| Literature DB >> 21481282 |
Jesper Kristiansen1, Mette Korshøj, Jørgen H Skotte, Tobias Jespersen, Karen Søgaard, Ole S Mortensen, Andreas Holtermann.
Abstract
OBJECTIVE: A number of small portable systems that can measure HRV are available to address questions related to autonomic regulation in free-living subjects. However, ambulatory HRV measurements obtained through use of these systems have not previously been validated against standard clinical measurements such as Holter recordings. The objective of this study was to validate HRV obtained using a commonly used system, Actiheart, during occupational and leisure-time activities.Entities:
Mesh:
Year: 2011 PMID: 21481282 PMCID: PMC3080340 DOI: 10.1186/1475-925X-10-27
Source DB: PubMed Journal: Biomed Eng Online ISSN: 1475-925X Impact factor: 2.819
Characteristics of the participants (n = 8)
| Mean | SD | Range | |
|---|---|---|---|
| Age (years) | 35.50 | 17.36 | 19 - 58 |
| Height (cm) | 169.88 | 6.01 | 158 - 177 |
| Weight (kg) | 75.03 | 15.46 | 51.2 - 100.2 |
| BMI (kg/m2) | 25.90 | 4.70 | 20.2 - 33.5 |
| Fat% | 34.36 | 9.53 | 17 - 45 |
| Waist circumference (cm) | 88.08 | 13.93 | 72 - 106 |
| Waist - hip ratio | 0.79 | 0.14 | 0.53 - 0.98 |
| Aerobic fitness (mL O2/min/kg) | 27.93 | 4.75 | 19.9 - 33.7 |
| Sleeping heart rate (beats/min.) | 54.04 | 5.94 | 47 - 66 |
| Systolic blood pressure (mmHg) | 117.38 | 10.64 | 98 - 132 |
| Diastolic blood pressure (mmHg) | 78.21 | 8.94 | 64 - 87 |
Figure 1Position of electrodes for the Holter monitor and Actiheart, respectively.
Figure 2Ambulatory HRV measurements of Actiheart versus Holter system; a) ln(TP); b) ln(LFP); c) ln(HFP); d) ln(LF/HF).
Figure 3Ambulatory HRV measurements of Actiheart versus Holter system; a) RRmean; b) LFn.u.; c) HFn.u.
Slopes and intercepts (95% CI) estimated by Deming regression analyses of Actiheart versus Holter (n = 489) HRV measures.
| Slope | 95%CI | Intercept | 95%CI | |
|---|---|---|---|---|
| RRmean | 0.998 | (0.994 - 1.001) | 1.91 | (-0.515 - 4.332) |
| LFn.u. | 0.981 | (0.962 - 1.001) | 0.01 | (-0.014 - 0.027) |
| HFn.u. | 0.980* | (0.964 - 0.997) | 0.01 | (-0.002 - 0.022) |
| ln(TP) | 0.977* | (0.962 - 0.994) | 0.17* | (0.047 - 0.289) |
| ln(LFP) | 0.995 | (0.977 - 1.013) | 0.03 | (-0.088 - 0.166) |
| ln(HFP) | 0.971* | (0.950 - 0.991) | 0.19* | (0.061 - 0.315) |
| ln(LF/HF) | 0.973* | (0.950 - 0.995) | 0.00 | (-0.043 - 0.034) |
Significantly different from 1 (slopes) or 0 (intercepts); * P < 0.05;
Figure 4Standardized differences of (Actiheart - Holter) HRV measures versus the mean value; a) ln(TP); b) ln(LFP); c) ln(HFP); d) ln(LF/HF); e) LFn.u.; f) HFn.u.
Number of ECG time segments for which the standardized difference between Actiheart and Holter HRV estimates (in units of repeatability standard deviations, z) exceeds 1 and 2, respectively.
| HRV metric | Number of segments z>1 | Percentage of totala | Number of segments z>2 | Percentage of totala | |
|---|---|---|---|---|---|
| ln(TP) | 0.41 | 19 | 3.9% | 4 | 0.8% |
| ln(LFP) | 0.39 | 28 | 5.7% | 10 | 2.0% |
| ln(HFP) | 0.50 | 37 | 7.6% | 12 | 2.5% |
| ln(LF/HF) | 0.41 | 47 | 9.6% | 12 | 2.5% |
| LFn.u. | 0.078 | 39 | 8.0% | 12 | 2.5% |
| HFn.u. | 0.078 | 32 | 6.5% | 8 | 1.6% |
a Total n = 489 ECG segments
Figure 5Mean heart period (RRmean) and heart rate variability indices of 8 cleaning workers during a workday. a) RRmean, b) ln(LF/HF), c) ln(TP) and d) ln(HFP) derived from 5 min recordings sampled every 30 min. Measurements are categorized according to work period (9:00 - 13:00), break during work (11:00), leisure period (13:30 - 22:00) and sleep periods (22:30 - 04:30). Error bars indicate standard error of the mean.