that mice are myopic, not hyperopic as is frequently reported. We have also measured longitudinal chromatic aberration (LCA) of the mouse eye and found that it follows predictions of the water-filled schematic mouse eye. Results indicate that the optical quality of the mouse eye assessed by measurement of its aberrations is remarkably good, better for retinal imaging than the human eye. The dilated mouse eye has a much larger numerical aperture (NA) than that of the dilated human eye (0.5 NA vs. 0.2 NA), but it has a similar amount of root mean square (RMS) higher order aberrations compared to the dilated human eye. These measurements predict that adaptive optics based on this method of wavefront sensing will provide improvements in retinal image quality and potentially two times higher lateral resolution than that in the human eye.
The Shack-Hartmann wavefront sensor (SHWS) spots upon which ocular aberration measurements depend have poor quality in mice due to light reflected from multiple retinal layers. We have designed and implemented a SHWS that can favor light from a specific retinal layer and measured monochromatic aberrations in 20 eyes from 10 anesthetized C57BL/6J mice. Using this instrument, we show that mice are myopic, not hyperopic as is frequently reported. We have also measured longitudinal chromatic aberration (LCA) of the mouse eye and found that it follows predictions of the water-filled schematic mouse eye. Results indicate that the optical quality of the mouse eye assessed by measurement of its aberrations is remarkably good, better for retinal imaging than the human eye. The dilated mouse eye has a much larger numerical aperture (NA) than that of the dilated human eye (0.5 NA vs. 0.2 NA), but it has a similar amount of root mean square (RMS) higher order aberrations compared to the dilated human eye. These measurements predict that adaptive optics based on this method of wavefront sensing will provide improvements in retinal image quality and potentially two times higher lateral resolution than that in the human eye.
Entities:
Keywords:
(110.1080) Active or adaptive optics; (170.4460) Medical optics and biotechnology: Ophthalmic optics and devices; (330.4300) Vision system - noninvasive assessment; (330.5370) Vision, color, and visual optics: Physiological optics; (330.7324) Vision, color, and visual optics: Visual optics, comparative animal models
The mouse is frequently adopted as the model system for investigating the normal and diseased
eye because it is inexpensive, reproduces and matures rapidly, and is amenable to genetic
manipulations that are not yet possible in other mammals. Despite its widespread use in vision
research, relatively little quantitative information is available about the optics of the mouse
eye. A few schematic eyes for the mouse have been proposed [1,2]. These incorporate the large spherical
crystalline lens that fills most of the ocular volume and an axial length of about 3.3 mm, 8
times smaller than that of the human eye. Consistent with the mouse’s nocturnal
habit, its eye has a numerical aperture (NA) of about 0.49 when the pupil is fully dilated,
about two times greater than that of the human eye.Knowledge of the optical aberrations of the mouse eye is important not only for understanding
the visual capabilities of the mouse, but also because correcting these aberrations can improve
the quality of in vivo images of the mouse retina. In vivo
imaging offers the advantage over post mortem histology of tracking in individual animals
developmental changes, disease progression, or the efficacy of therapy. This can reduce the
number of animals required and the animal to animal variability inherent in constructing the
dynamics of a process from different animals sacrificed at different time points. The optical
aberrations of the mouse eye can be corrected with adaptive optics (AO), which has enabled
diffraction-limited imaging of cellular and sub-cellular structures in the living human and
primate eyes [3-6]. Recently, a few pilot studies have demonstrated that some of the benefits of AO
found in primates can be realized in rodent eyes [7-9]. However, most investigations
that have attempted to measure the optical quality of the rodent eye have concluded either that
it is poor or have encountered technical difficulties in measuring it. One study found very poor
retinal image quality in both the rat and the mouse using the double pass method [10]. With the development of wavefront sensing, it is now
possible to directly measure aberrations in rodent eyes. Irving et al. and Bird et al. showed
distorted spots and qualitative wave aberration measurements in rat eyes using a Shack-Hartmann
wavefront sensor (SHWS) [11,12]. A recent study measured wave aberrations up to the 4th order Zernikes
over a 1.5 mm pupil in the mouse eye, showing clearly more degraded Shack-Hartman spots than
that of the typical human eye [13].Biss et al. found it difficult to measure the wavefront using back-scattered light due to very
large wavefront sensor spots [7]. To avoid this poor spot
quality, they used the fluorescence signal from a single layer of labeled retinal cells rather
than back-scattered light for wavefront sensing. This is a clever solution, but it has the
limitations that the signal available for wavefront sensing is reduced by an order of magnitude
or more and requires the use of intrinsic or extrinsic fluorophores, which may not be available
or feasible depending on the experiment. In the rat eye, we observed similar difficulties with
backscattered light but found that the wavefront sensor spots were good enough to correct
aberrations with an AO control loop, enabling sub-cellular imaging of ganglion cell axons and
dendrites [8].The inferior SHWS spot quality in rodent eyes is caused by the fact that the wavefront sensing
beacon is reflected from multiple layers of the optically thick retina. This phenomenon is
similar to the elongation of wavefront sensor spots from artificial guide stars in astronomy. In
that case, elongation is caused by the thickness of the sodium layer in the atmosphere that is
illuminated with a ground-based laser to produce the wavefront sensor beacon [14]. In the mouse eye (see Fig. 1
), the SHWS spots are often elongated radially, with the axis of elongation typically
pointing toward the point where the laser beacon enters the pupil of the eye. Figure 1 shows an image of the lenslet array superimposed on
the eye’s pupil to illustrate this point using only a single line of lenslets. When
the light from the beacon is reflected from the retina (shown in red), all the lenslets imaging
the illuminated part of retina off-axis with respect to the illuminating beam will produce a
radially elongated spot image, with increasing amounts of elongation the further off-axis the
lenslet happens to be. If there are two layers that have the brightest back reflections from the
retina (shown in blue), as is often the case to a first approximation, a double spot pattern
will appear in the raw wavefront image. The spot arising from the more posterior of the two
surfaces lies closer to the illumination axis than the spot arising from the more anterior
surface. This double spot pattern will be discussed in more detail later.
Fig. 1
Schematic explaining the elongation of wavefront sensor spots in the mouse eye.
θ: elongation angle. n: refractive index of the retina. Note that the lenslets are
in a plane conjugate to the eye’s pupil, and not physically on it.
Schematic explaining the elongation of wavefront sensor spots in the mouse eye.
θ: elongation angle. n: refractive index of the retina. Note that the lenslets are
in a plane conjugate to the eye’s pupil, and not physically on it.For a lenslet at the edge of the pupil, the amount of elongation of its wavefront sensor spot
can be described by an elongation angle θ, which can be approximated below in Eq. (1). Here NA is the
eye’s numerical aperture, z is the largest axial separation between
reflecting layers or the retinal thickness, and f is the focal length of the
eye.If there is no magnification factor between the eye’s pupil and the pupil plane of
the wavefront sensor, then the angle of elongation for the wavefront sensor spot on the edge of
the pupil is the same as θ. If the eye’s pupil is magnified to the
wavefront sensor pupil, the largest angle of elongation on the wavefront sensor becomes:where is the eye’s pupil size and is the pupil size of the wavefront sensor.Table 1
summarizes a few example calculations made for the elongation angle of a mouse eye
versus a human eye. Equation (1) makes it
immediately clear that there are at least two reasons why the wavefront sensor spot quality is
so much poorer in the mouse eye than in the human. First, the mouse eye has a larger NA (0.49
for a 2 mm pupil vs. 0.18 for a human 6 mm pupil). Second, it has a larger retinal thickness
when expressed relative to the eye’s focal length, z/f. It is approximately 9% for
the mouse vs. 1.7% for the thickest part of the human retina, and 0.6% for the thinnest part of
the human retina. Thus the elongation angle θ in the mouse retina is 15 times larger
than that of the human extrafoveal retina at its thickest point, and over 40 times larger than
that of the human foveal center.
