Haemoglobin A1c: evaluation of a new HbA1c point-of-care analyser Bio-Rad in2it in comparison with the DCA 2000 and central laboratory analysers.

Chin-Pin Yeo1, Carol Hui-Chen Tan, Edward Jacob.   

Abstract

BACKGROUND: Point-of-care-testing (POCT) of haemoglobin Alc (HbA1c) is popular due to its fast turnaround of results in the outpatient setting. The aim of this project was to evaluate the performance of a new HbA1c POCT analyser, the Bio-Rad in2it, and compare it with the Siemens DCA 2000, Bio-Rad Variant II and Roche Tina-quant HbA1c Gen 2 assay on the cobas c501.
METHODS: Imprecision of the four methods were compared by computing total imprecision from within-run and between-run data. A total of 80 samples were also compared and analysed by Deming regression and Altman-Bland difference test.
RESULTS: Study of total imprecision of the in2it at HBA1c levels of 6.0% and 10.4% produced a coefficient of variation (%CV) of 3.8% and 3.7%, respectively. These results were more favourable as compared with the DCA 2000 but did not match the low imprecision of the central laboratory methods, the Bio-Rad Variant II and the Roche cobas c501. Comparison between the in2it and the central laboratory analysers, Bio-Rad variant II and cobas c501, revealed positive bias of 12% and 10%, respectively, supported by corresponding Deming regression equation slopes of +1.18 and +1.14. Comparison between the DCA 2000 and the central laboratory analysers revealed a bias that became increasingly positive with rising HbA1c concentrations with Deming regression analysis also revealing proportional and constant differences.
CONCLUSIONS: The in2it is a suitable POCT analyser for HbA1c but its less than ideal precision performance and differences with the central laboratory analysers must be communicated to and noted by the users.

Entities:  

Mesh:

Substances:

Year:  2009        PMID: 19487414     DOI: 10.1258/acb.2009.009008

Source DB:  PubMed          Journal:  Ann Clin Biochem        ISSN: 0004-5632            Impact factor:   2.057


  4 in total

Review 1.  Point-of-Care Hemoglobin A1c Testing: An Evidence-Based Analysis.

Authors: 
Journal:  Ont Health Technol Assess Ser       Date:  2014-07-01

2.  [Capillary HbA1c determination on type 2 diabetes patients in a primary health centre].

Authors:  María Teresa Carrera Font; María Claustre Solé Brichs; María Clara Sala Álvarez; Jose María Navarro Olivella; Josefina Servent Turó; María Pilar Felipe Fernández
Journal:  Aten Primaria       Date:  2011-03-15       Impact factor: 1.137

3.  The Relationship between Generalized and Abdominal Obesity with Diabetic Kidney Disease in Type 2 Diabetes: A Multiethnic Asian Study and Meta-Analysis.

Authors:  Ryan Eyn Kidd Man; Alfred Tau Liang Gan; Eva Katie Fenwick; Preeti Gupta; Mark Yu Zheng Wong; Tien Yin Wong; Gavin Siew Wei Tan; Boon Wee Teo; Charumathi Sabanayagam; Ecosse Luc Lamoureux
Journal:  Nutrients       Date:  2018-11-05       Impact factor: 5.717

4.  Prevalence, determinants and association of unawareness of diabetes, hypertension and hypercholesterolemia with poor disease control in a multi-ethnic Asian population without cardiovascular disease.

Authors:  Ryan E K Man; Alvin Hong Wei Gan; Eva K Fenwick; Alfred Tau Liang Gan; Preeti Gupta; Charumathi Sabanayagam; Nicholas Tan; Kah Hie Wong; Tien Yin Wong; Ching-Yu Cheng; Ecosse L Lamoureux
Journal:  Popul Health Metr       Date:  2019-12-05
  4 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.