Literature DB >> 21457517

Assessment of fluid responsiveness during increased intra-abdominal pressure: keep the indices, but change the thresholds.

Benoit Tavernier1, Emmanuel Robin.   

Abstract

Dynamic variables of fluid responsiveness are useful guides for fluid management in patients under controlled positive pressure ventilation. In the previous issue of Critical Care, Jacques and colleagues show that these variables remain reliable predictors of fluid responsiveness in a porcine model of intra-abdominal hypertension, but threshold values are higher than during normal intra-abdominal pressure. Their threshold values, however, cannot be applied to clinical practice. This study suggests that intra-abdominal pressure must be measured in critically ill patients, and 'supranormal' values of dynamic variables should be analyzed with caution. The 'fluid responsive part' of an increased dynamic variable in such patients may be estimated by measuring its change during a fluid challenge.

Entities:  

Mesh:

Year:  2011        PMID: 21457517      PMCID: PMC3219334          DOI: 10.1186/cc10074

Source DB:  PubMed          Journal:  Crit Care        ISSN: 1364-8535            Impact factor:   9.097


In the previous issue of Critical Care, Jacques and colleagues [1] report that respiratory variations in stroke volume (SVV) and in arterial pulse pressure (PPV) remain reliable indices of fluid responsiveness in a porcine model of intra-abdominal hypertension (IAH). Threshold values, however, are higher than during normal intra-abdominal pressure (IAP), so that a 'supra-normal' SVV or PPV does not necessarily mean fluid responsiveness during IAH. Why is this study important? Dynamic indices such as PPV and SVV are accurate predictors of volume responsiveness in critically ill patients under controlled positive pressure ventilation, with an accuracy greater than that of traditional static indices of cardiac preload. The mean threshold values allowing optimal discrimination between fluid responders and non-responders were 12.5 ± 1.6% for PPV and 11.6 ± 1.9% for SVV in a recent systematic review of the literature [2]. These results, however, did not include patients where conditions prevented correct measurement of these indices (such as cardiac arrhythmias or spontaneous ventilation) or may have been associated with reduced accuracy (especially small tidal volumes or acute corpulmonale). A third category of circumstances, where dynamic indices remain robust indicators of fluid responsiveness if (and only if) used with different thresholds, may also be identified and constitute an additional refinement in PPV/SVV interpretation. It was suspected early on that the magnitude of tidal volume would influence the threshold value of functional hemodynamics [3]. More recently, Biais and coworkers [4] showed that prone position (for scoliosis surgery) does not alter the ability of both PPV and SVV to predict fluid responsiveness, but induces a significant increase in PPV and SVV, probably related to a decrease in static compliance of the respiratory system. Accordingly, the 'optimal' threshold value for PPV in this study was 11% in the supine position, and 15% in the prone position [4]. In recent years, elevated IAP has been increasingly recognized in medical and surgical critically ill patients [5]. Cardio-vascular dysfunction and failure are commonly encountered in the patient with IAH, and one of the bases of management is optimization of systemic perfusion and organ function, which includes accurate assessment of preload and preload responsiveness [5]. However, static indices of preload are difficult to interpret in patients with IAH [6]. Assessing fluid responsiveness with the passive leg raising maneuver in these patients results in false negative cases [7]. Experimental studies have shown that IAH increases dynamic variables [6] and, accordingly, 25% of patients with IAH may be non-responsive to volume infusion despite a PPV >12% [7]. Recently, Renner and colleagues [8] found in a porcine model that PPV, but not SVV (derived from pulse contour analysis), remained a sensitive and specific predictor of fluid responsiveness, although the threshold value for PPV increased from 11.5% (mean IAP = 7 mmHg) up to ≥20.5% (mean IAP = 26 mmHg). Thus, what does the study by Jacques and colleagues [1] add to previous knowledge? First, they confirm former results with a protocol that elegantly combines, for the first time, changes in IAP, blood withdrawal, and fluid loading. Second, they show that SVV, when directly measured using an ultrasound transit-time flow probe placed around the aortic root, is also predictive of fluid responsiveness in IAH. This contrasts with the results reported by Renner and colleagues [8], who acquired SVV via pulse contour monitoring, suggesting that the accuracy of the latter may be altered during IAH. Third, their data strongly suggest that, at least in their experimental conditions, the 'non-responsive' part of the increase in dynamic indices during IAH is due to an increase in right ventricular afterload. These new data should, however, be extrapolated to critically ill patients with caution. In this study, baseline PPV and SVV were much higher than in humans or in other experimental studies. High tidal volume as used in the study (13 ± 1 ml/kg in the presence of severe reduction in chest wall compliance) may explain, in part, these findings as well as the significant increase in right ventricular afterload, a result not found by Renner and colleagues [8]. In addition, the high IAP level used in the study (30 mmHg) corresponds to the more severe level (grade IV, IAH >25 mmHg [5]) of IAH in patients. Finally, a high but discriminative threshold could be identified by the authors because a precise and controlled value of IAP was induced in all animals, which would not be the case in clinical practice. Thus, what is demonstrated in this study is qualitative (there is an increase in dynamic indices in the presence of IAP, with one part responsive to fluid infusion) rather than quantitative. The threshold values reported by Jacques and colleagues have no direct clinical relevance. The consequences for clinical practice are thus that IAP must be measured in critically ill patients, and dynamic indices such as PPV should be used, but, in order to avoid excess of fluids, higher than classical (10 to 13%) thresholds should be considered when IAP is increased. As a gradual increase of threshold values with IAP is very likely, no precise value can be recommended so far. In a given patient with increased PPV, a fluid challenge may be performed and the resulting change in PPV/SVV quantified since only the 'preload-dependent part' of PPV is likely to be reduced with volume infusion. Whether the PPV value obtained at completion of such fluid load in such a patient may be used as the 'operational' threshold for the following hours of management, provided that other major determinants of PPV (IAP and tidal volume) are kept constant, can be reasonably hypothesized but deserves further validation.