Table 1
Elongation angle calculated for the mouse eye vs. the human eye*
Retina
Pupil size
DeDw
NA
f
z
z/f
θ'
θ'Mouseθ'Human
Mouse
2 mm
1
0.49
2.6 mm
230 µm
9%
2.51°
15–42
average
0.33
0.84°
5–14
Human thinnest
6 mm
1
0.18
22.3 mm
130 µm
0.6%
0.06°
Human thickest
370 µm
1.7%
0.17°
*Mouse eye f and z are values for the adult mouse eye
[2]. Human eye f is from the
Gullstrand-Le Grand model and z is from Polyak [15]. Values for are example values.
*Mouse eye f and z are values for the adult mouse eye
[2]. Human eye f is from the
Gullstrand-Le Grand model and z is from Polyak [15]. Values for are example values.If there is magnification between the eye’s and the wavefront sensor’s
pupil, then Eq. (2) is used to calculate the
elongation angle on the wavefront sensor spots. For example, if both the human pupil and the
mouse pupil are imaged onto the same human pupil size on the wavefront sensor, is the same for both species. Then the elongation angle calculated using the same parameters in the previous paragraph will be 5 to 14
times larger in the mouse eye than in the human eye.The elongation of the wavefront sensor spots will introduce error into the estimate of the
wave aberration to the extent that it displaces the centroid computed for each spot. If the only
difference between the light reflected from different retinal layers was a difference in axial
position, one might expect that their only effect would be on the defocus term of the wave
aberration. However, in our experience, light from each retinal layer can produce its own
idiosyncratic light distribution in the pupil. This means that light from a particular layer may
have more effect on the centroid in one part of the pupil than another, further complicating the
irregularities in the wavefront sensor spots across the pupil, and introducing spurious
aberrations in the wavefront sensor estimate that do not exist in the anterior optics of the
eye. By capitalizing on a few simple optical principles we can mitigate the influence on
wavefront sensing of light backscattered from multiple layers. This way we have measured the
wave aberration in the mouse eye over a fully dilated pupil up to 10th order Zernike polynomials
and have explored the implications of these aberrations especially for microscopic imaging of
the retina in the living eye.
2. Methods
2.1 Subjects
Ten black C57BL/6J wild type mice from 2 months to 5 months of age were used in wave
aberration measurement experiments, and the same mice from 7 months to 9 months of age were
used in the chromatic aberration study. Mice were housed in standard mouse cages under 12 hour
light/dark cycle. All animals were handled according to the Association for Research in Vision
and Ophthalmology Statement for the Use of Animals in Ophthalmic and Vision Research and to the
guidelines of the University Committee on Animal Resources at the University of Rochester.
2.2 The Shack-Hartmann wavefront sensor for the mouse eye
A SHWS was constructed for the mouse eye and the schematic layout of the system is shown in
Fig. 2
. A single mode fiber-coupled 789 nm superluminescent diode (SLD) with 11.5 nm bandwidth
(InPhenix, Livermore, CA USA) was used as the wavefront sensing light source. The
SLD’s fiber connector was mounted on a translation stage so that the light source
could be focused at different planes in the retina. The wavefront sensor consisted of a lenslet
array (Adaptive Optics Associates, Cambridge, MA USA) with 18 mm focal length and 325
μm lenslet pitch, and a 1004x1004 pixel CCD camera (Cohu, San Diego, CA USA). Wave
aberrations were measured using back scattered light at 10Hz.
Fig. 2
Schematic of the mouse eye wavefront sensor. SLD: fiber coupled Super Luminescent Diode.
SH Wavefront Sensor: Shack-Hartmann wavefront sensor. BS: Beam Splitter. RBS: Removable Beam
Splitter. FM: Fold Mirror. M1-6: Concave spherical Mirrors.
Schematic of the mouse eye wavefront sensor. SLD: fiber coupled Super Luminescent Diode.
SH Wavefront Sensor: Shack-Hartmann wavefront sensor. BS: Beam Splitter. RBS: Removable Beam
Splitter. FM: Fold Mirror. M1-6: Concave spherical Mirrors.The optical path consisted of three pupil planes (a wavefront sensor pupil plane, a plane for
inserting trial lenses, and a plane for a resonant scanner) conjugate to the eye’s
pupil using off axis spherical mirror 4-f telescopes. Defocus and astigmatism could be
corrected at the trial lens plane, but for this study it was not used when aberration data were
measured. Speckle in the wavefront sensing spot images was reduced by the wide bandwidth SLD
and was further reduced by the resonant scanner (Electro-Optical Products Corp, Glendale, NY
USA) which line-scanned the retina at 15 kHz [16]. In
comparison, the CCD camera exposure time typically used was 5 ms to 50 ms. The scanning angle
on the eye’s pupil was kept smaller than 0.5 degrees. To monitor the pupil
centration and the eye’s rotation, the mouse pupil was illuminated with an infrared
(IR) light emitting diode (LED) ring and imaged onto a CCD camera (The Imaging Source LLC.,
Charlotte, NC USA). A simple fundus camera consisting of an LED light source (Thorlabs Inc.,
Newton, NJ USA) and a CMOS camera (NET USA INC., Highland, IN USA) monitored the retinal
location during measurement.
2.3 Experiment design
Mice were anesthetized using a ketamine (~65 mg/kg) and xylazine (~8 mg/kg) cocktail and
covered by a heating pad during measurement. Pupils were dilated with one drop of 2.5%
phenylephrine (Neo-Synephrine) and one drop of 0.5% tropicamide.Mice were stabilized on a bitebar stage with two rotational degrees of freedom (Bioptigen,
Research Triangle Park, NC USA). The bitebar stage was mounted on a three-axis translation
stage to align the center of rotation of the bitebar stage with the exit pupil of the wavefront
sensor. The bite bar itself could be translated in two dimensions to align the mouse eye pupil
to the center of rotation of the stage. Then the stage could be rotated to measure different
retinal locations without misalignment of the pupil.The pupil camera provided easy alignment and monitoring of the pupil position. The bitebar
stage was rotated so that the first Purkinje image from the IR illuminator ring was centered on
the eye’s pupil. To measure along the optical axis, the optic disk was found with
the fundus camera. Wave aberration measurements were usually performed near the optic disk (at
4° temporal unless described otherwise). The cornea was periodically lubricated with
saline solution to prevent dehydration. After each saline drop, the excess was carefully
removed with a cotton swab, and 10 images were collected. The same procedure was repeated at
least 6 times, yielding over 60 images to average for each eye.
2.4 Wavefront spot image collection
In the current configuration, a total of 321 lenslets were used to sample a 2 mm pupil, and
44x44 CCD pixels were used for each lenslet. The Shack-Hartmann spot positions obtained with
the mouse eye were always referenced to spot positions obtained when the eye was replaced with
a point source in the back focal plane of a long focal length (140 mm EFL), high quality
achromatic doublet placed where the eye’s pupil would have been. The use of this
reference allowed the wave aberration measurements of the mouse eye to be unaffected by any
aberrations intrinsic to the wavefront sensor.For wavefront sensing in the human eye, especially in commercial instruments, a small
diameter illumination beam is typically used to increase the depth of focus, so that the
wavefront sensor spots are less susceptible to focus and aberration artifacts on the first pass
into the eye [17]. However, in the mouse eye, the
increased depth of focus provided by a small diameter beam exacerbates the problem of multiple
back reflections from multiple retinal layers. When a narrow beam is used as the illumination
beam, we often see two patterns of sharp spots superimposed. The two spot patterns follow a
similar radial elongation pattern shown in Fig. 1. Their
separations are largest at the edge of pupil and are superimposed at the point of beam entry.
