| Literature DB >> 1264041 |
Abstract
Measurements were made in water about the tandem and ovoid structures of the Fletcher-Suit applicator loaded with a 137Cs source. The dose rate was found to vary as much as 25% as a result of differential absorption in the ovoid. Empirically based trigonometric expressions are suggested which provide the absorption correction as a function of the two angles of a spherical coordinate system. Computer calculated dose rates, with and without the absorption corrections, are compared.Entities:
Mesh:
Substances:
Year: 1976 PMID: 1264041 DOI: 10.1118/1.594277
Source DB: PubMed Journal: Med Phys ISSN: 0094-2405 Impact factor: 4.071