Literature DB >> 21327489

Durability of repaired pacemaker leads in the pediatric population.

Sarah Chambers1, Alex Rusanov, Henry M Spotnitz, Eric S Silver, Leonardo Liberman.   

Abstract

PURPOSE: Lead failure is a common complication of pacemaker placement in the pediatric population and is often due to insulation breaks or lead fracture. Usually these leads are replaced with new leads. This study describes the surgical technique of lead repair and investigates the longevity of such repairs.
METHODS: Retrospective review of lead repairs performed in the pediatric population (<21 years old) between December 1995 and December 2008. Operative records, pre-operative interrogation data, and most recent follow-up data were reviewed.
RESULTS: Twenty-one lead repairs were performed on 17 patients. Sixteen were ventricular leads and five were atrial leads. There were 14 insulation defect repairs, six new lead tip splicings, and one lead extender splicing. The repaired leads were followed for a mean time of 3.6 ± 3.0 years (range, 0.3-10.6 years). The freedom from repaired lead failure by Kaplan-Meier survival analysis was 89% at 5 years and 22% at 10 years. One atrial lead and two ventricular leads have been replaced since the repairs. One atrial lead has had a repeat insulation break repair since the initial repair. One of the five atrial leads had an elevated threshold at follow-up. Five of 16 ventricular leads had elevated thresholds at follow-up. All leads had normal impedance at follow-up.
CONCLUSION: The option of pacemaker lead repair, if surgically feasible, offers an alternative to lead replacement in the pediatric population and may extend the longevity of pacemaker leads in this population.

Entities:  

Mesh:

Year:  2011        PMID: 21327489     DOI: 10.1007/s10840-011-9543-7

Source DB:  PubMed          Journal:  J Interv Card Electrophysiol        ISSN: 1383-875X            Impact factor:   1.900


  13 in total

1.  Cardiac lead extraction with a novel locking stylet.

Authors:  C Kennergren; R H Schaerf; T D Sellers; B L Wilkoff; C L Byrd; G F Tyres; S Coe; C W Coates; C Reiser
Journal:  J Interv Card Electrophysiol       Date:  2000-12       Impact factor: 1.900

Review 2.  Cardiac pacing in children.

Authors:  L Bevilacqua; A Hordof
Journal:  Curr Opin Cardiol       Date:  1998-01       Impact factor: 2.161

3.  Twenty years experience with pediatric pacing: epicardial and transvenous stimulation.

Authors:  J S Sachweh; J F Vazquez-Jimenez; F A Schöndube; S H Daebritz; H Dörge; E G Mühler; B J Messmer
Journal:  Eur J Cardiothorac Surg       Date:  2000-04       Impact factor: 4.191

4.  Transvenous pacing in pediatric patients.

Authors:  P C Gillette; C Shannon; H Blair; A Garson; C J Porter; D G McNamara
Journal:  Am Heart J       Date:  1983-05       Impact factor: 4.749

5.  Comparison of modern steroid-eluting epicardial and thin transvenous pacemaker leads in pediatric and congenital heart disease patients.

Authors:  Elizabeth B Fortescue; Charles I Berul; Frank Cecchin; Edward P Walsh; John K Triedman; Mark E Alexander
Journal:  J Interv Card Electrophysiol       Date:  2005-10       Impact factor: 1.900

6.  Salvage of a failing bifurcated bipolar epicardial lead with conductor fracture.

Authors:  Alexander Rusanov; Henry M Spotnitz
Journal:  Ann Thorac Surg       Date:  2010-08       Impact factor: 4.330

7.  Durability of repaired sensing leads equivalent to that of new leads in implantable cardioverter defibrillator patients with sensing abnormalities.

Authors:  Srijoy Mahapatra; Munthur K Homoud; Paul J Wang; N A Mark Estes; Mark S Link
Journal:  Pacing Clin Electrophysiol       Date:  2003-12       Impact factor: 1.976

8.  Permanent endovenous pediatric pacing: absence of lead failure--20 years follow-up study.

Authors:  Petar L Stojanov; Dragutin V Savic; Mirjana B Zivkovic; Zarko R Calovic
Journal:  Pacing Clin Electrophysiol       Date:  2008-09       Impact factor: 1.976

9.  Pacing activity, patient and lead survival over 20 years of permanent epicardial pacing in children.

Authors:  John D R Thomson; Michael E Blackburn; Carin Van Doorn; Ann Nicholls; Kevin G Watterson
Journal:  Ann Thorac Surg       Date:  2004-04       Impact factor: 4.330

10.  Predictors affecting durability of epicardial pacemaker leads in pediatric patients.

Authors:  Hiroomi Murayama; Masanobu Maeda; Hajime Sakurai; Akihiko Usui; Yuichi Ueda
Journal:  J Thorac Cardiovasc Surg       Date:  2008-02       Impact factor: 5.209

View more
  1 in total

1.  Reconstruction of the terminal of an abandoned fractured unipolar coronary sinus lead: a feasible solution to restore effective cardiac resynchronization therapy.

Authors:  Armando Gardini; Francesco Fracassi; Alberto Saporetti; Davide Mariggio
Journal:  Indian Pacing Electrophysiol J       Date:  2013-06-25
  1 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.