To reduce the depth of focus, we used the largest diameter beacon we could, 2 mm outer diameter
which nearly filled the eye’s pupil. (The depth of focus is ~3 μm for a
diffraction-limited mouse eye with a 2 mm pupil size and 0.49 NA.) To avoid back reflections
from the first surface of the cornea and possibly from the posterior surface of the lens, an
annular beacon with 0.3 mm inner diameter was used. The beacon was then focused on the retinal
layer of interest, and due to the small depth of focus, light from distant unwanted retinal
layers was blurred thus reducing its influence on the centroid. The best focus of the beacon
could be found in either of two ways: subjectively by translating the beacon while observing
the sharpness change of the wavefront spots, or objectively by using an automated metric to
find the sharpest spots. In the current study we did this objectively by using a simple
sharpness metric [18]. To remove its dependence on the
overall spot intensity of the Shack-Hartmann image, we normalized the metric as shown in Eq. (3). This normalized sharpness metric
S was computed for the wavefront spot pattern in real time to help find the
sharpest spots which corresponded to the best focus position. It calculates the normalized
intensity square for all the pixels in the entire Shack-Hartmann spot image, and yields a
larger number for an image with a sharper overall spot pattern.where x, y are coordinates in the Shack-Hartmann spot image, and
I(x,y) is the spot intensity.
2.5 Wave aberration data analysis
As mentioned previously, spurious back reflections from multiple layers can make the
wavefront sensing spots quite asymmetric for the mouse eye. Therefore, in addition to
decreasing the depth of focus, we chose to process the wavefront sensor images in a way that
reduced the impact of asymmetries in each spot image on the aberration estimate. For
centroiding the wavefront spots, an iterative window size center of mass algorithm was used
[19]. To further diminish possible contributions of
light from unwanted retinal layers, the final window size was set at 1/4 of the
diffraction-limited spot Airy-disk size. This was done because measurements showed that
aberration coefficients reach an asymptotic level at window sizes at or smaller than ~1/4 of
the airy-disk size. This, along with the observation that the spot profiles appeared symmetric
over these small windows, suggests that the influence of other retinal layers was largely
eliminated.Additionally, in order to capture the large differences in refractive error of individual
mice and different retinal layers within the same mouse, the centroiding algorithm required an
additional feature. This was accomplished by roughly centering the initial search window
positions on the spots using the estimated amount of defocus, thereby increasing the dynamic
range of the wavefront sensor.The wave aberration was expressed as a combination of Zernike polynomials up to and including
10th order according to the ANSI standards [20]. The
Zernike coefficients calculated from the 10 images obtained after each saline drop have very
low variability, thus these images are averaged before averaging data from image sets acquired
after different saline drops. The variability in the wave aberration measurements is indicative
of the variation between administrations of saline drops. Wave aberrations measured over a
large pupil size were sometimes truncated and renormalized to a smaller pupil size for
analysis. Strehl ratios and modulation transfer functions (MTFs) were calculated from the wave
aberration data.
2.6 Chromatic aberration measurement
One of our goals was to measure the refractive state of the mouse eye, which requires
correcting the wave aberration measurements obtained in IR light for the chromatic aberration
of the mouse eye. To measure chromatic aberration, the mouse eye’s spherical
equivalent values were measured through wavefront sensing using four different wavelengths (457
nm, 514 nm, 633 nm, 789 nm). Besides the wavefront sensing SLD that has a central wavelength of
789 nm, an air-cooled Argon laser (CVI Melles Griot, Albuquerque, NM USA) provided two lines at
457 nm and 514 nm. A He-Ne laser (CVI Melles Griot, Albuquerque, NM USA) was used to measure
the wavefront at 633 nm. Measurements were performed on the optic disk of 9 mouse eyes over a 2
mm pupil diameter. The focus control of the wavefront sensing beacon mount was adjusted for
each wavelength so that the beacon is focused on the layer of interest and the wavefront spots
are sharp. For each wavelength and each eye, a minimum of 4 measurements were averaged. To
minimize variability of the wavefront over time caused by factors including dissipation of eye
drops and slow eye movements, each measurement made with a visible wavelength was paired with
an adjacent reference measurement made with 789 nm. No eye drops were applied between the two
measurements. The refractive error differences between each visible wavelength and the
reference wavelength were recorded and averaged for all eyes.
2.7 OCT imaging
A Bioptigen OCT system (Bioptigen, Research Triangle Park, NC USA) was used to collect OCT
images in the mouse eye. The light source of the OCT system has a central wavelength of 860 nm
and a bandwidth of 130 nm. Mice were anesthetized and dilated the same way as stated in section
2.3. A custom made rigid contact lens (Unicon Corporation, Osaka, Japan) was placed on the eye
to maintain corneal hydration. Line scan images were registered and averaged to increase signal
to noise using a custom-written algorithm [21].
2.8 Retinal image simulation
To better understand the implications of the mouse wave aberration for the quality of retinal
images, we simulated the retinal image quality that could be achieved under ideal incoherent
imaging conditions with and without AO correction. Note that this simulation doesn’t
account for scattering, photon noise and other sources of noise that degrade retinal image
quality. The optical transfer function (OTF) of the system can be calculated using typical
higher order aberration data measured in a mouse eye. The Fourier transform of the OTF is the
point spread function (PSF). By convolving an original image with the PSF, an in
vivo retinal image taken under best conditions without using AO was simulated. By
convolving the original image with a diffraction-limited PSF, an image with ideal AO correction
in the absence of noise was simulated.
3. Results
3.1 Shack-Hartmann image quality
The typical poor quality of mouse Shack-Hartmann wavefront images is illustrated in Fig. 3
[13]. Figure
3(a) shows a 3x3 array of spots taken from a wavefront sensor raw image obtained by
using an SLD beacon that filled the whole pupil, without optimizing the focus of the beacon.
The spots are large and blurry, demonstrating a cross-section that has a bimodal appearance. An
OCT scan of the mouse retina also shows that the strongest back reflections come generally from
two components: the inner retina and from the outer retina, with a distinctly less reflective
region in between (Fig. 3(b)). This is particularly
clear in linear plots of the OCT signal vs. depth since the logarithmic representation used
ubiquitously in OCT images reduces the large differences in the reflectance of different
layers. We use the term outer retina loosely here because we have not undertaken a detailed
analysis of which specific layers contribute most to this light. OCT images of mouse published
elsewhere tend to suggest major contributions from the retinal pigment epithelium (RPE) and
choroid [22,23].
Back reflections from the inner retina have a large variability with retinal location and are
not as uniform as that from the deeper layers. Some of the back reflections have a specular
reflection component that could originate from blood vessels, nerve fibers or the vitreal
retinal interface. This direction dependent back reflection from the inner retina results in an
asymmetric wavefront spot pattern that may further degrade the accuracy of aberration
measurement.
Fig. 3
(a) Typical non-optimized wavefront sensor spot pattern in the mouse eye with a cross
section of a single spot shown below. (b) An OCT line scan image with a cross section on a
particular retinal location. OCT signal vs. depth for the cross section is shown below in
linear scale. Both the wavefront spots and the OCT image show back reflections from two
major layers from the retina. (c) Improved wavefront spots when the beacon is focused on the
outer retina, when source focus was set at ~-10 D. Cross section of a single spot is shown
below. Solid line: normalized intensity profile for the cross section. Dashed line:
diffraction-limited spot profile for a lenslet. (d) An even sharper spot pattern can be
obtained when the beacon is focused on the disk, when the source focus was set at ~20 D.
Cross section of a single spot is shown below. Solid line: normalized intensity profile for
the cross section. Dashed line: diffraction-limited spot profile for a lenslet. The
wavefront sensor spots are very close to diffraction-limited. (e) An OCT line scan image
taken on the optic disk with a cross section on the central retinal artery. OCT signal vs.
depth for the cross secion is shown below in linear scale.
