Length of endodontic files measured in digital radiographs with and without noise-suppression filters: an ex-vivo study.

D D Brüllmann1, B Röhrig, S L Sulayman, R Schulze.   

Abstract

OBJECTIVES: The aim of this study is to evaluate if theoretically possible edge shifts induced by noise-suppression filters potentially occur on objects found in digital radiographs. Most manufacturers carry out noise-suppression filtering of their images before they are displayed to the user. It is not usually possible for the user to disrupt the function of the filters. The use of these filters can lead to deletion of small image structures.
METHODS: K-files (ISO size 06, 08, 10 and 15) were placed in the root canals of 6 human teeth located in cadaver jaw segments. File tip positions were measured on original and filtered digital images by three observers. The file position was marked on each filtered image and compared with the unfiltered ones.
RESULTS: The 5 × 5 pixel-sized median and mean filters caused the largest underestimation of measured lengths between -7.87 pixels and -10.8 pixels (-306.93 μm and -421.2 μm). Maximum standard deviation for length differences was found for the calculated position ( = gold standard) and the original unfiltered images with 13.31 pixels. The standard deviation found for the 5 × 5 mean and median filter was 7.62 pixels and 8.68 pixels.
CONCLUSION: Different studies showed that noise-suppression filters can induce edge shifts in diagnostic radiographs. The high standard deviations found for length differences between the defined gold-standard and the original images indicate that theoretical edge shifts may not be clinically relevant for length measurements of endodontic files.

Entities:  

Mesh:

Year:  2011        PMID: 21346084      PMCID: PMC3611456          DOI: 10.1259/dmfr/84700141

Source DB:  PubMed          Journal:  Dentomaxillofac Radiol        ISSN: 0250-832X            Impact factor:   2.419


  14 in total

1.  Comparison of observer performance in determining the position of endodontic files with physical measures in the evaluation of dental X-ray imaging systems.

Authors:  R H Vandre; J C Pajak; H Abdel-Nabi; T T Farman; A G Farman
Journal:  Dentomaxillofac Radiol       Date:  2000-07       Impact factor: 2.419

2.  A comparison of phosphor-plate digital images with conventional radiographs for the perceived clarity of fine endodontic files and periapical lesions.

Authors:  Lara T Friedlander; Robert M Love; Nicholas P Chandler
Journal:  Oral Surg Oral Med Oral Pathol Oral Radiol Endod       Date:  2002-03

Review 3.  Radiographic digital image enhancement. Part II: transform domain techniques.

Authors:  M Analoui
Journal:  Dentomaxillofac Radiol       Date:  2001-03       Impact factor: 2.419

4.  Image processing and enhancement provided by commercial dental software programs.

Authors:  T M Lehmann; E Troeltsch; K Spitzer
Journal:  Dentomaxillofac Radiol       Date:  2002-07       Impact factor: 2.419

5.  In vitro comparison of root-canal measurements with conventional and digital radiology.

Authors:  A Lozano; L Forner; C Llena
Journal:  Int Endod J       Date:  2002-06       Impact factor: 5.264

6.  A comparison of 18 different x-ray detectors currently used in dentistry.

Authors:  Allan G Farman; Taeko T Farman
Journal:  Oral Surg Oral Med Oral Pathol Oral Radiol Endod       Date:  2005-04

7.  An optimal nonorthogonal separation of the anisotropic Gaussian convolution filter.

Authors:  Christoph H Lampert; Oliver Wirjadi
Journal:  IEEE Trans Image Process       Date:  2006-11       Impact factor: 10.856

8.  Dose reduction of two digital sensor systems measuring file lengths.

Authors:  X L Velders; G C Sanderink; P F van der Stelt
Journal:  Oral Surg Oral Med Oral Pathol Oral Radiol Endod       Date:  1996-05

Review 9.  Efficacy of digital intra-oral radiography in clinical dentistry.

Authors:  C H Versteeg; G C Sanderink; P F van der Stelt
Journal:  J Dent       Date:  1997 May-Jul       Impact factor: 4.379

10.  Radiographic determination of canal length direct digital radiography versus conventional radiography.

Authors:  R T Hedrick; S B Dove; D D Peters; W D McDavid
Journal:  J Endod       Date:  1994-07       Impact factor: 4.171

View more
  2 in total

1.  Effects of five levels of noise reduction applied to indirect digital radiography on diagnostic accuracy of external apical root resorption.

Authors:  Sanaz Sharifi Shoshtari; Seyed Arman Mohagheghi; Nastaran Farhadi; Negin Kheradmand; Lida Naderi
Journal:  Dent Res J (Isfahan)       Date:  2020-03-17

2.  Effect of digital noise reduction on the accuracy of endodontic file length determination.

Authors:  Mojdeh Mehdizadeh; Abbas Ali Khademi; Ali Shokraneh; Nastaran Farhadi
Journal:  Imaging Sci Dent       Date:  2013-09-23
  2 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.