Improved hardware for higher spatial resolution strain-encoded (SENC) breast MRI for strain measurements.

Ahmed A Harouni1, Jakir Hossain, Michael A Jacobs, Nael F Osman.   

Abstract

RATIONALE AND
OBJECTIVES: Early detection of breast lesions using mammography has resulted in lower mortality rates. However, some breast lesions are mammography occult, and magnetic resonance imaging (MRI) is recommended, but it has lower specificity. It is possible to achieve higher specificity by using strain-encoded (SENC) MRI and/or magnetic resonance elastography. SENC breast MRI can measure the strain properties of breast tissue. Similarly, magnetic resonance elastography is used to measure the elasticity (ie, shear stiffness) of different tissue compositions interrogating the tissue mechanical properties. Reports have shown that malignant tumors are three to 13 times stiffer than normal tissue and benign tumors.
MATERIALS AND METHODS: The investigators have developed a SENC breast hardware device capable of periodically compressing the breast, thus allowing for longer scanning time and measuring the strain characteristics of breast tissue. This hardware enables the use of SENC MRI with high spatial resolution (1 × 1 × 5 mm(3)) instead of fast SENC imaging. Simple controls and multiple safety measures were added to ensure accurate, repeatable, and safe in vivo experiments.
RESULTS: Phantom experiments showed that SENC breast MRI has higher signal-to-noise ratio and contrast-to-noise ratio than fast SENC imaging under different scanning resolutions. Finally, the SENC breast device reproducibility measurements resulted in a difference of <1 mm with a 1% strain difference.
CONCLUSIONS: SENC breast magnetic resonance images have higher signal-to-noise ratio and contrast-to-noise ratios than fast SENC images. Thus, combining SENC breast strain measurements with diagnostic breast MRI to differentiate benign from malignant lesions could potentially increase the specificity of diagnosis in the clinical setting.
Copyright © 2011 AUR. Published by Elsevier Inc. All rights reserved.

Entities:  

Mesh:

Year:  2011        PMID: 21440464      PMCID: PMC4197851          DOI: 10.1016/j.acra.2011.01.006

Source DB:  PubMed          Journal:  Acad Radiol        ISSN: 1076-6332            Impact factor:   3.173


  39 in total

1.  MR elastography of the prostate: initial in-vivo application.

Authors:  J Kemper; R Sinkus; J Lorenzen; C Nolte-Ernsting; A Stork; G Adam
Journal:  Rofo       Date:  2004-08

2.  An integrated system for real-time detection of stiff masses with a single compression.

Authors:  Ahmed S Fahmy; Axel Krieger; Nael F Osman
Journal:  IEEE Trans Biomed Eng       Date:  2006-07       Impact factor: 4.538

3.  Comparison of phase-difference and complex-difference processing in phase-contrast MR angiography.

Authors:  M A Bernstein; Y Ikezaki
Journal:  J Magn Reson Imaging       Date:  1991 Nov-Dec       Impact factor: 4.813

Review 4.  Phase contrast cine magnetic resonance imaging.

Authors:  N J Pelc; R J Herfkens; A Shimakawa; D R Enzmann
Journal:  Magn Reson Q       Date:  1991-10

5.  Magnetic resonance elastography for the noninvasive staging of liver fibrosis.

Authors:  Laurent Huwart; Christine Sempoux; Eric Vicaut; Najat Salameh; Laurence Annet; Etienne Danse; Frank Peeters; Leon C ter Beek; Jacques Rahier; Ralph Sinkus; Yves Horsmans; Bernard E Van Beers
Journal:  Gastroenterology       Date:  2008-04-04       Impact factor: 22.682

6.  Improved myocardial tagging contrast.

Authors:  S E Fischer; G C McKinnon; S E Maier; P Boesiger
Journal:  Magn Reson Med       Date:  1993-08       Impact factor: 4.668

7.  Characterization of neoplastic and normal human breast tissues with in vivo (1)H MR spectroscopy.

Authors:  K A Kvistad; I J Bakken; I S Gribbestad; B Ehrnholm; S Lundgren; H E Fjøsne; O Haraldseth
Journal:  J Magn Reson Imaging       Date:  1999-08       Impact factor: 4.813

8.  Elastic moduli of normal and pathological human breast tissues: an inversion-technique-based investigation of 169 samples.

Authors:  Abbas Samani; Judit Zubovits; Donald Plewes
Journal:  Phys Med Biol       Date:  2007-02-16       Impact factor: 3.609

9.  Assessment of liver viscoelasticity using multifrequency MR elastography.

Authors:  Patrick Asbach; Dieter Klatt; Uwe Hamhaber; Jürgen Braun; Rajan Somasundaram; Bernd Hamm; Ingolf Sack
Journal:  Magn Reson Med       Date:  2008-08       Impact factor: 4.668

10.  MRI evaluation of the contralateral breast in women with recently diagnosed breast cancer.

Authors:  Constance D Lehman; Constantine Gatsonis; Christiane K Kuhl; R Edward Hendrick; Etta D Pisano; Lucy Hanna; Sue Peacock; Stanley F Smazal; Daniel D Maki; Thomas B Julian; Elizabeth R DePeri; David A Bluemke; Mitchell D Schnall
Journal:  N Engl J Med       Date:  2007-03-28       Impact factor: 91.245

View more
  3 in total

1.  Strain-encoded breast MRI in phantom and ex vivo specimens with histological validation: preliminary results.

Authors:  Ahmed A Haruoni; Jakir Hossain; Riham El Khouli; Kant M Matsuda; David A Bluemke; Nael F Osman; Michael A Jacobs
Journal:  Med Phys       Date:  2012-12       Impact factor: 4.071

2.  Assessment of liver fibrosis using fast strain-encoded MRI driven by inherent cardiac motion.

Authors:  Ahmed A Harouni; Ahmed M Gharib; Nael F Osman; Caryn Morse; Theo Heller; Khaled Z Abd-Elmoniem
Journal:  Magn Reson Med       Date:  2014-07-31       Impact factor: 4.668

3.  Strain-encoded magnetic resonance: a method for the assessment of myocardial deformation.

Authors:  Grigorios Korosoglou; Sorin Giusca; Nina P Hofmann; Amit R Patel; Tomas Lapinskas; Burkert Pieske; Henning Steen; Hugo A Katus; Sebastian Kelle
Journal:  ESC Heart Fail       Date:  2019-04-25
  3 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.