(a) Typical non-optimized wavefront sensor spot pattern in the mouse eye with a cross
section of a single spot shown below. (b) An OCT line scan image with a cross section on a
particular retinal location. OCT signal vs. depth for the cross section is shown below in
linear scale. Both the wavefront spots and the OCT image show back reflections from two
major layers from the retina. (c) Improved wavefront spots when the beacon is focused on the
outer retina, when source focus was set at ~-10 D. Cross section of a single spot is shown
below. Solid line: normalized intensity profile for the cross section. Dashed line:
diffraction-limited spot profile for a lenslet. (d) An even sharper spot pattern can be
obtained when the beacon is focused on the disk, when the source focus was set at ~20 D.
Cross section of a single spot is shown below. Solid line: normalized intensity profile for
the cross section. Dashed line: diffraction-limited spot profile for a lenslet. The
wavefront sensor spots are very close to diffraction-limited. (e) An OCT line scan image
taken on the optic disk with a cross section on the central retinal artery. OCT signal vs.
depth for the cross secion is shown below in linear scale.By optimizing the focus of the wavefront sensing beacon, the spot quality can be improved
significantly. Figure 3(c) shows that for the same entry
beam diameter (2 mm), by changing the beacon focus by ~-10 diopters so that the outer retina is
in focus, a relatively symmetric single spot pattern can be obtained. An even better and
brighter spot pattern was obtained at the optic disk (Fig.
3(d)), presumably generated by reflection from the central retinal artery that shows up
in an OCT image taken on the disk (shown in Fig. 3(e))
[24]. Source focus was at ~20 diopters for the spots
obtained at the disk. Dashed lines in Fig. 3(c) and
Fig. 3(d) are the diffraction-limited spot profile for
each lenslet. Figure 3(d) shows that on the optic disk
(central retinal artery), where the retina’s multilayer reflection is absent, the
mouse eye wavefront sensor spots we recorded were very close to diffraction-limited.
3.2 Refractive state
Two different refractive states were obtained by analyzing the spot pattern focused on the
outer retina vs. the inner retina. The average measured values are shown in Fig. 4
. for two different pupil sizes, a dilated pupil diameter of 2 mm, and a smaller pupil
diameter of 0.8 mm. 18 eyes were measured on the optic disk, and the spherical equivalent was
consistently hyperopic for all eyes, with a value of + 21.3 ± 1.3 D (mean
± 2 SEM (Standard Error of the Mean)) for a 2 mm pupil, and + 23.2 ± 4.5
D for a 0.8 mm pupil. 20 eyes were measured with the focus at the outer retina, and all the
eyes were myopic with the exception of one eye, with an average spherical equivalent of
−7.4 ± 1.8 D for a 2 mm pupil, and −11.1 ± 3.7 D
for a 0.8 mm pupil.
Fig. 4
Average equivalent sphere of mice eyes measured in two different positions over two pupil
sizes. Left column: spherical equivalent measured on the outer retina. Right column:
spherical equivalent measured on the inner retina. Light gray bars: spherical equivalent for
2 mm pupil size. Dark gray bars: spherical equivalent for 0.8 mm pupil size. Error bar:
± 2 times SEM.
Average equivalent sphere of mice eyes measured in two different positions over two pupil
sizes. Left column: spherical equivalent measured on the outer retina. Right column:
spherical equivalent measured on the inner retina. Light gray bars: spherical equivalent for
2 mm pupil size. Dark gray bars: spherical equivalent for 0.8 mm pupil size. Error bar:
± 2 times SEM.
3.3 Chromatic aberration and its effect on the measured refractive state
Longitudinal chromatic aberration (LCA) measurements were performed on the optic disk
(central retinal artery). This was done because contributions of the reflecting layers in the
retina may vary with wavelength and therefore we measured LCA where the multilayer reflection
properties of the retina are absent. Average LCA over 9 eyes between three visible wavelengths
(457, 514 and 633 nm) and reference wavelength 789 nm (refraction at visible wavelengths minus
refraction at 789 nm) are plotted as circle data points in Fig.
5
. Measured LCA from 457 nm to 789 nm ranges from 6.2 D to 12.4 D, with an average (
± 2 SEM) of 10.3 ± 1.3 D. From 457 nm to 633 nm, the average LCA is 7.7
± 1.0 D. Intersubject variability was higher than measurement variability for all
three wavelengths.
Fig. 5
Experimental data for mouse eye longitudinal chromatic aberration (LCA) and comparisons to
theoretical estimates of the mouse eye LCA using human experimental data. Blue dashed lines:
estimates of the mouse eye LCA using human LCA data and the water-filled reduced schematic
eye. Circle data points: average LCA between wavelengths of 457, 514 and 633 nm, and
reference wavelength 789 nm. For example, an average of 10.3 D of chromatic aberration is
measured between 457 nm and 789 nm. Error bar: ± 2 SEM.
Experimental data for mouseeye longitudinal chromatic aberration (LCA) and comparisons to
theoretical estimates of the mouse eye LCA using human experimental data. Blue dashed lines:
estimates of the mouse eye LCA using humanLCA data and the water-filled reduced schematic
eye. Circle data points: average LCA between wavelengths of 457, 514 and 633 nm, and
reference wavelength 789 nm. For example, an average of 10.3 D of chromatic aberration is
measured between 457 nm and 789 nm. Error bar: ± 2 SEM.In comparison, estimates of the mouse eye LCA using human eye LCA data are also plotted in
Fig. 5 (dashed line). In the human eye, LCA from 400 nm
to 900 nm were measured experimentally by Thibos et al. and Fernández et al. [25,26]. Using
approximations of the water-filled reduced schematic eye, the LCA ratio between different
animals equals to the ratio of their total eye power [27]. The ratio of total eye power between the mouse eye and the human eye is ~9.6, thus
estimates of the mouse eye LCA would be 9.6 times the humanLCA. Since the mouse data is
measured with regards to 789 nm, the estimate using human data is also shifted laterally so
that LCA is 0 at 789 nm. Figure 5 shows that the
measured mouseLCA is in good agreement with estimates using humanLCA data.97% of the mouse photoreceptors are rods and their spectral sensitivity peaks at 497-500 nm
[28,29]. Cone
photoreceptors are either maximally responsive to UV light (peak sensitivity at 360 nm) or to
medium (M)-wavelength light (peak sensitivity at 508 nm) [30]. Compared to the measured refraction at 789 nm, the true refraction would be closer
to what would be measured at 514 nm for rods and M cones. The measured average LCA between 514
nm and 789 nm is −7.9 D. Therefore the average refraction at the optic disk for 514
nm would be 13.4 D, and at the outer retina the refractive error would be −15.3 D
(for a 2 mm pupil). Chromatic aberration will be even larger between a UV wavelength and 789
nm, thus refraction at the outer retina for UV sensitive cones will be even more myopic.
Therefore the mouse eye refraction at the outer retina appears myopic for all the photoreceptor
classes.
3.4 Wave aberration of the eye
A typical wavefront sensor spot pattern used for measurement is shown in Fig. 6(a)
. All images are captured over 2.2 mm pupil sizes and the images are truncated to 2 mm
for analysis. Higher order aberration wavefront maps for 6 eyes from 3 mice for a 2 mm pupil
size are shown as examples in Fig. 6(b).
Fig. 6
(a) Typical wavefront sensor spot pattern in the mouse eye used for measurement. Image is
captured at 2.2 mm pupil size and wavefront is analyzed at 2 mm pupil size. (b) Higher order
aberration wavefront maps for 6 eyes from 3 mice for 2 mm pupil size.
