Localizing intracavitary brachytherapy applicators from cone-beam CT x-ray projections via a novel iterative forward projection matching algorithm.

Damodar Pokhrel1, Martin J Murphy, Dorin A Todor, Elisabeth Weiss, Jeffrey F Williamson.   

Abstract

PURPOSE: To present a novel method for reconstructing the 3D pose (position and orientation) of radio-opaque applicators of known but arbitrary shape from a small set of 2D x-ray projections in support of intraoperative brachytherapy planning.
METHODS: The generalized iterative forward projection matching (gIFPM) algorithm finds the six degree-of-freedom pose of an arbitrary rigid object by minimizing the sum-of-squared-intensity differences (SSQD) between the computed and experimentally acquired autosegmented projection of the objects. Starting with an initial estimate of the object's pose, gIFPM iteratively refines the pose parameters (3D position and three Euler angles) until the SSQD converges. The object, here specialized to a Fletcher-Weeks intracavitary brachytherapy (ICB) applicator, is represented by a fine mesh of discrete points derived from complex combinatorial geometric models of the actual applicators. Three pairs of computed and measured projection images with known imaging geometry are used. Projection images of an intrauterine tandem and colpostats were acquired from an ACUITY cone-beam CT digital simulator. An image postprocessing step was performed to create blurred binary applicators only images. To quantify gIFPM accuracy, the reconstructed 3D pose of the applicator model was forward projected and overlaid with the measured images and empirically calculated the nearest-neighbor applicator positional difference for each image pair.
RESULTS: In the numerical simulations, the tandem and colpostats positions (x,y,z) and orientations (alpha, beta, gamma) were estimated with accuracies of 0.6 mm and 2 degrees, respectively. For experimentally acquired images of actual applicators, the residual 2D registration error was less than 1.8 mm for each image pair, corresponding to about 1 mm positioning accuracy at isocenter, with a total computation time of less than 1.5 min on a 1 GHz processor.
CONCLUSIONS: This work describes a novel, accurate, fast, and completely automatic method to localize radio-opaque applicators of arbitrary shape from measured 2D x-ray projections. The results demonstrate approximately 1 mm accuracy while compared against the measured applicator projections. No lateral film is needed. By localizing the applicator internal structure as well as radioactive sources, the effect of intra-applicator and interapplicator attenuation can be included in the resultant dose calculations. Further validation tests using clinically acquired tandem and colpostats images will be performed for the accurate and robust applicator/sources localization in ICB patients.

Entities:  

Mesh:

Year:  2011        PMID: 21452744      PMCID: PMC3045416          DOI: 10.1118/1.3544661

Source DB:  PubMed          Journal:  Med Phys        ISSN: 0094-2405            Impact factor:   4.071


  33 in total

1.  Catheter autoreconstruction in computed tomography based brachytherapy treatment planning.

Authors:  N Milickovic; S Giannouli; D Baltas; M Lahanas; C Kolotas; N Zamboglou; N Uzunoglu
Journal:  Med Phys       Date:  2000-05       Impact factor: 4.071

2.  On the validity of the superposition principle in dose calculations for intracavitary implants with shielded vaginal colpostats.

Authors:  J Markman; J F Williamson; J F Dempsey; D A Low
Journal:  Med Phys       Date:  2001-02       Impact factor: 4.071

3.  The American Brachytherapy Society recommendations for low-dose-rate brachytherapy for carcinoma of the cervix.

Authors:  Subir Nag; Clifford Chao; Beth Erickson; Jeffery Fowler; Nilendu Gupta; Alvaro Martinez; Bruce Thomadsen
Journal:  Int J Radiat Oncol Biol Phys       Date:  2002-01-01       Impact factor: 7.038

4.  Survey of the use of the ICRU 38 in recording and reporting cervical cancer brachytherapy.

Authors:  R Pötter; E Van Limbergen; N Gerstner; A Wambersie
Journal:  Radiother Oncol       Date:  2001-01       Impact factor: 6.280

5.  Dosimetry of 137Cs sources with the Fletcher-Suit gynecological applicator.

Authors:  W L Saylor; M Dillard
Journal:  Med Phys       Date:  1976 Mar-Apr       Impact factor: 4.071

6.  Reconstruction of brachytherapy seed positions and orientations from cone-beam CT x-ray projections via a novel iterative forward projection matching method.

Authors:  Damodar Pokhrel; Martin J Murphy; Dorin A Todor; Elisabeth Weiss; Jeffrey F Williamson
Journal:  Med Phys       Date:  2011-01       Impact factor: 4.071

7.  Prospects for quantitative computed tomography imaging in the presence of foreign metal bodies using statistical image reconstruction.

Authors:  Jeffrey F Williamson; Bruce R Whiting; Jasenka Benac; Ryan J Murphy; G James Blaine; Joseph A O'Sullivan; David G Politte; Donald L Snyder
Journal:  Med Phys       Date:  2002-10       Impact factor: 4.071

8.  Cervical tumor control evaluated with ICRU 38 reference volumes and integrated reference air kerma.

Authors:  P W Grigsby; J F Williamson; K S Clifford Chao; C A Perez
Journal:  Radiother Oncol       Date:  2001-01       Impact factor: 6.280

9.  An algorithm for the reduction of metal clip artifacts in CT reconstructions.

Authors:  G H Glover; N J Pelc
Journal:  Med Phys       Date:  1981 Nov-Dec       Impact factor: 4.071

10.  Accurate localization of intracavitary brachytherapy applicators from 3D CT imaging studies.

Authors:  F A Lerma; J F Williamson
Journal:  Med Phys       Date:  2002-03       Impact factor: 4.071

View more

北京卡尤迪生物科技股份有限公司 © 2022-2023.