| Literature DB >> 21330975 |
Maria Francesca Spadea1, Guido Baroni, David P Gierga, Julie C Turcotte, George T Y Chen, Gregory C Sharp.
Abstract
The purpose of this study is to assess the temporal and reconstruction accuracy of a surface imaging system, the GateCT under ideal conditions, and compare the device with a commonly used respiratory surrogate: the Varian RPM. A clinical CT scanner, run in cine mode, was used with two optical devices, GateCT and RPM, to detect respiratory motion. A radiation detector, GM-10, triggers the X-ray on/off to GateCT system, while the RPM is directly synchronized with the CT scanner through an electronic connection. Two phantoms were imaged: the first phantom translated on a rigid plate along the anterior-posterior (AP) direction, and was used to assess the temporal synchronization of each optical system with the CT scanner. The second phantom, consisting of five spheres translating 3 cm peak-to-peak in the superior-inferior direction, was used to assess the quality of rebinned images created by GateCT and RPM. Calibration assessment showed a nearly perfect synchronization with the scanner for both the RPM and GateCT systems, thus demonstrating the good performance of the radiation detector. Results for the volume rebinning test showed discrepancies in volumes for the 3D reconstruction (compared to ground truth) of up to 36% for GateCT and up to 40% for RPM. No statistical difference was proven between the two systems in volume sorting. Errors are mainly due to phase detection inaccuracies and to the large motion of the phantom. This feasibility study assessed the consistency of two optical systems in synchronizing the respiratory signal with the image acquisition. A new patient protocol based on both RPM and GateCT will be soon started.Entities:
Mesh:
Year: 2010 PMID: 21330975 PMCID: PMC5718580 DOI: 10.1120/jacmp.v12i1.3288
Source DB: PubMed Journal: J Appl Clin Med Phys ISSN: 1526-9914 Impact factor: 2.102
Figure 1Experimental setup for the assessment of the calibration of CT scanner and optical systems for 4D image reconstruction.
Figure 2Motion amplitude of the square set measured on the images (ground truth) and detected by RPM and GCT. The object extended over nine couch positions, from the second to the tenth one, as can be noted from the ground truth pattern. Both rpm‐amp and gct‐amp present a time offset with respect to the scanner due to different clocks of the systems. The offset was computed by aligning the X‐ray on/off pulses.
Figure 3Scanned object for evaluation of 4D volume reconstruction. The middle and the right panel refers to RPM and GCT reconstruction, respectively (exhale phase).
Figure 4Comparison of RPM and GCT motion amplitude vs ground truth after removing the clock offset. Amplitudes were aligned by their respective mean value.
X‐ray pulse width for the three systems, mean (std) [sec]. Data are averaged over nine couch positions.
|
|
|
|
|---|---|---|
| 4.3999 | 4.4045 | 4.4612 |
| (0) | (0.0005) | (0.0160) |
Selected threshold for each sphere and comparison between computed and real volume.
|
|
|
|
|
| |
|---|---|---|---|---|---|
| Threshold [HU] |
|
|
|
|
|
| Ground Truth [ | 54.95 | 25.09 | 7.64 | 1.99 | 2.19 |
| Computed Volume [ | 55.00 | 25.10 | 7.60 | 2.00 | 2.20 |
| Computed vs. Ground Truth [ |
|
| 0.04 |
|
|
Real and estimated volume after image sorting. Volume values and discrepancies were averaged on 10 phases, mean (std)[].
|
|
|
|
|
| |
|---|---|---|---|---|---|
| Ground Truth | 55.0 | 25.1 | 7.6 | 2.0 | 2.2 |
| RPM | 51.8 (3.0) | 24.7 (3.3) | 7.7 (0.8) | 1.8 (0.3) | 2.0 (0.3) |
| GCT | 51.7 (2.9) | 23.6 (2.0) | 8.0 (1.0) | 2.0 (0.2) | 2.0 (0.3) |
| RPM vs Ground Truth | ‐3.1 (3.0) | ‐0.4 (3.3) | 0.0 (0.8) | ‐0.2 (0.3) | 0.2 (0.3) |
| GCT vs Ground Truth | ‐3.3 (2.9) | ‐1.5 (2.5) | 0.4 (2.0) | ‐0.1 (0.2) | 0.2 (0.3) |
Comparison in terms of volume discrepancies and CM between RPM and GCT in each phase.
|
|
|
|
|
|
| |||||
|---|---|---|---|---|---|---|---|---|---|---|
|
|
|
|
|
|
|
|
|
|
| |
| 1 | 0.0 | 0.0 | 0.0 | 0.0 | 0.0 | 0.0 | 0.0 | 0.0 | 0.0 | 0.0 |
| 2 | 0.0 | 0.0 | 0.0 | 0.0 | 0.0 | 0.0 | 0.0 | 0.0 | 0.0 | 0.0 |
| 3 | 0.0 | 0.0 |
| 0.7 | 0.1 | 0.2 |
| 1.3 | 0.0 | 0.0 |
| 4 | 0.0 | 0.0 |
| 2.9 | 0.0 | 0.0 | 0.8 | 2.5 | 0.0 | 0.0 |
| 5 | 0.0 | 0.0 |
| 4.8 | 0.0 | 0.0 | 0.2 | 1.0 | 0.0 | 0.0 |
| 6 | 0.0 | 0.0 | 0.0 | 0.0 | 3.4 | 5.3 | 0.0 | 0.0 | 0.0 | 0.0 |
| 7 | 0.0 | 0.0 | 0.0 | 0.0 | 0.0 | 0.0 |
| 0.7 | 0.0 | 0.0 |
| 8 |
| 0.1 | 0.0 | 0.6 | 0.0 | 0.3 | 0.0 | 0.0 | 0.0 | 0.0 |
| 9 | 0.0 | 0.0 | 0.0 | 0.0 | 0.0 | 0.0 | 0.0 | 0.0 | 0.0 | 0.0 |
| 10 |
| 0.0 | 0.0 | 0.0 | 0.0 | 0.0 | 0.6 | 2.3 | 0.0 | 0.0 |
| mean |
|
|
|
| 0.4 | 0.4 | 0.1 | 0.1 | 0.0 | 0.0 |
| std | 0.3 | 0.3 | 2.4 | 2.4 | 1.1 | 1.1 | 0.4 | 0.4 | 0.0 | 0.0 |
Figure 5Relative error in volume reconstruction for both RPM and GCT sorting ((1‐RPM/GroundTruth)%) and (1‐GCT%/ GroundTruth)%).