(a) Typical wavefront sensor spot pattern in the mouse eye used for measurement. Image is
captured at 2.2 mm pupil size and wavefront is analyzed at 2 mm pupil size. (b) Higher order
aberration wavefront maps for 6 eyes from 3 mice for 2 mm pupil size.The average Zernike coefficients for each mode were calculated in 20 eyes across a 2 mm pupil
size using wavefront images focused on the outer retina (Fig. 7
). Even though we found indications of correlation between the Zernike coefficients and
wavefront maps of the two eyes, we did not find obvious mirror symmetry [31]. Therefore average data from the left eyes and right eyes are shown
separately. Lower and higher order aberrations are shown using different scales. Zernike
defocus (single index j = 4 or ) has the largest magnitude of all the modes. The next largest mode is Zernike
spherical aberration (single index j = 12 or ), with a value of −0.09 ± 0.06 μm. In
comparison, the diffraction-limited RMS for 789 nm wavelength needs to be <0.06
μm (λ/14) according to the Maréchal criterion. Intersubject
variability was high for almost all higher order modes. The total root mean square (RMS)
wavefront error ( ± 2 SEM) was 1.28 ± 0.19 μm. The average
higher order RMS wavefront errors were 0.38 ± 0.03 μm.
Fig. 7
Mean values of 2nd to 5th order Zernike modes for 20 eyes across a 2 mm pupil, using
wavefront spots focused on the outer retina. Zernikes for the 10 left eyes are shown in (a)
and ones for the 10 right eyes are shown in (b). Error bars represent ± 2 times
SEM. Lower and higher order aberrations are shown using different scales.
Mean values of 2nd to 5th order Zernike modes for 20 eyes across a 2 mm pupil, using
wavefront spots focused on the outer retina. Zernikes for the 10 left eyes are shown in (a)
and ones for the 10 right eyes are shown in (b). Error bars represent ± 2 times
SEM. Lower and higher order aberrations are shown using different scales.
3.5 Strehl ratio and higher order aberrations
These wave aberration measurements can be used to estimate the benefit of correcting the
mouse aberrations with adaptive optics. The role of higher order aberrations in reducing the
retinal image quality of the mouse eye was assessed by computing the Strehl ratio. Figure 8
shows how the Strehl ratio improves when we successively remove Zernike orders from the
wave aberration for a 2 mm pupil. Note that Strehl ratios are calculated for a wavelength of
789 nm, and for visible wavelengths the calculated Strehl ratios will be lower than shown in
Fig. 8. Values on the abscissa correspond to the
highest Zernike order removed. For the 20 eyes, Zernike orders up to and including 6th order
need to be corrected to achieve diffraction-limited imaging (Strehl>0.8) for the 789 nm
wavelength. If aberrations are measured using a SHWS and corrected using a deformable mirror,
this sets a lower bound for the number of lenslets in the wavefront sensor and the number of
actuators in the deformable mirror required to achieve diffraction-limited imaging. The maximum
number of Zernike modes that can be reliably reconstructed is approximately the same as the
number of lenslets [32]. Since Zernike coefficients up
to 6th order correspond to 25 total coefficients without piston, tip and tilt, at least 25
lenslets are needed to recover the important aberrations in the mouse eye. The number of
actuators required depends on the type of deformable mirror to be used [32,33].
Fig. 8
Strehl ratio for the 20 eyes over a 2 mm pupil at a wavelength of 789 nm. Error bars
represent ± 2 SEM. Each data point shows the calculated Strehl ratio after a
number of lower order aberrations are corrected. For example, the number 5 means that
second, third, fourth and fifth order aberrations (or all aberration modes shown in Fig. 4) are corrected. Dashed line represents a Strehl
ratio of 0.8, above which imaging is considered to be diffraction-limited.
Strehl ratio for the 20 eyes over a 2 mm pupil at a wavelength of 789 nm. Error bars
represent ± 2 SEM. Each data point shows the calculated Strehl ratio after a
number of lower order aberrations are corrected. For example, the number 5 means that
second, third, fourth and fifth order aberrations (or all aberration modes shown in Fig. 4) are corrected. Dashed line represents a Strehl
ratio of 0.8, above which imaging is considered to be diffraction-limited.
3.6 Pupil size, higher order aberrations and modulation transfer functions (MTF)
To determine aberrations present for a pupil size smaller than 2 mm, wave aberrations
measured at 2 mm can be truncated and renormalized. Figure 9
shows how the total higher order RMS increases and how the Strehl ratio decreases as the
pupil size increases. Note that again Strehl ratios are calculated for a wavelength of 789 nm.
Also note that different pupil sizes may include different number of Zernike coefficients,
because as the SHWS image is truncated, the number of SHWS spots available in the images is
reduced. For pupil sizes larger than 1.4 mm, the aberration coefficients include up to 10th
order Zernikes; for smaller pupil sizes, the number of Zernike orders included is reduced. For
example, 21 lenslets are included in a 0.6 mm pupil, enough to sample the wavefront up to 5th
order Zernikes. For a 0.6 mm pupil size, the higher order RMS is smaller than λ/14,
which is considered diffraction-limited according to the Maréchal criterion.
Fig. 9
The total higher order aberration RMS increases, and the Strehl ratio decreases as the
eye’s pupil size increases. Strehl ratio is calculated for the measurement
wavelength of 789 nm. Diamond data points: average measured higher order RMS (3rd to 10th
order) in µm. Dashed line: diffraction-limited RMS of λ/14, according
to the Maréchal criterion. Triangular data points: average Strehl ratio for each
pupil size. Error bars represent ± 2 SEM for the 20 eyes.
The total higher order aberration RMS increases, and the Strehl ratio decreases as the
eye’s pupil size increases. Strehl ratio is calculated for the measurement
wavelength of 789 nm. Diamond data points: average measured higher order RMS (3rd to 10th
order) in µm. Dashed line: diffraction-limited RMS of λ/14, according
to the Maréchal criterion. Triangular data points: average Strehl ratio for each
pupil size. Error bars represent ± 2 SEM for the 20 eyes.Aberrations decrease as pupil size decreases, but the effect of diffraction is larger for
smaller pupil sizes. In the human eye, this well known trade off results in a pupil size of 2-3
mm providing the optimum lateral resolution [3,34]. To determine what pupil size will provide the best trade
off in the mouse eye, radially-averaged modulation transfer functions (MTFs) were computed
using the higher order aberration data in the 20 eyes for different pupil sizes (Fig. 10
). At middle and low frequencies, the MTF is the highest for a 0.8 mm pupil size, where
the MTF is close to diffraction-limited. However, larger pupil sizes provide better modulation
transfer at the very highest spatial frequencies that only these large pupils can pass at
all.
Fig. 10
Mean of radial averaged MTF for 20 eyes over different pupil sizes (0.6 mm to 2 mm) for
the measurement wavelength of 789 nm. Defocus and astigmatism are removed before MTF is
calculated.
Mean of radial averaged MTF for 20 eyes over different pupil sizes (0.6 mm to 2 mm) for
the measurement wavelength of 789 nm. Defocus and astigmatism are removed before MTF is
calculated.
3.7 Comparison between MTF of the human eye and the mouse eye
The average mouse eye MTF was compared with the average MTF computed from higher order
aberration measurements obtained in 378 healthy normal human eyes over a 6 mm pupil size [35]. The human higher order aberration data were measured up
to and including 5th order Zernike polynomials. To characterize the angular resolution of the
eye viewing the world, the MTF was plotted in cycles/degree (Fig. 11(a)
). The mouse eye has a lower diffraction-limited MTF for angular resolution than the
human eye as expected. Based on higher order aberrations, the mouse eye and the human eye MTF
drop from the diffraction-limited MTF in a similar way, and the human eye modulation transfer
is higher than that of the mouse eye for all frequencies. To characterize the spatial
resolution when an investigator is viewing the retina, the MTF was plotted in cycles/mm. This
way, eyes with higher NAs have larger diffraction-limited MTFs, and eyes with the same NA share
the same diffraction-limited MTFs despite their differing pupil sizes and focal lengths. Figure 11(b) shows that the mouse eye has a better
diffraction-limited MTF in cycles/mm than the human eye due to its larger NA. Even with higher
order aberrations present, the mouse eye still has a higher modulation transfer for all the
spatial frequencies expressed in cycles/mm than the human eye.
