Dose reduction of two digital sensor systems measuring file lengths.

X L Velders1, G C Sanderink, P F van der Stelt.   

Abstract

OBJECTIVE: The aim of this study was to determine the effect of dose reduction on the image quality of two sensor systems with automatic greyscale adjustment. STUDY
DESIGN: Two digital sensor systems, Sidexis (Siemens, Bensheim, Germany) and Digora (Soredex, Helsinki, Finland), were tested. The image quality of the systems was determined by comparing lengths of endodontic files (025, 020, 015, and 010 files) and a premolar root in the digital images with lengths on conventional radiographic films. For the experiments the exposure was changed to 100%, 50%, 25%, 12.5%, 6.25%, and 3.125% of that for Ektaspeed films.
RESULTS: The lengths of 025 and 020 files and the premolar root in the digital images were comparable with the lengths on films even if the exposure was reduced to 6% of that used for Ektaspeed films. The lengths of 015 files reduced as the exposure was reduced. The lengths of 010 files were significantly shorter than the lengths on films.
CONCLUSIONS: With the systems tested, a dose reduction of approximately 95% compared with Ektaspeed films is possible to determine the lengths of a premolar root and 025 and 020 files. For thinner objects a dose reduction is questionable.

Mesh:

Year:  1996        PMID: 8734712     DOI: 10.1016/s1079-2104(96)80057-6

Source DB:  PubMed          Journal:  Oral Surg Oral Med Oral Pathol Oral Radiol Endod        ISSN: 1079-2104


  5 in total

1.  Detection of artificial occlusal caries in a phosphor imaging plate system with two types of LCD monitors versus three different films.

Authors:  Mehmet Ilgüy; Semanur Dinçer; Dilhan Ilgüy; Gündüz Bayirli
Journal:  J Digit Imaging       Date:  2008-10-24       Impact factor: 4.056

2.  Length of endodontic files measured in digital radiographs with and without noise-suppression filters: an ex-vivo study.

Authors:  D D Brüllmann; B Röhrig; S L Sulayman; R Schulze
Journal:  Dentomaxillofac Radiol       Date:  2011-03       Impact factor: 2.419

3.  The usage of digital radiography and cone beam computed tomography among Turkish dentists.

Authors:  Semanur Dölekoğlu; E Fişekçioğlu; M İlgüy; D İlgüy
Journal:  Dentomaxillofac Radiol       Date:  2011-09       Impact factor: 2.419

4.  Comparative Evaluation of Intraoral and Extraoral Periapical Radiographic Techniques in Determination of Working Length: An In Vivo Study.

Authors:  Arvind Sridhara; Sapna Konde; Sunil Raj Noojadi; Narayan Chandra Kumar; Ashwitha C Belludi
Journal:  Int J Clin Pediatr Dent       Date:  2020 May-Jun

5.  Comparison of the accuracy of conventional and digital radiography in root canal working length determination: An invitro study.

Authors:  Mohammad Mahdi Yaghooti Khorasani; Hamed Ebrahimnejad
Journal:  J Dent Res Dent Clin Dent Prospects       Date:  2017-09-20
  5 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.