Salvage of a failing bifurcated bipolar epicardial lead with conductor fracture.

Alexander Rusanov1, Henry M Spotnitz.   

Abstract

Cardiac rhythm abnormalities in children are often managed by using epicardial pacing systems. Lead failure poses a serious problem in pacemaker-dependent patients. We describe a lead repair that was not intuitive and avoided repeat thoracotomy in a pacemaker-dependent patient with lead fracture. Copyright 2010 The Society of Thoracic Surgeons. Published by Elsevier Inc. All rights reserved.

Entities:  

Mesh:

Year:  2010        PMID: 20667373     DOI: 10.1016/j.athoracsur.2010.02.033

Source DB:  PubMed          Journal:  Ann Thorac Surg        ISSN: 0003-4975            Impact factor:   4.330


  2 in total

1.  Acute and chronic high-frequency properties of cardiac pacing and defibrillation leads.

Authors:  Danko Tomasic; Bozidar Ferek-Petric; Sandro Brusich; Agostino P Accardo
Journal:  Med Biol Eng Comput       Date:  2012-06-24       Impact factor: 2.602

2.  Durability of repaired pacemaker leads in the pediatric population.

Authors:  Sarah Chambers; Alex Rusanov; Henry M Spotnitz; Eric S Silver; Leonardo Liberman
Journal:  J Interv Card Electrophysiol       Date:  2011-02-16       Impact factor: 1.900