Fig. 11
Modulation transfers for a human eye (6 mm pupil) and a mouse eye (2 mm pupil) plotted in
(a) cycles/degree to characterize angular resolution for vision and (b) cycles/mm to
characterize spatial resolution for imaging the retina. Only higher order aberrations (HOA)
are considered. Solid black line: mouse eye with 2 mm pupil size with HOA. Solid green line:
human eye with 6 mm pupil size with HOA. Dashed black line: Diffraction-limited mouse eye
with 2 mm pupil size. Dashed green line: Diffraction-limited human eye with 6 mm pupil
size.
Modulation transfers for a human eye (6 mm pupil) and a mouse eye (2 mm pupil) plotted in
(a) cycles/degree to characterize angular resolution for vision and (b) cycles/mm to
characterize spatial resolution for imaging the retina. Only higher order aberrations (HOA)
are considered. Solid black line: mouse eye with 2 mm pupil size with HOA. Solid green line:
human eye with 6 mm pupil size with HOA. Dashed black line: Diffraction-limited mouse eye
with 2 mm pupil size. Dashed green line: Diffraction-limited human eye with 6 mm pupil
size.
3.8 Simulated retinal images
The high spatial resolution in the mouse eye as shown in Fig.
11(b) provides the potential of achieving even higher resolution retinal images than
can be obtained in the human eye. To simulate retinal images of fluorescent ganglion cells, a
fluorescent ganglion cell from a flat-mounted rat retina imaged with confocal microscopy with
1.2 NA is used as the original image [Fig. 12(a)
]. Note that simulation used aberration data measured with 789 nm wavelength; if imaging
is performed using shorter wavelength, image degradation will be worse with aberrations, and
the benefit of AO correction will be greater. For imaging without AO, after focus and
astigmatism is corrected, Fig. 10 shows that 0.8 mm
pupil gives the best MTF for low to middle spatial frequencies. For a 0.8 mm pupil size, the
higher order aberration RMS for a typical eye is only 0.07 μm. Figure 12(b) shows the simulated best ganglion cell image obtained without
AO over a 0.8 mm pupil. Even though image contrast is reduced, all the ganglion cell bodies and
most of the axons and dendrites can be resolved. Figure
12(c) shows the same cells imaged with ideal AO correction over a 2 mm pupil size with
increased contrast. Figure 12(d)-(f) shows the same
simulation for incoherent imaging of rods. The rod image in Fig. 12(d) is a simulated triangularly packed mosaic [36] with a nearest neighbor distance of 1.6 μm (calculated using average
rod density [28]). For the rod image, if imaged without
AO using a 0.8 mm pupil (Fig. 12(e), the spatial
frequency corresponding to the average rod spacing is already beyond the cut off frequency of
the optics. Here the improvement afforded by AO is critical to resolve individual cells.
Fig. 12
Simulated retinal images of ganglion cells and rod photoreceptors under ideal imaging
conditions and in the absence of noise with/without AO correction. (a) A fluorescent
ganglion cell image from a flat-mounted rat retina imaged with confocal microscopy with 1.2
NA. (b) A simulation of the retina in (a) being imaged without AO correction over a 0.8 mm
pupil under the best possible condition. (c) A simulation of the same ganglion cell being
imaged with AO correction over a 2 mm pupil under the best possible condition. (d) A
simulated rod photoreceptor mosaic in the mouse retina. (e) & (f) are the same
without AO and with AO simulations of the rod photoreceptor mosaic. Scale bar: 20
µm.
Simulated retinal images of ganglion cells and rod photoreceptors under ideal imaging
conditions and in the absence of noise with/without AO correction. (a) A fluorescent
ganglion cell image from a flat-mounted rat retina imaged with confocal microscopy with 1.2
NA. (b) A simulation of the retina in (a) being imaged without AO correction over a 0.8 mm
pupil under the best possible condition. (c) A simulation of the same ganglion cell being
imaged with AO correction over a 2 mm pupil under the best possible condition. (d) A
simulated rod photoreceptor mosaic in the mouse retina. (e) & (f) are the same
without AO and with AO simulations of the rod photoreceptor mosaic. Scale bar: 20
µm.
4. Discussion
4.1 Method for measuring the wave aberration
Measuring the wave aberration of the mouse eye with a replica of the method used in the human
eye is difficult due to the severely degraded Shack-Hartmann spots [13]. The current study suggests that the origin of the degraded spot pattern
lies in the multiple back reflections from the optically thick retina, instead of imperfections
such as higher order aberrations present in the mouse eye optics. By carefully controlling the
beacon focus, our wavefront sensor spots shown in Fig.
3(c) and 3(d) provide good quality approaching
the diffraction-limit. This wavefront sensor that favors backscattered light from a specific
retinal layer improves the wavefront spot image quality and hence aberration measurements in
the mouse eye. Another method that may be worth exploring for wavefront sensing in the mouse
eye on a particular layer is to utilize coherence [37].Even though according to Eqs. (1) and (2), the elongation of the wavefront spots in the
human eye caused by the retinal thickness is a lot smaller than that in the mouse, it would be
interesting to find out if this would affect the wavefront sensing accuracy in the human eye,
especially at eccentricities with thicker nerve fiber layer and larger overall thickness. This
will probably be a small effect, but it has been observed during AOSLO imaging of the human
retina that the wavefront sensor is sometimes confused, locking on the nerve fiber layer
instead of on the photoreceptor layer [38].
4.2 Refractive error and the small eye artifact
As one would expect, we measured different refractive errors when the wavefront sensor beacon
was focused on different layers and at different retinal eccentricities. The average equivalent
sphere values are + 21.9 D hyperopic when the beacon is focused on the optic disk and
−7.4 D myopic when focused on the outer retina for 789 nm wavelength. The refraction
value on the optic disk (central retinal artery) was further confirmed by the ~20 D of trial
lens focus needed on the fundus camera for the disk to be in focus. This is consistent with the
so called “small eye artifact” in retinoscopy [39], in which the true measured refraction is confounded by reflection from a
retinal layer different from the photoreceptor layer. The small eye artifact is especially
pronounced in eyes where the retinal thickness is large relative to the effective focal length
of the eye. It has been calculated that the dioptric difference of the positions from the
vitreo-retinal interface (where the retinoscopic reflection is supposed to originate [39]) to the RPE for the adult C57BL/6 mice is ~39 D [2]. The dioptric difference we measured from the optic disk to
the outer retina was 29.2 D and lies within the total optical thickness of the retina.All previous studies provided hyperopic readings for the the mouse eye. However, with the
confounding small eye artifact, the true refraction is likely to be less hyperopic than
reported [1,2]. Our
measurements indicate that the real refraction for the mouse eye is probably myopic. White
light streak retinoscopy studies have found + 13.5 D to over + 15 D of hyperopia [39-41]. An
IR photoretinoscopy study found about + 7 D of hyperopia [2], and finally the previous Shack-Hartmann study found + 9.7 D of hyperopia [13].In the present study, we also observed hyperopic refractions in the case where the beacon was
focused on the central retinal artery located on the optic disk. The OCT image in Fig. 3(c) shows that the central retinal artery is highly
reflective and lies at a depth that is close to that of the vitreo-retinal interface. In that
case, taking into account the measured chromatic aberration of the mouse eye, the average
refraction at the disk (central retinal artery) for 514 nm was calculated to be 13.4 D for a 2
mm pupil, or 15.3 D for a 0.8 mm pupil. This value is consistent with retinoscopy results,
suggesting that the retinoscopic reflection is probably generated from the vitreo-retinal
interface as suggested by Glickstein [39]. Both the IR
photoretinoscopy study [2] and the previous SHWS study
[13] measured a value between our 21.9 D hyperopia for
the inner retina (central retinal artery) and −7.4 D myopia for the outer retina.
