Adam M Graichen1, Richard W Vachet.
Abstract
A multiplexed method for performing MS/MS on multiple ions simultaneously in a miniature rectilinear ion trap (RIT) mass spectrometer has been developed. This method uses an ion encoding procedure that relies on the mass bias that exists when ions are externally injected into an RIT operated with only a single phase rf applied to one pair of electrodes. The ion injection profile under such conditions ions is Gaussian-like over a wide range of rf amplitudes, or low mass cutoff (LMCO) values, during ion accumulation. We show that this distribution is related to ion m/z and is likely caused by ions having an optimal range of pseudo-potential well depths for efficient trapping. Based on this observation, precursor ion intensity changes between two different injection LMCO values can be predicted, and these ion intensity changes are found to be carried through to their corresponding product ions, enabling multiplexed MS/MS spectra to be deconvoluted. © American Society for Mass Spectrometry, 2011Entities:
Year: 2011 PMID: 21472607 DOI: 10.1007/s13361-010-0068-6