Literature DB >> 25753973

Design and performance evaluation of a linear ion trap mass analyzer featuring half round rod electrodes.

Xiaoxu Li1, Xiaohua Zhang, Rujiao Yao, Yang He, Yongyong Zhu, Jie Qian.   

Abstract

A novel linear ion trap mass analyzer featuring half round rod electrodes (HreLIT) has been built. It is mainly composed of two pairs of stainless steel electrodes which have a cross-section of half round rod and a pair of end electrodes. The HreLIT has a simple structure and so it could be assembled by hand with relatively high mechanical accuracy. The external dimension of HreLIT is 50 mm × 29.5 mm × 28 mm (length × width × height) and its internal volume is about 3.8 cm(3). A home-made HreLIT mass spectrometer with three-stage vacuum system was built and the performance of HreLIT was characterized using reserpine solution and PPG standard solution. When the scan rate was 254 u/s, mass peak with FWHM of 0.14 u was achieved for ions with m/z 609, which corresponds to a mass resolution of 4350. The HreLIT was also operated at a low q value of 0.28 to extend its mass range. The experiment result showed a mass range of over 2800 u and the amplitude of radio frequency (rf) signal was only 1560 V (0-p). Three-stage tandem mass spectrometry was successfully performed in the HreLIT, and the collision-induced dissociation (CID) efficiencies of MS(2) (CID of ions with m/z 609) and MS(3) (CID of ions with m/z 448) were 78% and 59%, respectively.

Entities:  

Year:  2015        PMID: 25753973     DOI: 10.1007/s13361-015-1085-2

Source DB:  PubMed          Journal:  J Am Soc Mass Spectrom        ISSN: 1044-0305            Impact factor:   3.109


  24 in total

1.  Linear ion trap MS(n) of enzymatically synthesized 13C-labeled fructans revealing differentiating fragmentation patterns of β (1-2) and β (1-6) fructans and providing a tool for oligosaccharide identification in complex mixtures.

Authors:  Scott Harrison; Hong Xue; Geoff Lane; Silas Villas-Boas; Susanne Rasmussen
Journal:  Anal Chem       Date:  2012-01-12       Impact factor: 6.986

2.  Dual buffer gases for ion manipulation in a miniature ion trap mass spectrometer with a discontinuous atmospheric pressure interface.

Authors:  Tsung-Chi Chen; Wei Xu; Zheng Ouyang
Journal:  Rapid Commun Mass Spectrom       Date:  2011-11-15       Impact factor: 2.419

3.  Quadrupole mass filters with added hexapole fields.

Authors:  Zilan Xiao; XianZhen Zhao; D J Douglas
Journal:  Rapid Commun Mass Spectrom       Date:  2010-07-30       Impact factor: 2.419

4.  Rectilinear ion trap mass spectrometer with atmospheric pressure interface and electrospray ionization source.

Authors:  Qingyu Song; Sameer Kothari; Michael A Senko; Jae C Schwartz; Jonathan W Amy; George C Stafford; R Graham Cooks; Zheng Ouyang
Journal:  Anal Chem       Date:  2006-02-01       Impact factor: 6.986

5.  Monitoring of toxic compounds in air using a handheld rectilinear ion trap mass spectrometer.

Authors:  Adam Keil; Heriberto Hernandez-Soto; Robert J Noll; Miriam Fico; Liang Gao; Zheng Ouyang; R Graham Cooks
Journal:  Anal Chem       Date:  2008-01-09       Impact factor: 6.986

6.  Ion trap array mass analyzer: structure and performance.

Authors:  Xiaoxu Li; Gongyu Jiang; Chan Luo; Fuxing Xu; Yuanyuan Wang; Li Ding; Chuan-Fan Ding
Journal:  Anal Chem       Date:  2009-06-15       Impact factor: 6.986

7.  Handheld miniature ion trap mass spectrometers.

Authors:  Zheng Ouyang; Robert J Noll; R Graham Cooks
Journal:  Anal Chem       Date:  2009-04-01       Impact factor: 6.986

8.  Mass resolution of linear quadrupole ion traps with round rods.

Authors:  D J Douglas; N V Konenkov
Journal:  Rapid Commun Mass Spectrom       Date:  2014-11-15       Impact factor: 2.419

9.  Novel linear ion trap mass analyzer built with triangular electrodes.

Authors:  Yu Xiao; Zhengzhi Ding; Chongsheng Xu; Xinhua Dai; Xiang Fang; Chuan-Fan Ding
Journal:  Anal Chem       Date:  2014-06-05       Impact factor: 6.986

10.  Quadrupole mass filters with octopole fields.

Authors:  Chuanfan Ding; N V Konenkov; D J Douglas
Journal:  Rapid Commun Mass Spectrom       Date:  2003       Impact factor: 2.419

View more
  3 in total

Review 1.  Miniature and Fieldable Mass Spectrometers: Recent Advances.

Authors:  Dalton T Snyder; Christopher J Pulliam; Zheng Ouyang; R Graham Cooks
Journal:  Anal Chem       Date:  2015-10-21       Impact factor: 6.986

2.  Simulation of Unidirectional Ion Ejection in an Asymmetric Half-Round Rod Electrode Linear Ion Trap Mass Analyzer.

Authors:  HaiYan Wu; LiPeng Zhang; ZaiYue Zhang; Jie Qian; ShuGuang Zhang; YingJun Zhang; SaiJin Ge; XiaoXu Li
Journal:  J Am Soc Mass Spectrom       Date:  2018-07-11       Impact factor: 3.109

3.  Portable Digital Linear Ion Trap Mass Spectrometer Based on Separate-Region Corona Discharge Ionization Source for On-Site Rapid Detection of Illegal Drugs.

Authors:  Lingfeng Li; Tianyi Zhang; Deting Wang; Yunjing Zhang; Xingli He; Xiaozhi Wang; Peng Li
Journal:  Molecules       Date:  2022-05-30       Impact factor: 4.927

  3 in total