Literature DB >> 15063268

Pacing activity, patient and lead survival over 20 years of permanent epicardial pacing in children.

John D R Thomson1, Michael E Blackburn, Carin Van Doorn, Ann Nicholls, Kevin G Watterson.   

Abstract

BACKGROUND: We report on pediatric epicardial pacing activity, patient and lead survival for more than two decades in a single center.
METHODS: The data cover 96 pacing leads implanted in 59 patients. Median age at implantation was 1.9 years (1 day to 18.2 years). Forty-four percent had structural cardiac disease. Most frequent indications for pacing were postoperative (42%) and congenital complete heart block (42%).
RESULTS: Median activity was 3 pacing leads per year; 326 patient pacing years were observed (median 11.9 years; range, 1.1 to 22 years). Death due to pacemaker failure occurred in a single patient. Lead failure occurred in 33 of 96 leads (median of 28 months postimplantation) with lead fracture the commonest cause (47%). Risk factors for lead failure were decade of implant and nonsteroid eluting leads. Acute implant energy thresholds were significantly lower for steroid than nonsteroid eluting leads but did not predict subsequent lead failure.
CONCLUSIONS: The epicardial approach has offered an effective solution to pacing problems in the pediatric age range. Lead survival has improved with more than 75% of modern steroid eluting leads surviving to 5 years.

Entities:  

Mesh:

Year:  2004        PMID: 15063268     DOI: 10.1016/j.athoracsur.2003.08.057

Source DB:  PubMed          Journal:  Ann Thorac Surg        ISSN: 0003-4975            Impact factor:   4.330


  6 in total

1.  Durability of repaired pacemaker leads in the pediatric population.

Authors:  Sarah Chambers; Alex Rusanov; Henry M Spotnitz; Eric S Silver; Leonardo Liberman
Journal:  J Interv Card Electrophysiol       Date:  2011-02-16       Impact factor: 1.900

Review 2.  Resynchronization Therapy for Patients with Congenital Heart Disease: Are We Ready for Prime Time?

Authors:  Scott Anjewierden; Peter F Aziz
Journal:  Curr Cardiol Rep       Date:  2018-07-18       Impact factor: 2.931

3.  Permanent complete heart block following surgical correction of congenital heart disease.

Authors:  F Edwin; E Aniteye; M Tettey; L Sereboe; D Kotei; M Tamatey; K Entsua-Mensah; K Frimpong-Boateng
Journal:  Ghana Med J       Date:  2010-09

4.  Fractally coated myocardial pacemaker leads in children.

Authors:  Thomas Borth-Bruhns; Matthias Gass
Journal:  J Interv Card Electrophysiol       Date:  2005-10       Impact factor: 1.900

5.  Comparison of modern steroid-eluting epicardial and thin transvenous pacemaker leads in pediatric and congenital heart disease patients.

Authors:  Elizabeth B Fortescue; Charles I Berul; Frank Cecchin; Edward P Walsh; John K Triedman; Mark E Alexander
Journal:  J Interv Card Electrophysiol       Date:  2005-10       Impact factor: 1.900

6.  Right-sided subcutaneous implantable cardioverter-defibrillator placement in a patient with dextrocardia, tetralogy of Fallot, and conduction disease.

Authors:  Scott R Ceresnak; Kara S Motonaga; Ian S Rogers; Mohan N Viswanathan
Journal:  HeartRhythm Case Rep       Date:  2015-06-12