Literature DB >> 21409611

Development of quality control system for flat-panel detectors.

Atsushi Teramoto1, Takahiko Kajihara, Shoichi Suzuki, Kazuo Kinoshita, Masatoshi Tsuzaka, Hiroshi Fujita.   

Abstract

The characteristics of flat-panel detectors (FPD) are degraded by exposure to radiation. Degradation in a FPD progresses locally and has a nonlinear relationship to the radiation dose. In order to manage FPD systems properly, one must perform quality control (QC) such as evaluation of image degradation. However, no evaluation method for degradation has been established. In this paper, we first review the structure and degradation mechanism of FPDs, and then we propose a daily QC system for FPDs. To evaluate the degradation of FPDs, we investigated the number of defective pixels and lines, as well as the offset level of the pixel output. Furthermore, we developed daily QC software for FPD that can evaluate the image quality and is operationally simple. In the experiments, an indirect-conversion type FPD was evaluated by our proposed system. The offset level of FPD increased exponentially with X-ray exposure; no trends were seen for the number of defective pixels and defective lines. The required time for the evaluation of an FPD was about 1 min, and no special skills were needed for the analysis. These results indicate that our system may be useful for daily QC of FPDs.

Mesh:

Year:  2011        PMID: 21409611     DOI: 10.1007/s12194-011-0117-2

Source DB:  PubMed          Journal:  Radiol Phys Technol        ISSN: 1865-0333


  5 in total

1.  An experimental comparison of detector performance for direct and indirect digital radiography systems.

Authors:  Ehsan Samei; Michael J Flynn
Journal:  Med Phys       Date:  2003-04       Impact factor: 4.071

2.  Image quality in two phosphor-based flat panel digital radiographic detectors.

Authors:  Ehsan Samei
Journal:  Med Phys       Date:  2003-07       Impact factor: 4.071

3.  Performance of a 41 x 41 cm2 amorphous silicon flat panel x-ray detector designed for angiographic and R&F imaging applications.

Authors:  Paul R Granfors; Richard Aufrichtig; George E Possin; Brian W Giambattista; Zhong S Huang; Jianqiang Liu; Bing Ma
Journal:  Med Phys       Date:  2003-10       Impact factor: 4.071

4.  Signal, noise power spectrum, and detective quantum efficiency of indirect-detection flat-panel imagers for diagnostic radiology.

Authors:  J H Siewerdsen; L E Antonuk; Y el-Mohri; J Yorkston; W Huang; I A Cunningham
Journal:  Med Phys       Date:  1998-05       Impact factor: 4.071

5.  Radiation damage of amorphous silicon, thin-film, field-effect transistors.

Authors:  J M Boudry; L E Antonuk
Journal:  Med Phys       Date:  1996-05       Impact factor: 4.071

  5 in total
  2 in total

1.  Simulation system for understanding the lag effect in fluoroscopic images.

Authors:  Rie Tanaka; Hiroki Kawashima; Katsuhiro Ichikawa; Kosuke Matsubara; Hiroji Iida; Shigeru Sanada
Journal:  Radiol Phys Technol       Date:  2012-12-29

2.  Investigation of image lag and modulation transfer function in fluoroscopy images obtained with a dynamic flat-panel detector.

Authors:  Hiroki Kawashima; Rie Tanaka; Katsuhiro Ichikawa; Kosuke Matsubara; Hiroji Iida; Shigeru Sanada
Journal:  Radiol Phys Technol       Date:  2013-04-09
  2 in total