Abbreviations

IAH: intra-abdominal hypertension; IAP: intra-abdominal pressure; PPV: respiratory variation in arterial pulse pressure; SVV: respiratory variation in stroke volume.

Competing interests

The authors declare that they have no competing interests.
  8 in total

1.  The effect of tidal volume and intravascular volume state on systolic pressure variation in ventilated dogs.

Authors:  A Szold; R Pizov; E Segal; A Perel
Journal:  Intensive Care Med       Date:  1989       Impact factor: 17.440

2.  Functional hemodynamics and increased intra-abdominal pressure: same thresholds for different conditions ...?

Authors:  Manu L N G Malbrain; Inneke de Laet
Journal:  Crit Care Med       Date:  2009-02       Impact factor: 7.598

3.  Abilities of pulse pressure variations and stroke volume variations to predict fluid responsiveness in prone position during scoliosis surgery.

Authors:  M Biais; O Bernard; J C Ha; C Degryse; F Sztark
Journal:  Br J Anaesth       Date:  2010-02-26       Impact factor: 9.166

4.  The passive leg-raising maneuver cannot accurately predict fluid responsiveness in patients with intra-abdominal hypertension.

Authors:  Yazine Mahjoub; Jérémie Touzeau; Norair Airapetian; Emmanuel Lorne; Mustapha Hijazi; Elie Zogheib; François Tinturier; Michel Slama; Hervé Dupont
Journal:  Crit Care Med       Date:  2010-09       Impact factor: 7.598

Review 5.  Dynamic changes in arterial waveform derived variables and fluid responsiveness in mechanically ventilated patients: a systematic review of the literature.

Authors:  Paul E Marik; Rodrigo Cavallazzi; Tajender Vasu; Amyn Hirani
Journal:  Crit Care Med       Date:  2009-09       Impact factor: 7.598

6.  Abdominal compartment syndrome.

Authors:  Manu Malbrain
Journal:  F1000 Med Rep       Date:  2009-11-16

7.  Influence of increased intra-abdominal pressure on fluid responsiveness predicted by pulse pressure variation and stroke volume variation in a porcine model.

Authors:  Jochen Renner; Matthias Gruenewald; Rene Quaden; Robert Hanss; Patrick Meybohm; Markus Steinfath; Jens Scholz; Berthold Bein
Journal:  Crit Care Med       Date:  2009-02       Impact factor: 7.598

8.  Pulse pressure variation and stroke volume variation during increased intra-abdominal pressure: an experimental study.

Authors:  Didier Jacques; Karim Bendjelid; Serge Duperret; Joëlle Colling; Vincent Piriou; Jean-Paul Viale
Journal:  Crit Care       Date:  2011-01-19       Impact factor: 9.097

  8 in total
  7 in total

Review 1.  Study on acute recent stage pancreatitis.

Authors:  Ye-Chen Feng; Min Wang; Feng Zhu; Ren-Yi Qin
Journal:  World J Gastroenterol       Date:  2014-11-21       Impact factor: 5.742

2.  Predictive values of pulse pressure variation and stroke volume variation for fluid responsiveness in patients with pneumoperitoneum.

Authors:  Marko Zlicar; Vesna Novak-Jankovic; Rok Blagus; Maurizio Cecconi
Journal:  J Clin Monit Comput       Date:  2017-11-17       Impact factor: 2.502

3.  Dynamic preload indicators decrease when the abdomen is opened.

Authors:  Martijn van Lavieren; Jeroen Veelenturf; Charlotte Hofhuizen; Marion van der Kolk; Johannes van der Hoeven; Peter Pickkers; Joris Lemson; Benno Lansdorp
Journal:  BMC Anesthesiol       Date:  2014-10-14       Impact factor: 2.217

4.  Predicting and measuring fluid responsiveness with echocardiography.

Authors:  Ashley Miller; Justin Mandeville
Journal:  Echo Res Pract       Date:  2016-05-24

Review 5.  Assessment of volume responsiveness during mechanical ventilation: recent advances.

Authors:  Xavier Monnet; Jean-Louis Teboul
Journal:  Crit Care       Date:  2013-03-19       Impact factor: 9.097

6.  Predicting fluid responsiveness with the passive leg raising test: don't be fooled by intra-abdominal hypertension!

Authors:  Andrea Minini; Paul Abraham; Manu L N G Malbrain
Journal:  Ann Transl Med       Date:  2020-06

7.  Validation of new marker of fluid responsiveness based on Doppler assessment of blood flow velocity in superior vena cava in mechanically ventilated pigs.

Authors:  Tomas Kovarnik; Miroslav Navratil; Jan Belohlavek; Mikulas Mlcek; Martin Chval; Zhi Chen; Stepan Jerabek; Otomar Kittnar; Ales Linhart
Journal:  Intensive Care Med Exp       Date:  2018-09-24
  7 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.