Our result may be consistent with those results if the signal they measured is a weighted
average from different retinal layers.To estimate the true refraction of the mouse eye, the average refraction at the outer retina
was calculated to be quite myopic at −15 D for a wavelength of 514 nm (that is, in
which the chromatic aberration has been taken into account). In the mouse, the available
evidence seems to suggest that the most reflective layer in the outer retina may lie deeper
than the photoreceptor, in the RPE and choroid [22,23]. If the spot centroid lies on the RPE or choroid instead
of the photoreceptor layer, the true refraction would be less than −15 D, but would
still be myopic. This is the first study that indicates the mouse eye may be myopic. It would
make sense for the animal to be myopic or emmetropic, since for a hyperopic eye, no object
vergence will give a sharp image on the retina. With a large depth of field of at least
± 11 D [1], even with very myopic eyes the
mouse should still be able to achieve reasonable focus for far objects.Our mice were anesthetized and it is not clear how much anesthesia affects measured
refraction. It was reported that in one mouse eye, the hyperopic spherical equivalent was
reduced by 7 D in the anesthetized eye [13]. However, a
recent photorefraction study measured the opposite effect, that ketamine–xylazine
anesthesia caused on average 6.9 D of hyperopic refractive shift in 9 mice eyes [42]. Another retinoscopy test of five
C57BL/6J mice measured no difference larger than ± 2 D in the same
eyes between awake vs. anesthetized state (Marsha L. Kisilak, unpublished data). In any case, a
focal shift in the range of −7 D to 7 D to the anesthetized refractive state of
−15 D will not render the mouse eye hyperopic.
4.3 Chromatic aberration
To our knowledge this is the first in vivo measurement of the mouse eye
chromatic aberration. The current in vivo measurement and previous estimations
of the mouse eye chromatic aberration are summarized in Table
2
. Remtulla and Hallett developed a homogeneous schematic eye from measuring frozen
mouse eye sections [1]. With refractive indices measured
at 4 wavelengths and paraxial ray tracing, they estimated a chromatic aberration of 20.2 D
(13-20 D) between 488 nm and 655 nm. In contrast, we measured only 7.7 D ± 1.0 D
(mean ± 2SEM) chromatic aberration between 457 nm and 633 nm. Except for
experimental uncertainties, we are not sure about the origins of the large difference between
this schematic eye and our result. Even though schematic eye studies assume a narrow paraxial
beam while we experimentally measured the aberrations over a fully dilated pupil, our
refraction measurements in a 2 mm and 0.8 mm pupil size (Fig.
4) didn’t show much change. The water-filled reduced schematic mouse eye
model gives 8.8 D for the same wavelength range as the current study using total eye power
values from Remtulla and Hallet [1,27], which agrees very well with our experimental value.
Table 2
Current in vivo measurement and previous estimations of the mouse eye
chromatic aberration
Rat eye × 2
Mouse eye
Schematic eye
from measuring frozen sections [43]
Retinoscopy [44,45]
Schematic eye from
measuring frozen sections [1]
Water-filled schematic eye [27]
Current study
Wavelength Range (nm)
486–656
Broadband blue to red
488–655
457–633
457–633
Chromatic Aberration
11.6 D
5.0–7.0 D
20.2 D (13–20 D)
8.8 D
7.7 ± 1.0 D
If the only significant difference between mouse and rat eyes is a linear scale difference of
two times, the mouse eye chromatic aberration should be two times as large as that of the rat
eye. Similar to the mouse schematic eye, Chaudhuri et al. had a rat homogeneous model eye
measured from enucleated eyes and their estimation for chromatic aberration was 5.8 D between
486 nm and 656 nm [43]. Our measured value (7.7 D) is
closer to two times the value from the Chaudhuri et al. rat schematic eye model (11.6 D) than
the mouse schematic eye value (20.2 D). There had also been retinoscopy measurements using
relatively broad-band red and blue light, and their estimated rat eye LCA was 2.5-3.5 D [44,45]. Our result is
consistent with a value that is roughly two times the rat eye retinoscopy results (5.0 D
– 7.0 D).
4.4 Comparison of measured higher order aberrations and MTF with other studies
The magnitude of higher order aberrations measured in anesthetized mice in this study is
smaller than that measured in awake behaving mice by de la Cera et al. [13]. If the measured data in this study is cropped to 1.5 mm pupil size and
includes only up to 4th order Zernikes, the average RMS higher order aberration is 0.20
± 0.05 μm (mean ± standard deviation) compared with 0.32
± 0.08 μm reported previously [13]. The average spherical aberration measured in this study is negative, but with a
large variability, compared to a measured larger positive spherical aberration of 0.15
± 0.06 μm reported previously. If mice being used are genetically
identical, there are several possible explanations for the differences in the magnitude of
higher order RMS. Different wavefront spot quality and centroiding algorithms could affect the
final measurement. Anesthetized mice may have different aberrations compared with awake mice
however awake mice may have better optical quality, as indicated in studies done in mice and
cats [13,46]. The
eccentricities at which wave aberrations are being measured can be different. Last but not
least, depending on the different contribution to the wavefront spots from various retinal
layers, aberrations measured can be different.IR photorefraction in the mouse eye typically show non-uniform intensity distributions with
ring-shaped areas in the pupil [47]. This pattern is
different from the smooth pupil illumination pattern observed in human eyes using the same
technique [48]. It is possible that higher order
aberrations contribute to this non-uniform intensity, even though we did not find the mouse eye
higher order aberrations to be significantly larger than that of the human eye. As was
described in 3.1, the mouse eye SHWS spot intensity on the pupil is non-uniform and contains
reflections from multiple layers in the retina, some of which having a specular reflection
component. These reflections are also eccentricity dependent, with an especially bright
reflection coming from the central retinal artery. We do not know whether these
non-uniformities have an impact on other methods of measuring the eye’s optics, such
as IR photorefraction. Geometrical theory to predict eccentric photorefraction intensity
profiles assumes a single retinal layer which diffusely scatters light, with reflection
properties independent of eccentricity [49,50]. It might be possible that the multilayer properties of
the mouse retina may also contribute to the non-uniform intensity distribution observed during
IR photorefraction.Artal et al. calculated the MTF for a single mouse eye using a double-pass method and found
the modulation to be lower than 0.1 for 1 cycle/degree for a 1 mm pupil [10]. Our data gives a modulation of above 0.5 for the same frequency and
pupil size for an average of 20 eyes even before defocus and astigmatism are removed, and a
modulation of above 0.8 when defocus and astigmatism are removed. Our MTF is significantly
higher than that measured using the double-pass method. However, comparisons of double-pass and
wavefront sensor MTFs are complicated by the failure of the wavefront sensor to capture
scatter, and the possibility that the double-pass method exaggerates the effect of scatter.
Following the lower higher order RMS, the calculated MTF in this study is also slightly higher
than previous results obtained using Shack-Hartmann wavefront sensor [13].
4.5. Optical and neural factors in mouse spatial vision
The visual acuity of the mouse is only 0.5-0.6 cycles/deg [51,52], about a factor of 100 worse than human.
The optics of the mouse eye could support substantially higher visual acuity than this. We have
found that a pupil size of 0.8 mm provide the best MTF for low and middle spatial frequencies
(Fig. 10). Figure 13
shows that for a 0.8 mm pupil and 514 nm wavelength, when both lower and higher order
aberrations are taken into account, the modulation transfer is ~0.6 for the maximum acuity of
0.6 cycles/deg, indicating that optics is not the limiting factor for mouse vision.
Fig. 13
Average modulation transfer functions for the mouse eye over a 0.8 mm pupil diameter. Note
that MTFs are calculated for a wavelength of 514 nm, where rods and M-cones are near their
spectral sensitivity peak.. To calculate MTF for a visible wavelength using wave aberration
data measured at 789 nm, we assume other than defocus, variations in the other aberration
coefficients between visible and IR wavelengths are small [26,56]. For all 20 eyes, defocus was
increased by 7.9 D based on measured average LCA between the two wavelengths and Zernike
coefficients for other aberrations remained the same.
Average modulation transfer functions for the mouse eye over a 0.8 mm pupil diameter. Note
that MTFs are calculated for a wavelength of 514 nm, where rods and M-cones are near their
spectral sensitivity peak.. To calculate MTF for a visible wavelength using wave aberration
data measured at 789 nm, we assume other than defocus, variations in the other aberration
coefficients between visible and IR wavelengths are small [26,56]. For all 20 eyes, defocus was
increased by 7.9 D based on measured average LCA between the two wavelengths and Zernike
coefficients for other aberrations remained the same.The rod Nyquist limit in the mouse is 9.5 cycles/deg (calculated using average rod density
data from Jeon et al. assuming triangular packing [28]).
The visual acuity is, not surprisingly, much lower than that expected from the rod mosaic.
Presumably, as in the human, the mouse rod mosaic pools photons over many rods to increase the
signal-to-noise ratio at low light levels. However, different from the human eye, mouse eye
spatial vision seems to be governed by rods instead of cones. It has been shown that knock-out
mice with only rods have normal grating acuity, but mice with only cones have much worse than
normal acuity [53]. At the rod Nyquist limit, Fig. 13 shows that when all aberrations are present, the
average modulation transfer for an eye with 0.8 mm pupil is at ~0.02. This is a conservative
estimate since the defocus values used assumes fixation at infinity. For an object closer to
the mouse which it is probably more interested in, the myopic mouse could have a modulation as
high as 0.12 (if defocus is excluded from MTF calculations). The MTF is sufficiently reduced at
the rod Nyquist limit that there is essentially no risk of aliasing.The cone Nyquist limit in the mouse is 1.6 cycles/deg (also calculated using density data
from Jeon et al. assuming triangular packing [28]). At
1.6 cycles/deg, Fig. 13 shows that even when all
aberrations are present, the average modulation transfer for an eye with 0.8 mm pupil is 0.25.
For an object closer to the mouse, the myopic mouse could have a modulation as high as 0.75 for
the cone Nyquist limit. Therefore cones undersample the mouse retinal image, unless there are
additional major degradations to the image quality not accounted for, such as scattering.Similarly the ganglion cell Nyquist limit (even if we assume that each cell independently
samples the retinal image) is 0.9-1.3 cycles/deg [28,54,55], which undersamples the retinal image even more than the cones do. If only a subset
of the total ganglion cell population is used for behavioral acuity, that subset could set the
resolution limit for the mouse. Alternatively, additional high frequency loss caused by
ganglion cell receptive field size or from more central stages in the mouse visual pathway
might account for the poor behavioral acuity of the mouse.In any case, the mouse eye optics is very good with respect to the retinal sampling
properties and the behavioral performance of the entire animal. It may be that the optical
quality of the mouse eye has evolved to be as good as it is to ensure high contrast at the very
low frequencies that the nervous system is sensitive to. This relationship is consistent with
observations in vertebrates that the optical quality of the eye tends to outperform the
sampling limitations imposed by the retina, especially in animals that live in dim light [57-59]. One
such example is the barn owl, which has almost diffraction-limited optics and an axial length
not much smaller than the human eye (17.5 mm), but with a behavioral acuity of only about 4
cycles/degree [60,61].
4.6 Optical quality of the mouse eye: visual performance vs. retinal imaging
A comparison of the optical quality of the human and mouse eye depends on whether the basis
for the comparison is the organism’s visual performance or the
experimenter’s ability to resolve structure in the retina. For a scene that subtends
a certain visual angle, the small mouse eye can only use a very small size retina to sample the
scene and thus is at a clear disadvantage. The MTF expressed in cycles per deg in Fig. 11(a) shows that the human optics can support better
visual performance than the mouse optics can.On the other hand, the NA of the dilated mouse eye is approximately two times larger than the
dilated human eye, presumably because the increase in light gathering power benefits the
nocturnal mouse. From the point of view of retinal imaging, this larger NA makes it
theoretically possible to image anatomical structures in the mouse retina that are half the
size of the smallest feature that can be resolved in the human eye. When higher order
aberrations are taken into account, and for the case of imaging the retina rather than visual
performance, the mouse still outperforms the human eye as shown in Fig. 11(b). This runs counter to the view that mice have large amounts of
higher order aberrations and encourages the prospects for resolving cellular structures in the
rodent retina even without AO.
4.7 Adaptive optics retinal imaging in the mouse eye
With the remarkably good optical quality of the mouse eye for imaging the retina,
it’s possible to obtain high quality retinal images even without using AO [62-65].
However AO can provide additional improvements in both lateral and axial resolution and retinal
image contrast. The diffraction-limited in vivo resolution is ~0.7
µm for lateral resolution, and ~6 µm for axial resolution (calculated for
550 nm wavelength and 2 mm pupil size). Apart from providing diffraction-limited imaging, AO
can also increase the pupil size that can be used for imaging, correct both lower and higher
order aberration at video rates, and increase light collection efficiency for confocal or
two-photon imaging. The ability to capture the entire wave aberration in the anesthetized mouse
eye over a fully dilated pupil with reflected light is promising for high-speed adaptive
correction of mouse retinal images. The amount of higher order aberrations measured over the
0.49 NA (2 mm pupil) mouse eye is similar to what is measured in the normal human eye over a
0.18 NA (6 mm pupil) [66] thus is correctible by AO. An
adaptive optics instrument customized for in vivo imaging of the mouse retina
with this improved wavefront sensing method is currently under development.
5. Conclusions
Using a wavefront sensor that favors backscattered light from a specific retinal layer in the
mouse eye, we have improved the quality of wavefront sensor spots and thus improved aberration
measurements in the mouse eye. Contrary to common belief, we measured the mouse eye to be
myopic, and the optical quality of the mouse eye to be remarkably good. An instrument
constructed with this improved wavefront sensor technique may provide a faster and more
effective correction for the mouse eye aberrations using AO. This instrument can potentially
achieve a lateral resolution at least two times higher than that of the human eye, and an axial
resolution at least four times higher than that of the human eye. Such an instrument could allow
microscopic imaging and monitoring of retinal development, disease progression, or the efficacy
of therapy in single animals over time.
Authors: Daniel C Gray; William Merigan; Jessica I Wolfing; Bernard P Gee; Jason Porter; Alfredo Dubra; Ted H Twietmeyer; Kamran Ahamd; Remy Tumbar; Fred Reinholz; David R Williams Journal: Opt Express Date: 2006-08-07 Impact factor: 3.894
Authors: Omer P Kocaoglu; Stephen R Uhlhorn; Eleut Hernandez; Roger A Juarez; Russell Will; Jean-Marie Parel; Fabrice Manns Journal: Invest Ophthalmol Vis Sci Date: 2007-03 Impact factor: 4.799
Authors: Vivek J Srinivasan; Tony H Ko; Maciej Wojtkowski; Mariana Carvalho; Allen Clermont; Sven-Erik Bursell; Qin Hui Song; Janis Lem; Jay S Duker; Joel S Schuman; James G Fujimoto Journal: Invest Ophthalmol Vis Sci Date: 2006-12 Impact factor: 4.799
Authors: Jessica I W Morgan; Alfredo Dubra; Robert Wolfe; William H Merigan; David R Williams Journal: Invest Ophthalmol Vis Sci Date: 2008-10-24 Impact factor: 4.799
Authors: Peter Charbel Issa; Mandeep S Singh; Daniel M Lipinski; Ngaihang V Chong; François C Delori; Alun R Barnard; Robert E MacLaren Journal: Invest Ophthalmol Vis Sci Date: 2012-02-29 Impact factor: 4.799
Authors: Anna Fasoli; James Dang; Jeffrey S Johnson; Aaron H Gouw; Alex Fogli Iseppe; Andrew T Ishida Journal: J Comp Neurol Date: 2017-02-13 Impact factor: 3.215